CT-Series Freeze Dryers Manual

Technical Reference Summary

Covered Models & Document Scope

Covers Labconco CT Freeze Dryers, 2025—Present: 8700xxx (-50C Series), 8701xxx (-84C Series), 8702xxx (-105C Series). Catalog numbers use X = 1, 2, 5 or 6; voltage/frequency variants are indicated by suffix 00 (115V/60Hz), 10 (230V/60Hz), 15 (230V/50Hz), and 30/40/50/70/80/90 (230V/50Hz with different plug types). Manifold options are 4-port (7232800) and 8-port (7232900).

Section-by-Section Technical Breakdown

pp. 7, 18
Location, Space and Clearance Requirements

The Cold Trap must be placed on a stable, flat, level surface. Refrigeration performance is adversely affected by high ambient temperature; optimal ambient is 70–75°F (21–24°C). A minimum clearance of 3 inches (7.6 cm) is required between the sides/back and adjacent walls or equipment because airflow enters the right side and exits the left side of the cabinet.

  • Stable, flat, level surface required.
  • Optimal ambient room temperature: 70–75°F (21–24°C).
  • Minimum clearance: 3 inches (7.6 cm) between sides/back and adjacent surfaces.
  • Airflow: drawn into right side of cabinet, exhausts out left side.
  • Restriction of cabinet airflow adversely affects performance.
pp. 8
Electrical Requirements

Electrical requirements are tabulated by catalog number family. All models are single-phase and require a dedicated outlet with an appropriate circuit breaker and a protective earthing ground connection. Typical operating currents range from 2.6 A to 13 A depending on model and voltage.

  • 8700X00: 5.9 A typical, 115V/60Hz, 7A circuit, 1 phase.
  • 8700X30/-40/-50/-70/-80/-90: 2.6 A typical, 230V/50Hz, 3A circuit, 1 phase.
  • 8700X10/-15: 2.9 A typical, 230V/60Hz, 3A circuit, 1 phase.
  • 8701X00: 11.2 A typical, 115V/60Hz, 12A circuit, 1 phase.
  • 8701X30/-40/-50/-70/-80/-90: 5.2 A typical, 230V/50Hz, 6A circuit, 1 phase.
  • 8701X10/-15: 5.6 A typical, 230V/60Hz, 6A circuit, 1 phase.
  • 8702X00: 13 A typical, 115V/60Hz, 13A circuit, 1 phase.
  • 8702X30/-40/-50/-70/-80/-90: 6.3 A typical, 230V/50Hz, 7A circuit, 1 phase.
  • 8702X10/-15: 6.5 A typical, 230V/60Hz, 7A circuit, 1 phase.
  • Dedicated outlet must provide protective earthing ground.
pp. 8, 9
Exhaust and Vacuum Pump Requirements

The Cold Trap should be located in a fume hood when hazardous or flammable solvents are used, and the vacuum pump should be vented to a fume hood. The vacuum pump inlet fitting must accept 0.50 ID hose, and the pump must achieve less than 0.02 mbar vacuum for freeze drying. Flammable solvents require an explosion-proof or solvent-suitable pump; aggressive vapors require PTFE or PTFE-coated wetted parts.

  • Vacuum pump inlet fitting suitable for 0.50 ID hose.
  • Pump capable of achieving less than 0.02 mbar vacuum.
  • Flammable solvents: explosion-proof or solvent-suitable pump recommended.
  • Aggressive solutions: internal wetted parts should be PTFE or PTFE-coated.
  • Contact Labconco before freeze drying acids.
  • Cold Trap not evaluated for biological, radio toxins, or flammable liquids/materials.
pp. 18, 19
Installation and Manifold Assembly

The Cold Trap must always be placed before the vacuum pump. Installation requires a 5/16-inch nut driver or flat-head screwdriver. The manifold gasket is centered on the chamber, the manifold is secured with two latches, a 3-inch vacuum tubing piece connects to a vacuum gauge (or a plug fitting replaces the barb), and the other barb fitting connects to the vacuum pump.

