Separation Resolution Capabilities of the Agilent ZAG DNA Analyzer

Sommaire

  1. Overview p. 1
    Introduces separation resolution capabilities of the ZAG DNA Analyzer
  2. Procedure p. 2
    Describes experimental conditions, kits, and methods used
  3. pp. 3-9
    Presents resolution results for various fragment sizes and kits
  4. Spec p. 10
    Table of sizing results for single fragments and mixes
  5. Overview p. 11
    Summarizes separation resolution performance findings
  6. p. 12
    Provides website and contact details
  7. p. 2 table, 7 rows
  8. p. 10 table, 26 rows

Table 1. Specification comparisons for the Agilent ZAG 135 dsDNA (1–1500 bp) and the Fragment Analyzer dsDNA 935 Reagent kits (1–1500 bp) and the ZAG 110 dsDNA (35–5000 bp) and the Fragment Analyzer dsDNA 915 Reagent kits (35–5000 bp).

ZAG 135 dsDNA kitdsDNA 935 Reagent kitZAG 110 dsDNA kitdsDNA 915 Reagent kit
Sizing Range100-1,500 bp100-1,500 bp35-5,000 bp35-5,000 bp
Sizing Accuracy5%5%5%5%
Sizing Precision2%2%2%2%
Sizing Resolution100-1,500 bp ≤ 10%100-1,500 bp ≤ 10%35-100 bp ≤ 10%35-100 bp ≤ 10%
100-1,000 bp ≤ 5%100-1,000 bp ≤ 5%
1,000-5,000 bp ≤ 10%1,000-5,000 bp ≤ 10%
Concentration Range0.5-50 ng/µL0.5-50 ng/µL0.5-50 ng/µL0.5-50 ng/µL

Table 2. Sizing of single fragments and mixes of fragments on the Agilent ZAG DNA Analyzer and Fragment Analyzer systems with the ZAG 135 dsDNA kit (1–5000 bp) and dsDNA 935 Reagent kit (1–5000 bp), respectively. All fragments are within the sizing accuracy specifications of the kits. UM = upper marker, *n = 6.

Theoretical Fragment Size (bp)ZAG 135 kit (bp)Fragment Analyzer 935 kit (bp)ZAG 110 kit (bp)Fragment Analyzer 915 kit (bp)
306--302305
311--308310
Mix 306311---
-305309305
309400406400
398400425435429
428428Mix 400425403
431400428398427
3984281,000968972
9839781,0501,0181,015
1,0261,0221,1001,0661,062
1,0991,095Mix 1,0001,0501,100
9751,015-9701,013
-9861,0291,099981
1,0211,091Mix 1,0001,100976
1,0779731,0669851,100
9781,0951,3501,3561,347
1,2971,2891,4251,4221,415
1,3931,3891,500UMUM
1,4991491Mix 1,3501,4251,500
1,3601,414UM1,3621,410
UM1,3061,3901,5021,295
1,3821,492Mix 3,0004,000-
--2,9434,0222,932
3,9114,500--4,225
4,131

UM = upper marker, *n = 6.

Engineering Facts

12 facts · every claim linked to a page
Selection

Choose the kit based on the required sizing range: 1–1500 bp for ZAG 135/dsDNA 935 or 35–5000 bp for ZAG 110/dsDNA 915.

Defines which kit covers the sample fragment sizes and prevents out-of-range sizing errors.

Source
“The ZAG 135 and Fragment Analyzer 935 kits were specifically designed for accurate sizing of fragments less than 1,500 bp, while the ZAG 110 and Fragment Analyzer 915 kits size up to 5,000 bp.”
Selection

Select capillary array length according to the resolution versus run-time trade-off: long for best resolution, ultrashort for fastest runs, short for a balance.

Array length directly determines achievable separation of closely sized fragments and throughput.

