mo304
Technical Reference Summary
Covered Models & Document Scope
Espectrómetro de masas MALDI AXIMA Assurance; configuración estándar con láser de N2 de 50 Hz, modos de iones positivos y negativos, beam blanker patentado; formatos de portaobjetos: placa de microtitulación de 96 o 384 pocillos, portaobjetos de microscopio FlexiMass (lisos o de 48 pocillos), adaptadores para disposiciones no convencionales
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Ion source | variable repetition rate 50Hz N2 laser p.2 ↗ | |
| Ion modes | positive and negative p.2 ↗ | standard |
| Resolution | >6000 FWHM p.2 ↗ | linear mode, ACTH 18-39 |
| Mass range | 2000-12000 m/z p.2 ↗ | typical oligonucleotide spectrum |
| Sample target formats | 96 or 384 well microtitre plate; FlexiMass microscope slides (plain or 48 well) p.2 ↗ | |
| Beam blanker | patented, optionally removes unwanted low mass ions p.2 ↗ | |
| Ion optics | near-axis laser irradiation and advanced ion optics p.2 ↗ | |
| Extraction | pulsed extraction p.2 ↗ | |
| Software | Launchpad suite; Polymer Analysis p.3 ↗ | |
| Operation | manual or fully automated p.3 ↗ |