  • Cold Trap always placed before the vacuum pump.
  • Minimum 3 inches between sides and adjacent walls.
  • Tools: 5/16-inch nut driver or flat-head screwdriver.
  • Center manifold gasket on chamber; secure manifold with two latches.
  • 3-inch tubing for vacuum gauge; replace barb with plug if no gauge.
  • Other barb fitting must connect to vacuum pump.
  • Secure hoses with supplied clamps.
pp. 20
Electrical Connection and Grounding

The power cord is plugged into the receptacle on the back of the Cold Trap and into a suitable power receptacle; the vacuum pump is plugged into a separate suitable receptacle. A ground wire must be attached to the stainless-steel elbow on the drain tubing, and the grounding clip must be attached to the solvent catch pot when draining to eliminate electrostatic spark ignition risk.

  • Power cord connects to rear receptacle of Cold Trap.
  • Vacuum pump has its own power cord to a suitable receptacle.
  • Ground wire attaches to stainless-steel elbow on drain tubing.
  • When draining, attach grounding clip to solvent catch pot.
  • Grounding prevents electrostatic spark ignition.
pp. 21, 22
Freeze Dry Process and Rates

Freeze drying removes water or solvent from frozen material by sublimation. Efficiency depends on sample surface area and thickness, collector temperature, vacuum obtained, eutectic point, and solute concentration. The collector should be at least 15–20°C colder than the eutectic temperature, and a vacuum of 0.133 mbar or less is recommended in many applications. Table 5-1 lists safe temperatures and approximate drying times for selected materials.

  • Collector at least 15–20°C colder than eutectic temperature.
  • Recommended vacuum: 0.133 mbar or less.
  • Greater surface area increases rate; thicker samples dry slower.
  • Flask volume should be 2–3 times sample volume.
  • Table 5-1: Milk -5°C safe, -40°C collector, 10 h; Blood Plasma -10 to -25°C, -40°C, 16 h; Serum -25°C, -40°C, 18 h; Vaccinia -30 to -40°C, -50°C, 22 h; Influenza Vaccine -30°C, -50°C, 24 h; Human Tissue -30 to -40°C, -50°C, 48 h; Vegetable Tissue -50°C, -80°C, 60 h.
pp. 23
Freeze Dry Capacity

Capacity depends on sample surface area, eutectic temperature and concentration, and heat transfer. Eutectic temperature is the most important factor; samples with eutectic temperatures of –20°C or lower should be added one flask at a time to allow vacuum recovery. If vacuum does not recover, capacity has been exceeded. Samples with eutectic temperatures of –40 to –60°C require cascade refrigeration capable of below –75°C or a dry ice/solvent trap.

  • Samples with eutectic temperature ≤ –20°C: add one flask at a time.
  • If vacuum does not recover, remove sample; capacity exceeded.
  • Eutectic temperature –40 to –60°C: use cascade refrigeration below –75°C or dry ice/solvent trap.
  • Most biological samples have eutectic temperature –10 to –25°C.
  • Sugar-containing or tissue samples may have eutectic temperature –30 to –50°C.
pp. 24
Volatile Samples and Chemical Resistance

Samples containing volatile substances such as acetonitrile, methanol, acetic acid, formic acid, or pyridine tend to melt more readily; dilution with water helps keep them frozen. The Cold Trap is chemically resistant but some materials are degraded by corrosive chemicals; the manual provides a compatibility table with C (moderate degradation/questionable use) and D (severe degradation/infrequent use, immediate cleaning) ratings for chamber, lid, and gasket materials.

  • Volatile substances increase vapor pressure and melting tendency.
  • Dilute volatile samples with water to keep frozen.
  • Chemical resistance table rates components for acids, bases, and solvents.
  • C = moderate degradation/questionable use; D = severe degradation/infrequent use, immediate thorough cleaning required.
  • After each run, clean and rinse all residues from chamber.
pp. 26, 27
Start-up and Shutdown

Start-up includes verifying hose connections, securing the manifold with two latches, turning power on (amber indicator illuminates; green when at operating temperature), closing all valves, starting the vacuum pump, confirming vacuum after 5 minutes, and adding pre-frozen samples one at a time with about 5 minutes recovery. Shutdown includes venting samples, turning off pump and Cold Trap, allowing ice to melt without chipping, draining immediately, flushing with water, and not starting the pump with liquid in the trap.

