Unistat T350w HT Temperature Control System Datasheet
Sommaire
Overviewp. 1
Product name and general introduction to the Unistat T350w HT unit
Specpp. 1-2
Technical data per DIN 12876 including performance and dimensions
p. 2table, 3 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
min. ambient temperature
5 °C
from Serial-No.:
S619282
1.0/26
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
65...300 °C
Minimum temperature with water cooling
15 °C
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
96 kW
Cooling power at 300°C
60 kW
Cooling power at 200°C
60 kW
Cooling power at 100°C
30 kW
Circulation pump: MK pump
max. delivery
200 l/min
max. delivery pressure
5,5 bar
Pump connection
M38x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1 1/4 male
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
33 l
Filling capacity expansion tank
56 l
Overall dimensions WxDxH **
800x1007x1606 mm
Net weight
460 kg
Power supply (3 Phase)
460V 3~ 60Hz
max. current (3 Phase)
130 A
Fuse (3 phase)
3x160A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
max. ambient temperature
40 °C
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 460 V 3-phase 60 Hz supply with a 130 A max current and 3x160 A fusing, so the electrical infrastructure must be sized accordingly.
Determines feeder, breaker and fuse sizing before the unit can be installed.
Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation
Three-phase devices drawing more than 63 A are delivered without cable and without plug, so the installer must provide the power connection.
Affects installation scope and cost; cabling and plug must be procured separately.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Installation
Cooling water must be supplied at 1 bar minimum differential pressure and no more than 6 bar, with G1 1/4 male connections.
Defines the cooling water supply requirements and connection size for the site.
Source
“min. cooling water differential pressure 1 bar”
Operation
The unit is designed for cooling water intake between 5 °C and 20 °C, and cooling power drops as the cooling water temperature rises.
Cooling water temperature directly affects achievable cooling capacity and must be verified at the site.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.
Prevents condensation and corrosion damage inside the unit during cold water operation.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Selection
The Unistat T models have no active refrigeration system, so reaching 65 °C requires adequate heat loss from the application or suitable water cooling.
Critical for selection: the unit cannot cool below 65 °C on its own and depends on the process heat loss.
Source
“These devices do not have an active refrigeration system.”
Selection
For air-cooled versions the application heat loss must be between 0.1 kW and 2.8 kW depending on the unit, and ambient temperature must be well below 60 °C to reach 65 °C.
Determines whether an air-cooled variant is viable for the application's heat load.
Source
“The heat loss must be between 0.1kW and 2.8kW, depending on the unit.”
Safety
Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection.
Affects electrical protection design and compliance with local installation rules.
Source
“Attention: leakage current > 3,5mA”
Conformité
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.
Must be checked against the site power quality to avoid invalid operating conditions.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The pump connection is M38x1.5 male and pipework must use ball socket per DIN 3863 with sleeve nut per DIN 3870.
Ensures correct mating of hoses and piping to the circulation pump.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation
The maximum permissible kinematic viscosity of the thermal fluid is 50 mm²/s, limiting fluid choice at low temperatures.
Constrains thermofluid selection and low-temperature operability of the pump.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The unit has IP20 protection and a maximum ambient temperature of 40 °C, so it must be installed in a dry, protected location.
Restricts installation environment and enclosure requirements.
Source
“Degree of Protection IP20”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.
Guides cooling water quality and treatment to avoid corrosion and fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Performance
Cooling power is 60 kW at 300 °C and 200 °C but drops to 30 kW at 100 °C, which matters for sizing at lower set points.
Cooling capacity is temperature-dependent and must be checked at the actual operating point.