NovoCyte Flow Cytometer User Guide

Sommaire

  1. Overview pp. 6-9
    Instrument overview, workstation connection, compensation, laser configuration
  2. pp. 10-16
    Front matter, glossary of terms, and introductory material
  3. Warning pp. 17-28
    Warnings, laser classification, biosafety, EMC, regulatory compliance
  4. Overview pp. 29-50
    System introduction, fluidics, components, workstation, fluidics cart
  5. Procedure pp. 51-56
    Space requirements, instrument installation, workstation and cart connections
  6. Procedure pp. 57-70
    Power-up, configuration, QC test, and shutdown procedures
  7. Procedure pp. 71-104
    Experiment creation, acquisition parameters, compensation, data analysis
  8. Procedure pp. 105-134
    Preventative and unscheduled maintenance, cleaning, priming, filter care
  9. Procedure pp. 135-138
    Troubleshooting tables for common instrument issues
  10. pp. 139-140
    Document revision history
  11. Spec pp. 141-146
    Laser and detection channel configuration appendix
  12. Spec pp. 147-150
    Instrument, power adapter, workstation, and monitor specifications
  13. p. 141 table, 8 rows
  14. p. 147 table, 19 rows
  15. p. 143 table, 10 rows
  16. p. 150 table, 6 rows
  17. p. 113 table, 6 rows
  18. p. 126 table, 9 rows
  19. p. 148 table, 19 rows
  20. p. 149 table, 17 rows

Table 15 Laser Configuration

Model Number405 nm488 nm561 nm640 nm
1000/1040/1050√
1000Y√
2000R/2060R/2070R√√
2070V/2090V/2110V√√
2100YB√√√
3000/3005√√√
3000VYB√√√
3000RYB√√√√

Table 21 NovoCyte Technical Specifications1

ParameterSpecification
Optics
LaserSolid state laser with on-board thermal-electric cooling and guaranteed thermal stability and beam quality.
Laser Power20 or 60 mW for 488 nm laser, 50 mW for 405 nm and 561 nm lasers, 40 mW for 640 nm laser
Laser Beam Profile10 x 80 µm elliptical beam.
Optical Alignment ProcedureFixed, no operator alignment required
Fluorescence DetectionPhotomultiplier Tube (PMT)
FSC/SSC SensitivityFSC: 0.5 µm; SSC: 0.2 µm
Fluorescence Resolution<3% CV for CEN
Optical FiltersUser Exchangeable
Cell Size0.2 - 50 µm
Fluidics
Flow Cell170 x 290 µm rectangular quartz flow cell
Sample Acquisition Rate35,000 events/second
Sample DeliveryPositive-displacement syringe pump enabling direct absolute count without the need for reference counting beads.
Volumetric Absolute Count Precision< 5%
Sample Flow Rate5-120 µL/min, continuously adjustable
Sheath Flow Rate6.5 mL/min
Sample Aspiration Volume5 µL – 5 mL
Compatibility to AutosamplerNo fluidic tubing disassembly or re-connection required.

1 Сноска в тексте страницы не приведена.

Table 17 Detection Channel Configuration1

2090V-Config 22090V-Config 32110V-Default2110V-Config 12110V-Config 22100YB-Default3000-Default3000-Config 13000VYB-Default3000VYB-Config 13000VYB-Config 2
VVVVVVVVVVVVV
VBVBVBVBVBBVBVBVBVBVBBBBBVBVBV
VBVBVBVBVBVBBVBV
BYVBVBVBYVBYVBYBYBYBY
VBVBVBVBVBBYVBVBVBYVBYBYBYBYR
VBBYVBRVBRVBYVBYBYRBYRBYRBVRBVRBVR
VBVBVBVBYBYRBYRBYR
VBVBVBBY
VBVBVBRVBRVYVYVYYRYRYRBVRBVRBVR
VBYBVRBVRBVR

1V: 405 nm Laser; B: 488 nm Laser; Y: 561 nm Laser; R: 640 nm Laser. Different models of NovoCyte system use different lasers for FSC and SSC detection.