Source
“The long array will provide the best resolution with a longer run time, while the ultrashort array has the shortest run times, but compromises on resolution.”
Selection

The ZAG accommodates nine plates for qualitative DNA kits and is intended for ultrahigh-throughput laboratories, while the Fragment Analyzer holds three plates and supports both quantitative and qualitative DNA/RNA kits.

Plate capacity and kit type determine instrument fit for sample volume and assay portfolio.

Source
“Alternatively, the ZAG accommodates nine plates for qualitative DNA analysis kits, with faster run times and is ideal for laboratories requiring ultrahigh sample throughput6.”
Performance

The 10% resolution specification means that at 1,000 bp a ±100 bp difference is required for complete separation; a 5% difference gives only limited separation.

Sets realistic expectations for distinguishing closely sized fragments near 1,000 bp.

Source
“Resolution for both kits is 10%, with resolution at 1,000 bp requiring a ±100 bp difference.”
Performance

At the smaller base-pair end of the ZAG 135/dsDNA 935 kits, better resolution than the specified 10% can be achieved.

Allows tighter fragment discrimination at low bp than the nominal specification suggests.

Source
“At the smaller base pair end of the kits, better resolution can be achieved than the 10% resolution specified in the kits.”
Operation

When peaks are only partially or limitedly separated, the ProSize software may not select them as separate peaks and manual peak addition/splitting is needed to assign sizes.

Operators must know the manual peak-editing workflow to obtain sizes for unresolved mixtures.

Source
“ProSize software allows the user to “add peak or split peak” in order for a size to be assigned to both fragments.”
Operation

The upper marker is a 1,500 bp fragment used for sizing and alignment of ladder and samples on the ZAG 135 and dsDNA 935 kits.

Understanding marker identity is essential for correct sizing and alignment interpretation.

Source
“The upper marker (UM) is a 1,500 bp fragment used for sizing and alignment of the ladder and samples on both the ZAG 135 dsDNA and the Fragment Analyzer 935 kits.”
Performance

The wide concentration range of the kits often eliminates a dilution step and enables direct processing of samples with unknown concentrations.

Reduces sample prep time and avoids dilution errors for unknown-concentration samples.

Source
“The wide concentration range offered by both kits often eliminates a dilution step and enables direct processing of samples with unknown concentrations, saving time in the workflow.”
Selection

The ZAG 110/dsDNA 915 kits have a lower percent resolution (5%) in the 100–1,000 bp range than the 135/935 kits, allowing resolution requirements to be matched to the application.

Guides kit choice when better resolution in the 100–1,000 bp range is required.

Source
“While there is an overlap in the sizing range between the kits, the percent resolution is lower at the 100– 1,000 bp range of the ZAG 110 dsDNA and Fragment Analyzer 915 Reagent kits, allowing for the resolution requirements of the user to be met.”
Performance

A 10% size difference from the 5,000 bp upper marker meets the 10% resolution specification for complete separation; partial separation is possible only with fragments larger than 4,500 bp.

Defines the resolvable limit near the top of the sizing range for the 110/915 kits.

Source
“The 10% size difference from the upper marker meets both kit specifications of 10% resolution for complete separation.”
Operation

The 1,425 bp fragment had limited separation from the 1,500 bp upper marker and required manual peak selection in ProSize to assign an accurate size.

Alerts operators that fragments close to the upper marker may need manual peak selection.

Source
“However, the 1,425 bp fragment had limited separation from the upper marker and required manual selection of the peak in the ProSize software, allowing for an accurate size to be assigned”
Performance

The ZAG 110 and Fragment Analyzer 915 kits can provide partial and limited separation with accurate sizing below the specified 5% resolution and complete separation at or above 5%.

Clarifies that accurate sizing may still be possible even when peaks are not fully resolved.

Source
“The ZAG 110 and Fragment Analyzer 915 kits can provide partial and limited separation with accurate sizing below the specified 5% resolution and complete separation at or above the specified resolution of 5%.”