Induztr
Résumé technique Visionneuse de documents Modèles associés

Hydrogen Carrier Gas Safety - GCMS-QP2010 Series / GCMS-QP2020 / GCMS-TQ Series

Technical Reference Summary

Covered Models & Document Scope

Manual de seguridad 225-35068A (jun. 2016) que cubre la seguridad del gas portador de hidrógeno para todos los modelos de las series GCMS-QP2010, GCMS-QP2020 y GCMS-TQ.

Technical Data

Parameter Value Condition
Hydrogen combustion range in air 4 to 75 vol% p.3 ↗
Maximum ambient temperature for high-pressure gas cylinder location 40 °C p.7 ↗ high-pressure gas cylinder setup
Pressure gauge inspection interval once every three months p.8 ↗
Hydrogen expulsion wait time after power outage approximately 30 p.12 ↗ after power outage

Operating Limits

  • Hydrogen combustible concentration range in air: 4 vol% to 75 vol% p.3 ↗
  • Max temperature at high-pressure gas cylinder location: 40 °C p.7 ↗
  • Pressure gauge inspection frequency: at least once every three months p.8 ↗
  • Wait time for complete hydrogen expulsion after power outage: approximately 30 (unit not printed) p.12 ↗

Modèles liés