Table 24 NovoCyte Fluidics Station Specifications

ParameterSpecification
Dimensions (with reagents containers) (W × D × H)11.4 × 9.8 × 16.5 in (29.0 × 25.0 × 42.0 cm)
Weight13.2 lb (6.0 kg)
NovoFlow sheath fluid container3 L
NovoClean solution container500 mL
NovoRinse solution container500 mL
Waste container3 L

Table 11 List of Fluidic System Consumables to be Replaced

NovoCyte V1 Fluidics SystemNovoCyte V2 Fluidics System
NovoFlow sheath fluid in-line filter (2 pcs)NovoFlow sheath fluid in-line filter (1 pcs)
Waste filter (1 pcs)Waste filter (1pcs)
NovoFlow sheath fluid inlet filter (1 pcs)NovoFlow sheath fluid inlet filter (1 pcs)
NovoRinse solution inlet filter (1 pcs)NovoRinse solution inlet filter (1 pcs)
NovoClean solution inlet filter (1 pcs)NovoClean solution inlet filter (1 pcs)
Sheath Pump Tubing (1 pcs)

Table 12 Unscheduled Maintenance and When to Perform

Unscheduled MaintenanceWhen to Perform
Priming• The CV of the data is larger than normal. • QC test failed. • No events showed up after sample collection started for certain period of time. • Instrument was not used for over one week.
Clean the External Surface of the InstrumentKeep the external surface of the instrument clean.
DecontaminationWhen suspecting contamination in the system.
Prime Fluidics System before Long-Term StorageNo plan to run the instrument for more than two weeks.
Prime Fluidics System after Long-Term StorageInstrument has not been used for more than two weeks.
Purge Fluidics System before ShipmentThe instrument needs to be shipped.
Calibrate the NovoCyte Fluidics Station or NovoCyte Fluidics CartAbnormal warning about the liquid level in the reagent containers is observed.
Clean the Optical FiltersThe optical filters get dirty.
Clean Fluidics System• The CV of the data is larger than normal. • QC test failed. • Excessive debris in display.

Table 21 NovoCyte Technical Specifications1

CharacteristicSpecification
Fluid Level SensingActive sensing using weight sensors with automated warnings when any fluid level is out of specified range.
Fluid Container Capacity3 L sheath, 3 L waste, 500 mL cleaning, 500 mL rinse. Optional large container for sheath (15 L) and waste (15 L)
Carryover< 0.1%
Sample Injection Probe (SIP) RinseAutomated flying collar wash of inner and outer SIP surface after each sampling.
Fluidics System MonitoringIn-line pressure sensor monitors the pressure in real time. Automated system recovery when pressure is out of range due to clogging.
Fluidics MaintenanceAutomated startup and shutdown with fluidic system cleaning. Automated user executable cleaning, debubble, rinse, unclog, priming. Automated execution of user definable maintenance sequence.
Data Acquisition SoftwareNovoExpress
ParametersHeight and Area for FSC, SSC and all Fluorescent Channels, Width off FSC and Time.
Dynamic Range24 bits; 7.2 decades logarithmic scale
Fluorescence Photodetector Gain Control0-1000 User adjustable, optimized default gain setting for each individual channel.
CompensationFull inter-beam matrix, during or post acquisition
File ExportFCS 3.0, FCS 3.1; CSV; Batch PDF report
Data ReportAutomatic report. Customizable report. Batch PDF report.
Usage MonitorComprehensive Transaction Log and System Log.
User ManagementAdministrative creation of individual user accounts and user groups with privilege control. Configurable roles for individual users. User operation time tracking.
Quality Control Instrument Daily QCAutomatic daily quality control with Agilent Flow Cytometer QC Particles. Levey-Jennings Plot.
Production QCEach instrument is verified for assembly integrity and performance to specifications.
Sampling Manual Sample Loading12 X 75 mm tube; 1.5 mL or 2.0 mL tube with Agilent tube holder
Automatic Sample Loading with NovoSampler (Optional)Agilent Flow Cytometer 24 tube rack for 12 x 75 mm tube, 24-well, 48-well, 96-well (flat, V-, U-bottom) microtiter plates

1 Сноска в тексте страницы не раскрыта.

Table 21 NovoCyte Technical Specifications1

SpecificationValue
Automatic Sample Loading with NovoSampler Pro (Optional)Agilent Flow Cytometer 24 tube rack for 12 x 75 mm tube, 24-well, 48-well, 96-well (flat, V-, U-bottom) microtiter plates
Operating Conditions
Instrument Dimension (W × D × H)23.6 × 17.7 × 15.4 in (60.0 × 45.0 ×39.0 cm)
29.5 × 17.7 × 15.4 in (75.0 × 45.0 ×39.0 cm) with NovoSampler
31.9 × 17.7 × 15.4 in (81.0 × 45.0 ×39.0 cm) with NovoSampler Pro
Instrument Weight86.0 lb (39.0 kg)
101.4 lb (46.0 kg) with NovoSampler
105.8 lb (48.0 kg) with NovoSampler Pro
Operation RequirementsTemperature:15-32℃; Relative Humidity: 80% maximum; Atmosphere Range: 76-106 kPa
Storage RequirementsTemperature:1-40℃; Relative Humidity: 80% maximum; Atmosphere Range: 76-106 kPa
Power Requirements100-240 VAC, 50-60 Hz
Regulatory ComplianceIndoor and Outdoor Use
Indoor use only
Overvoltage CategoryClass II
Equipment ClassClass I Laboratory Equipment
Pollution degree of the intended environment2
Degree of Ingress protectionIPXO

1Specification and performance claims would be based on certain specific experiment conditions.

