Glycerophospholipids Analysis by Comprehensive HPLC Coupled with a Triple Quadrupole Mass Spectrometer

Technical Reference Summary

Covered Models & Document Scope

Nota de aplicación n.º L462 (LAAN-A-LC-E239) que describe un método completo de HPLC-ESI-MS/MS para el análisis de glicerofosfolípidos; abarca las condiciones analíticas para la separación 1D en fase normal y 2D en fase inversa con detección Shimadzu LCMS-8050. No se definen modelos de producto, tamaños ni códigos de pedido.

Technical Data

Parameter Value Condition
1D column Nucleosil SIL (150 mm L. × 1.0 mm I.D., 3 µm) p.1 ↗ 1D separation
1D mobile phase A Isooctane / Acetone / Ethyl Acetate / Acetic acid = 40/20/20/0.03 v/v/v/v p.1 ↗ 1D separation
1D mobile phase B Isooctane / 2-propanol / Water / Acetic acid / 28 % Ammonia aq.sol. = 40/51/9/0.03/0.03 v/v/v/v/v p.1 ↗ 1D separation
1D flow rate 0.02 mL/min p.1 ↗ 1D separation
1D time program B Conc. 30 % (0 min) → 40 % (25 min) → 100 % (40 min) → 100 % (55 min) → 30 % (55.1 min) → STOP (70 min) p.1 ↗ 1D separation
Column temperature 40 °C p.1 ↗ 1D separation
Injection volume 5 µL p.1 ↗
Loop volume 20 µL p.1 ↗
2D column Phenomenex Kinetex C18 (50 mm L. × 4.6 mm I.D., 2.6 µm) p.1 ↗ 2D separation
2D mobile phase A Methanol / Water / Acetic acid / 28 % Ammonia aq.sol. = 90/10/0.05/0.05 v/v/v/v p.1 ↗ 2D separation
2D mobile phase B 2-propanol / Acetic acid / 28 % Ammonium hydroxide = 100/0.05/0.05 V/V/V p.1 ↗ 2D separation
2D flow rate 3.5 (50 % split to MS) mL/min p.1 ↗ 2D separation
2D time program B Conc. 10 % (0 min) → 50 % (0.75 min) → 10 % (0.76 min) → STOP (1 min) p.1 ↗ 2D separation
Detector Shimadzu LCMS-8050 (ESI positive, MRM mode) p.1 ↗
MRM transition 16:0-16:0 PC 734.6 > 184.1 m/z p.2 ↗
MRM transition 18:0-18:1 PC 788.6 > 184.1 m/z p.2 ↗
MRM transition 18:0p-18:1 PC 772.6 > 184.1 m/z p.2 ↗
Total retention time 16:0-16:0 PC 0.0072 min p.2 ↗
Total retention time 18:0-18:1 PC 0.013 min p.2 ↗
Total retention time 18:0p-18:1 PC 0.013 min p.2 ↗
Correlation coefficient (R) 16:0-16:0 PC 0.999799 p.2 ↗
Correlation coefficient (R) 18:0-18:1 PC 0.999947 p.2 ↗
Correlation coefficient (R) 18:0p-18:1 PC 0.999656 p.2 ↗

Operating Limits

  • Column temperature: 40 °C p.1 ↗
  • 1D flow rate: 0.02 mL/min p.1 ↗
  • 2D flow rate: 3.5 mL/min (50 % split to MS) p.1 ↗
  • Injection volume: 5 µL p.1 ↗
  • Loop volume: 20 µL p.1 ↗