  • Top amber indicator illuminates on power ON; green when operating temperature reached.
  • After 5 minutes of pump operation, confirm manifold lid is pulled down by vacuum.
  • Add pre-frozen samples one at a time; allow approx. 5 minutes vacuum recovery.
  • Max instantaneous ice load capacity: -50C 1.6 L; -84C 2.1 L; -105C 1.6 L.
  • Do not chip ice off Cold Trap walls.
  • Drain immediately after ice melts and flush with water.
  • Do not start vacuum pump with liquid in Cold Trap or vacuum tubing.
  • For rotary vane pump, change oil if cloudy, particles, or discolored; gas ballast open for extended period after run with inlet blocked.
pp. 29, 30
Recommended Maintenance Schedule

Maintenance includes emptying the trap after each run, cleaning spills, cleaning manifold and gasket with mild non-abrasive cleaner, checking vacuum pump oil level and condition, immediate cleaning after acid use, checking protective earth continuity annually, checking hoses and gaskets, cleaning exterior, and cleaning the condenser with a vacuum brush.

  • After each run: completely empty trap; clean spills; clean manifold and gasket.
  • Monthly: check oil level between MIN and MAX; replace if cloudy/particles/discolored.
  • Acid use requires immediate cleaning and neutralization after each run.
  • Annually: check continuity of protective earth ground; clean condenser with vacuum brush.
  • Check hoses and gaskets; replace if hardened, permanently set, or deteriorated.
  • Clean exterior with cloth and non-abrasive cleaner; solvents may damage finish.
pp. 33, 34, 35, 36
Parts List

Appendix A lists replaceable parts for -50C, -84C, and -105C Cold Traps and for the 4/8-port manifold. Key parts include circuit boards, switches, circuit breakers, valves, ground wire assemblies, timers, temperature sensors, manifold knobs, valve bodies, trim rings, gaskets, tee fittings, tubing, hose clamps, and plugs.

  • -50C: circuit board 7394400; circuit breaker 115V 1327208 (3A), 230V 1327204; valve 1360500; ground wire 7464600; temperature sensor 7515300.
  • -84C: circuit board 7394402; circuit breaker 115V 1327207 (3A), 230V 1327208; timer 6-minute delay 115V 7474200, 230V 7474400.
  • -105C: circuit board 7394402; circuit breaker 115V 1289315 (3A), 230V 1289308; timer 6-minute delay 115V 7474200, 230V 7474400; ON/OFF recycle timer 115V 7480000, 230V 7480001.
  • Manifold: 4-port complete 7232800; 8-port complete 7232900; knob & stem 7508600; valve body 7505600; trim ring 7506500; gasket 7399400; tee fitting 1548800; stem tube-to-hose 1554700; tubing 7828608; hose clamp 1488800; plug 1548302.
pp. 37, 38, 39, 40
Dimensions

Appendix B provides overall product dimensions in inches (centimeters) for -50C, -84C, and -105C Cold Traps and for the 4-port and 8-port manifolds.

  • -50C dimensions (in/cm): 24.3/61.6, 11.1/28.1, 13.8/35.0, 16.0/40.7, 14.5/36.8.
  • -84C dimensions (in/cm): 22.1/56.1, 18.2/46.2, 21.0/53.3, 14.5/36.8, 16.3/41.3.
  • -105C dimensions (in/cm): 23.9/60.8, 14.9/38.0, 17.6/44.8, 24.3/61.6, 22.5/57.2.
  • 4-port manifold: 8.7/22.2, 13.7/34.9, 8.8/22.3, 8.0/20.3.
  • 8-port manifold: 8.8/22.3, 8.0/20.3.
pp. 41
Specifications

Appendix C provides normal operating power and environmental conditions. Normal operating power is 280 W for 8601xxx, 650 W for 8611xxx, and 700 W for 8621xxx. Environmental conditions include indoor use only, ambient 41° to 104°F (5° to 40°C), maximum 80% RH decreasing linearly to 50% at 104°F, main supply voltage fluctuations not exceeding ±10%, transient overvoltages per Installation Category II, and Pollution Degree 2.

