NovoCyte Penteon Flow Cytometer User Guide

Sommaire

  1. Overview pp. 6-9
    Document contents, version history, and manual overview
  2. Overview pp. 11-15
    Glossary and introductory information for the guide
  3. Warning pp. 16-28
    General, electrical, laser safety and regulatory compliance statements
  4. Overview pp. 29-52
    System requirements, components, instrument, workstation, and fluidics cart description
  5. Procedure pp. 53-58
    Space requirements and installation of instrument, workstation, and fluidics cart
  6. Procedure pp. 59-72
    Instrument configuration, operation procedures, and QC testing
  7. Procedure pp. 73-108
    Experiment creation, acquisition parameters, compensation, and data handling
  8. Procedure pp. 109-134
    Preventative and unscheduled maintenance, consumables replacement, and cleaning
  9. Procedure pp. 135-138
    Troubleshooting guidance and version history
  10. pp. 139-140
    Document revision history
  11. Spec pp. 141-144
    Laser and detection channel configuration appendix
  12. Spec pp. 145-148
    Instrument and workstation technical specifications
  13. p. 141 table, 1 rows
  14. p. 127 table, 9 rows
  15. p. 146 table, 19 rows
  16. p. 147 table, 30 rows
  17. p. 49 table, 1 rows

Table 13 Laser Configuration1

Detection Channel445/45525/45586/20615/20667/30695/40725/40780/60
U7V7B6Y6R4UVUVBUVBYUVBYUVBYRBYRUVBYRUVYR

1U: 349 nm Laser; V: 405 nm Laser; B: 488 nm Laser; Y: 561 nm Laser; R: 637 nm Laser.

Table 10 Unscheduled Maintenance and When to Perform

Unscheduled MaintenanceWhen to Perform
Debubble• The CV of the data is larger than normal. • QC test failed. • No events showed up after sample collection started for certain period of time. • Instrument was not used for over one week.
Clean the External Surface of the InstrumentKeep the external surface of the instrument clean.
DecontaminationWhen suspecting contamination in the system.
Prime Fluidics System before Long-Term StorageNo plan to run the instrument for more than two weeks.
Prime Fluidics System after Long-Term StorageInstrument has not been used for more than two weeks.
Purge Fluidics System before ShipmentThe instrument needs to be shipped.
Calibrate the NovoCyte Fluidics Station II or NovoCyte Fluidics CartAbnormal warning about the liquid level in the reagent containers is observed.
Clean the Optical FiltersThe optical filters get dirty.
Clean Fluidics System• The CV of the data is larger than normal. • QC test failed. • Excessive debris in display.

Table 16 NovoCyte Penteon Technical Specifications1

ParameterSpecification
Sheath Flow Rate6.5 mL/min
Sample Aspiration Volume5 µL - 5 mL
Sample RecoveryAvailable
Compatibility to AutosamplerNo fluidic tubing disassembly or re-connection required
Fluid Level SensingActive sensing using weight sensors with automated warnings when any fluid level is out of specified range
Fluid Container Capacity3 L sheath, 3 L waste, 500 mL cleaning, 500 mL rinse with Fluidics Station II. Optional large container for sheath (15 L) and waste (15 L) with Fluidics Cart.
Carryover< 0.1%
Sample Injection Probe (SIP) RinseAutomated flying collar wash of inner and outer SIP surface after each sampling
Fluidics System MonitoringIn-line pressure sensor monitors the pressure in real time. Automated system recovery when pressure is out of range due to clogging
Fluidics MaintenanceAutomated startup and shutdown with fluidic system cleaning. Automated user executable cleaning, debubble, rinse, unclog, priming. Automated execution of user definable maintenance sequence.
Data Acquisition SoftwareNovoExpress
ParametersHeight and Area for FSC, SSC and all Fluorescent Channels, Width off FSC and Time
Dynamic Range24 bits; 7.2 decades logarithmic scale
Fluorescence Photodetector Gain Control0-1000 User adjustable, optimized default gain setting for each individual channel
CompensationFull inter-beam matrix, during or post acquisition
File ExportFCS 3.0, FCS 3.1; CSV; Batch PDF report
Data ReportAutomatic report. Customizable report. Batch PDF report
Usage MonitorComprehensive Transaction Log and System Log
User ManagementAdministrative creation of individual user accounts and user groups with privilege control. Configurable roles for individual users. User operation time tracking

1 Сноска в тексте страницы не приведена.