Engineering Facts

13 facts · every claim linked to a page
Selection

The instrument is available with either V1 or V2 fluidics, which differ in pump type, filter count, and pressure sensor count, so the specific variant must be confirmed before specifying consumables or service procedures.

Determines which fluidics consumables, filters, and pressure sensors apply to the purchased unit.

Source
“NovoCyte flow cytometer is equipped with either V1 or V2 fluidics system.”
Installation

The instrument weighs more than 23 kg, so two-person lifting or mechanical assistance is required during installation and any relocation.

Affects site preparation, manpower planning, and bench/floor load during installation.

Source
“This product is large and the weight is more than 51 lb (23 kg).”
Installation

The instrument must be connected only to a three-pin socket with reliable protective ground, otherwise shock protection is compromised.

Electrical installation must provide a grounded three-pin outlet at the instrument location.

Source
“the power cord must be inserted in the three-pin socket with a protective ground connection”
Safety

The instrument is a Class I laser product with lasers fully enclosed, so no special laser work area is needed except during service by qualified Agilent personnel.

Laser safety classification determines lab signage, access control, and service restrictions.

Source
“Agilent flow cytometer is a Class I laser product according to IEC/EN 60825-1:2014.”
Operation

The fluidics station LED turns red and prohibits operation when NovoFlow is below 0.7 L, NovoRinse/NovoClean below 100 mL, or waste exceeds 3 L.

Defines minimum reagent volumes and maximum waste capacity that must be planned for uninterrupted runs.

Source
“Reagent volume is too low to run experiments (NovoFlow < 0.7 L; NovoRinse < 100 mL; NovoClean < 100 mL), or the waste container is full (> 3 L).”
Operation

The fluidics station LED turns orange when NovoFlow is between 0.7 and 0.9 L, NovoRinse/NovoClean between 100 and 150 mL, or waste between 2.8 and 3 L, warning that refill or empty is needed soon.

Sets the refill/empty thresholds that operators must monitor to avoid aborted experiments.

Source
“0.7 L < NovoFlow < 0.9 L; 100 mL < NovoRinse < 150 mL; 100 mL < NovoClean < 150 mL”
Operation

The optical configuration should be verified before each experiment and can be changed by replacing optical filters, after which a QC test with automatic photodetector gain adjustment is required.

Filter changes affect which fluorochromes can be measured and require a QC run to restore valid performance.

Source
“It is recommended for users to view and verify the current optical configuration before each experiment.”
Performance

For automatic fluorescence compensation, the default stop condition is 10,000 events or 20 µL at slow flow rate with FSC-H threshold above 100,000.

Defines the acquisition settings that must be used to obtain valid automatic compensation matrices.

Source
“Stop condition: total 10,000 events or 20μL sample volume.”
Performance

Compensation samples should be prepared at 1,000,000 to 2,000,000 per mL to ensure correct automatic fluorescence compensation.

Sample concentration directly affects the validity of the compensation matrix and downstream data quality.

Source
“it is recommended to prepare the compensation sample concentration to be 1,000,000 ~ 2,000,000 /mL.”
Maintenance

A QC test must be performed every time fluidic system consumables are replaced to verify system performance.

Makes QC test a mandatory step in the consumable replacement procedure, affecting maintenance workflow.

Source
“A QC Test must be conducted each time when fluidic system consumables are replaced in order to check the performance of the system.”
Maintenance

The SIP cleaning apparatus must be wiped with NovoClean (1X) and then with Milli-Q or 0.2 µm filtered DI water at least five times while the instrument is powered off.

Specifies the cleaning agents and repetition required to avoid carryover and SIP contamination.

Source
“Prepare a damp cloth soaked with Milli-Q or 0.2 µm filtered DI water. Wipe the SIP cleaning apparatus back and forth at least five times.”
Conformité

The instrument is Class A ISM equipment and may cause radio interference in domestic environments, so it should be installed in a commercial/industrial setting.

EMC class restricts the installation environment and may require coordination with other sensitive equipment.

Source
“This product complies with the requirement of CISPR 11, Group 1, Class A as radiation professional equipment.”
Safety

If a leak is detected, power must be cut by holding the front power button for about 6–7 seconds and Agilent support contacted immediately.

Defines the emergency shutdown action for fluid leaks that could damage electronics or create biohazard.

Source
“immediately turn off the power by pressing and holding the power button on the front of the instrument for about 6-7 seconds”