  • Normal operating power: 8601xxx 280 W; 8611xxx 650 W; 8621xxx 700 W.
  • Indoor use only.
  • Ambient temperature range: 41° to 104°F (5° to 40°C).
  • Maximum relative humidity: 80% up to 88°F (31°C), decreasing linearly to 50% at 104°F (40°C).
  • Main supply voltage fluctuations not to exceed ±10% of nominal.
  • Transient overvoltages per Installation Category II: up to 1500V for 115V models, 2500V for 230V models.
  • Pollution Degree 2 environment.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication S'applique à
Optimal ambient room temperature for the Cold Trap is 70–75°F (21–24°C). p.7 ↗ Install in a temperature-controlled room; higher ambient reduces refrigeration performance. All CT Freeze Dryer models
Minimum clearance of 3 inches (7.6 cm) between sides/back and adjacent walls or equipment. p.7 ↗ Maintain airflow; air enters right side and exits left side of cabinet. All CT Freeze Dryer models
8702X00 requires 115V, 60Hz, 13A dedicated circuit; typical operating current 13 A. p.8 ↗ Use a dedicated 13A circuit with protective earth for -105C 115V models. 8702X00
8700X30/-40/-50/-70/-80/-90 require 230V, 50Hz, 3A circuit; typical operating current 2.6 A. p.8 ↗ Lowest current draw among models; 3A circuit sufficient for -50C 230V/50Hz variants. 8700X30, -40, -50, -70, -80, -90
Vacuum pump must achieve less than 0.02 mbar vacuum and inlet fitting must accept 0.50 ID hose. p.8 ↗ Select a pump meeting these minimum vacuum and hose connection requirements. All CT Freeze Dryer models
Recommended operating vacuum for many applications is 0.133 mbar or less. p.22 ↗ Maintain vacuum at or below this level for efficient freeze drying. All CT Freeze Dryer models
Collector should be at least 15–20°C colder than the eutectic temperature of the sample. p.22 ↗ Select -50C, -84C, or -105C model based on sample eutectic temperature. All CT Freeze Dryer models
Max instantaneous ice load capacity: -50C 1.6 L; -84C 2.1 L; -105C 1.6 L. p.26 ↗ Do not exceed these volumes; add samples one at a time and allow vacuum recovery. -50C, -84C, -105C models
After starting vacuum pump, wait 5 minutes and confirm manifold lid is pulled down by vacuum. p.26 ↗ Verifies vacuum integrity before adding samples. All CT Freeze Dryer models
Allow approximately 5 minutes vacuum recovery between adding pre-frozen samples. p.26 ↗ Prevents collector overload and sample melting. All CT Freeze Dryer models
Samples with eutectic temperatures of –20°C or lower should be added one flask at a time. p.23 ↗ If vacuum does not recover, capacity is exceeded and sample should be removed. All CT Freeze Dryer models
Samples with eutectic temperatures of –40 to –60°C require cascade refrigeration capable of below –75°C or a dry ice/solvent trap. p.23 ↗ Standard -50C model may be insufficient; use -84C/-105C or dry ice trap. All CT Freeze Dryer models
Flask volume should be two to three times the sample volume. p.22 ↗ Prevents overfilling and allows proper shell freezing. All CT Freeze Dryer models
Do not store flammable or hazardous solvents within 12 inches (300 mm) of the CentriVap Cold Trap. p.15 ↗ Maintain safe distance to reduce fire/ignition risk. All CT Freeze Dryer models
When draining, attach grounding clip to solvent catch pot to eliminate electrostatic spark ignition risk. p.20 ↗ Mandatory grounding step during drain procedure. All CT Freeze Dryer models
Do not start vacuum pump when Cold Trap contains liquid or if liquid is present in vacuum tubing. p.27 ↗ Liquid can be drawn into pump and damage it. All CT Freeze Dryer models
For rotary vane pump, change oil if cloudy, shows particles, or is discolored; oil level between MIN and MAX. p.29 ↗ Frequent oil changes required; gas ballast open for 15 min after run with inlet blocked. All CT Freeze Dryer models with rotary vane pump
Normal operating power: 8601xxx 280 W; 8611xxx 650 W; 8621xxx 700 W. p.41 ↗ Use for heat load and facility power planning. 8601xxx, 8611xxx, 8621xxx
Main supply voltage fluctuations must not exceed ±10% of nominal; transient overvoltages up to 1500V (115V models) or 2500V (230V models) allowed per Installation Category II. p.41 ↗ Verify facility power quality meets these limits. All CT Freeze Dryer models
Ambient operating range: 41° to 104°F (5° to 40°C); max RH 80% up to 88°F (31°C), decreasing linearly to 50% at 104°F (40°C). p.41 ↗ Indoor use only; Pollution Degree 2 environment. All CT Freeze Dryer models

