** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit covers a very wide high-temperature range up to 300 °C, so it suits processes needing hot-side temperature control rather than only near-ambient work.
Defines whether the circulator can meet the process temperature window at the selection stage.
Source
“Operating temperature range 65... 300 °C”
Performance
With 48 kW heating and 30 kW cooling at 300 °C, the unit is a high-capacity machine that requires proper utility and heat-rejection planning.
Heating and cooling capacity drive utility sizing, heat exchanger selection and plant load calculations.
Source
“Heating power 48 kW”
Installation
The unit needs a three-phase 400 V 50 Hz supply, so the electrical installation must be planned for industrial three-phase power.
Determines the electrical supply, protection devices and cabling required at the installation site.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide the final connection.
Affects site wiring, isolator provision and who supplies the plug or terminal connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are linked, and some combinations such as -5% voltage with +2% frequency are not permitted.
Prevents commissioning on a supply that appears within individual tolerances but is outside the allowed combined envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
Leakage current exceeds 3.5 mA, which matters for protective earthing and RCD selection.
Influences earth connection design and residual current device choice for safe operation.
Source
“Attention: leakage current > 3,5mA”
Conformité
The equipment is EMC Class A Group 1, so it is intended for industrial environments and may cause disturbance in residential areas.
Determines whether the installation location is appropriate from an EMC standpoint.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit is rated IP20, so it must be installed in a dry, protected location away from water ingress.
Restricts where the unit can be placed and what environmental protection is needed.
Source
“Degree of Protection IP20”
Operation
Ambient temperature must stay between 5 °C and 40 °C, so the room climate must be controlled.
Out-of-range ambient conditions can affect performance, reliability and warranty.
Source
“min. ambient temperature 5 °C”
Installation
The unit is large at roughly 800 x 1007 x 2093 mm, so space and access must be planned before delivery.
Needed for room layout, transport route and maintenance access planning.
Source
“800x1007x2093 mm”
Selection
The pump delivers up to 200 l/min at 5.5 bar, so external circuit pressure drop and flow requirements must be checked.
Confirms whether the pump can overcome the pressure drop of the customer's loop.
Source
“max. delivery pressure 5,5 bar”
Selection
The maximum permissible kinematic viscosity is 50 mm²/s, which constrains the choice of thermofluid at low temperatures.
Prevents selecting a fluid too viscous for the pump at the lowest operating temperature.
Source
“max. permissible kin. viscosity 50 mm²/s”
Maintenance
The expansion tank holds 56 l and is not thermoregulated, which affects fluid volume planning and expansion allowance.
Determines fill volume, expansion space and whether the tank suits the chosen system type.
Source
“Expansion tank (not thermoregulated) for closed systems, lockable for open baths.”
Safety
The unit has adjustable overtemperature protection per DIN 12876, which must be set correctly for the process.
The safety cut-out setting is a key commissioning step to protect the process and equipment.
Source
“With adjustable overtemperature protection according to DIN 12876.”