High-Sensitivity Analysis of Aldosterone in Low-Volume Serum Samples using Micro-Flow LC-MS/MS for Clinical Research
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación / folleto de método para el análisis de alta sensibilidad de aldosterona en muestras de suero de bajo volumen utilizando un sistema LC-MS/MS de microflujo Shimadzu Nexera Mikros (LC-Mikros, LC-30AD, SIL-30AC, FCV-32AH, CTO-Mikros) acoplado a un espectrómetro de masas de triple cuadrupolo LCMS-8060 con fuente micro-ESI, en configuración trap-and-elute. Research Use Only; no para procedimientos de diagnóstico; no disponible en China.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Mass spectrometer system | LCMS-8060 p.1 ↗ | |
| Ionization | Micro-ESI p.1 ↗ | negative ionization |
| Probe voltage | -2.4 kV p.1 ↗ | negative ionization |
| Interface temperature | 125 °C p.1 ↗ | |
| Desolvation line temperature | 250 °C p.1 ↗ | |
| Heater block temperature | 300 °C p.1 ↗ | |
| Nebulizing gas flow | 3 L/min p.1 ↗ | |
| Heating gas flow | 3 L/min p.1 ↗ | |
| Drying gas | Off p.1 ↗ | |
| Dwell time | 115 ms p.1 ↗ | |
| MRM transition (quantifier), aldosterone | 359.20 > 189.30 m/z p.1 ↗ | negative |
| MRM transition (qualifier), aldosterone | 359.20 > 331.35 m/z p.1 ↗ | negative |
| MRM transition (quantifier), d4-aldosterone | 363.00 > 190.20 m/z p.1 ↗ | |
| Trapping column | CERI C8 5 µm 5*0.3 mm p.1 ↗ | |
| Trapping flow rate | 100 µL/min p.1 ↗ | |
| Microbore column | Shimadzu PLONAS C18 2.7 µm 100*0.2 mm p.1 ↗ | |
| Analytical flow rate | 4 µL/min p.1 ↗ | |
| Serum sample volume | 10 µL p.1 ↗ | |
| Calibration range | 1 to 2 500 pg/mL p.1 ↗ | |
| Lower limit of quantitation (LOQ) | 1 pg/mL p.1 ↗ | QC, n=3 replicates |
| Intra-day precision (%RSD) | < 8 % p.1 ↗ | real samples |
| LOQ accuracy | 103 (mean), 94.4 (min), 114 (max) % p.1 ↗ | QC LOQ 1 pg/mL |
| LOQ precision (%RSD) | 5.2 intra-day, 4.9 inter-day % p.1 ↗ | QC LOQ 1 pg/mL |
| Matrix effect recovery | 99 (QC), 103 (solution) % p.1 ↗ |
Operating Limits
- Lower limit of quantitation (LOQ): 1 pg/mL p.1 ↗
- Calibration range: 1 to 2 500 pg/mL p.1 ↗
- Minimum serum sample volume: 10 µL p.1 ↗
- Organic solvent addition to plasma: max 0.5% (v/v) p.1 ↗
- Intra-day precision (%RSD) for real samples: < 8% p.1 ↗
- LOQ acceptance criteria: accuracy 80-120%, %RSD < 20% p.1 ↗