Faults & Troubleshooting

Code / Symptôme Possible Cause Remedy Note
Unit will not operate p.31 ↗ Unit not connected to electrical power Connect unit to proper electrical receptacle. Chapter 7
Unit will not operate p.31 ↗ Circuit breaker blown Correct electrical problem and reset circuit breaker by pressing button. Chapter 7
Sample odor in lab p.31 ↗ Vent hose exhausting into lab area Redirect hose to fume hood. Chapter 7
No vacuum/poor vacuum p.31 ↗ Pump not on Turn on pump. Chapter 7
No vacuum/poor vacuum p.31 ↗ Leaks in lines or connectors or gasket Locate and repair. Chapter 7
No vacuum/poor vacuum p.31 ↗ Foreign material on Manifold gasket Clean gasket. Chapter 7
No vacuum/poor vacuum p.31 ↗ Sample valve open Close sample valve Chapter 7
No vacuum/poor vacuum p.31 ↗ Drain valve open Close drain valve Chapter 7
No vacuum/poor vacuum p.31 ↗ Pump is not functioning properly Check pump by locating vacuum gauge closer to pump and close off rest of system. Check pump oil for cloudiness or particles and change. Chapter 7
No vacuum/poor vacuum p.31 ↗ Manifold gasket not seated Check position of manifold gasket on Cold Trap chamber. Chapter 7
Recovery of condensate in Cold Trap is less than normally expected p.32 ↗ Cold Trap is not ON Check to make sure switch is ON and condensing unit fan is moving air out of the rear of the cabinet. Chapter 7
Recovery of condensate in Cold Trap is less than normally expected p.32 ↗ Cold Trap does not cool down Turn vacuum pump OFF and allow Cold Trap to cool for at least 30 min. to reach temperature. Chapter 7
Recovery of condensate in Cold Trap is less than normally expected p.32 ↗ Temperature can be checked with a solvent thermometer or digital thermometer. Temperature can be checked with a solvent thermometer or digital thermometer. Chapter 7
Frequent oil change needed in pump p.32 ↗ Cold Trap is not emptied after each run and dried Empty the stainless steel cold trap and drain line and ensure they are dry before starting vacuum pump. Chapter 7
Frequent oil change needed in pump p.32 ↗ No outgassing of vacuum pump being performed After FD run is complete, continue to run vacuum pump with gas ballast open for 15 min Chapter 7
Frequent oil change needed in pump p.32 ↗ High solvent content in frozen samples Evaporate solvents out of sample before freeze drying process Chapter 7
Samples Melting p.32 ↗ Poor vacuum (vacuum level not low enough) Ensure pump is operating properly and/or check for leaks in the vacuum system. Chapter 7
Samples Melting p.32 ↗ Collector temperature not cold enough for samples Collector needs to be 10-15°C colder than the sample freezing point. Chapter 7
Samples Melting p.32 ↗ High solvent content Dilute samples. Evaporate solvents out of sample before freezing. Isolate the flask from room temp. with insulation or putting the flask inside a larger flask. Chapter 7