Table 16 NovoCyte Penteon Technical Specifications1

Specification
Quality ControlInstrument Daily QC
Quality ControlAutomatic daily quality control with Agilent Flow Cytometer QC Particles. Levey-Jennings Plot
Quality ControlProduction QC
Quality ControlEach instrument is verified for assembly integrity and performance to specifications
SamplingManual Sample Loading
Sampling12 X 75 mm tube; 1.5 mL or 2.0 mL tube with Agilent tube holder
SamplingAutomatic Sample Loading with NovoSampler Q (Optional)
SamplingAgilent Flow Cytometer 40 tube rack for 12 x 75 mm tube, 24-well, 48-well, 96-well (flat, V-, U-bottom), 384-well microtiter plates, Agilent Flow Cytometer 24 tube cooling rack for 12 x 75 mm tube, Agilent Flow Cytometer 96-well (flat, U-, V-bottom) plate cooling box.
Operating ConditionsInstrument Dimension (W × D × H)
Operating Conditions24.4 × 18.1 × 18.8 in (62.0 × 46.0 × 48.0 cm)
Operating Conditions33.5 × 18.1 × 18.8 in (85.0 × 46.0 × 48.0 cm) with NovoSampler Q
Operating ConditionsInstrument Weight
Operating Conditions109.2 lb (49.5 kg)
Operating Conditions137.8 lb (62.5 kg) with NovoSampler Q
Operating ConditionsOperating Environment
Operating ConditionsTemperature:15-32℃; Relative Humidity: 80% maximum; Atmosphere Range: 76-106 kPa
Operating ConditionsStorage Environment
Operating ConditionsTemperature:1-40℃; Relative Humidity: 80% maximum
Operating ConditionsPower Requirements
Operating Conditions100-240 VAC, 50-60 Hz
Regulatory ComplianceIndoor and Outdoor Use
Regulatory ComplianceIndoor use only
Regulatory ComplianceOvervoltage Category
Regulatory ComplianceClass II
Regulatory ComplianceEquipment Class
Regulatory ComplianceClass I Laboratory Equipment
Regulatory CompliancePollution degree of the intended environment
Regulatory Compliance2
Regulatory ComplianceDegree of Ingress protection
Regulatory ComplianceIPXO

1Specification and performance claims would be based on certain specific experiment conditions.

Table 6 NovoCyte Fluidics Station II LED Status Indicator Information

NameColorMeaning
NovoCyte Fluidics Station II LED Status IndicatorGreenReagent volume is above the designated threshold and is sufficient to run experiments (NovoFlow > 0.9 L; NovoRinse > 150 mL; NovoClean > 150 mL), and the waste container is not full (< 2.8 L).

Engineering Facts

14 facts · every claim linked to a page
Installation

The instrument weighs over 23 kg, so plan for two-person lifting and proper technique during installation.

Determines manpower and rigging needed for safe placement during installation.

Source
“This product is large and the weight is more than 51 lb (23 kg).”
Installation

Moving the instrument after professional installation is discouraged and requires Agilent personnel to re-install.

Affects site planning and layout decisions to avoid costly relocation later.

Source
“moving the instrument is not recommended. If you need to move the instrument, be sure to contact Agilent Technologies professionals.”
Safety

The unit is a Class I laser product with lasers fully enclosed, so no special work area is needed except during service.

Confirms laser safety classification for lab siting and EHS approval.

Source
“Agilent flow cytometer is a Class I laser product according to IEC/EN 60825-1:2014.”
Safety

Only Agilent-supplied power cables may be used to maintain safety and EMC compliance.

Constrains power connection hardware during installation.

Source
“Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.”
Safety

A reliable protective earth on a three-pin grounded outlet is mandatory to avoid shock hazard.

Requires verifying grounded outlets at the installation site.

Source
“the power cord must be inserted in the three-pin socket with a protective ground connection.”
Operation

Priming is mandatory after adding reagents, otherwise air in the tubing degrades instrument performance.

Defines a required operational step to avoid bad data after reagent refill.

Source
“It's necessary to run Priming after adding instrument reagent.”
Operation

Reagent containers must be color-coded and correctly matched to fluidics ports to prevent malfunction.

Guides correct plumbing hookup during setup and reagent changes.

Source
“Incorrect connection will cause the instrument malfunction. The tubing and containers are color coded to prevent such errors.”
Operation

Sample flow rate is selectable from 5 to 120 µL/min with three presets, and Slow is advised for high-resolution work.

Informs assay setup and throughput versus resolution trade-offs.

Source
“You can also manually set a desired flow rate within the range of 5 µL/min to 120 µL/min by adjusting the sliding bar.”
Performance

The ADC digitizes pulses at 80 million samples per second with 24-bit resolution.

Defines signal digitization performance relevant to data quality expectations.

Source
“A high-accuracy high-speed 24-bit analog-to-digital convertor is used to process the pulse in a speed of 80 million times per second.”
Maintenance

Fluidic system consumables must be replaced every two months or when software prompts.

Sets a recurring maintenance schedule and consumable budget.

Source
“Replacing Fluidic System Consumables Every two months or when prompted by NovoExpress Software.”
Maintenance

Preventive maintenance including SIP cleaning and flow cell unclogging is required monthly.

Establishes monthly service workload for reliable operation.

Source
“Every month.”
Operation

Software warns when sheath, rinse, or clean volumes drop below thresholds and when waste exceeds 2.8 L.

Defines fluid capacity limits that drive lab logistics and waste handling.

Source
“when the waste volume exceeds a set threshold (2.8 L), the software will give a warning to remind the user to empty the waste container.”
Conformité

The instrument is Class A Group 1 ISM equipment, which may cause radio interference in domestic environments.

Affects EMC planning and placement in non-residential lab environments.

Source
“This product complies with the requirement of CISPR 11, Group 1, Class A as radiation professional equipment.”
Safety

Data is stored on the workstation hard drive and is unrecoverable if deleted or corrupted, so regular backups are required.

Drives IT backup provisioning decisions at installation.

Source
“Please regularly backup the data to a server or other mobile storage devices.”