Induztr
Résumé technique Visionneuse de documents Modèles associés

Instruction Manual - Nexera UC Piping Kit -

Technical Reference Summary

Covered Models & Document Scope

Nexera UC System Piping Kits (document 228-92325D). Covers all piping kit variants identified by symbol: B (Basic, 228-61492-41), M (MS, 228-61493-41), O (Online, 228-61494-41), E (SFE, 228-61496-41), C (Column Switching 6, 228-61105-43 for FCV-34AH / -45 for FCV-0607), P (Column Switching +6, 228-61105-44 for FCV-34AH / -46 for FCV-0607), H (High pressure gas, 228-61497-41), m (MS for I.D. 0.1 mm, 228-63782-41), c (Column Switching 6 for I.D. 0.1 mm, 228-63783-41 for FCV-0607 / -43 for FCV-34AH), p (Column Switching +6 for I.D. 0.1 mm, 228-63783-42 for FCV-0607 / -44 for FCV-34AH), F (Fraction Collector Modification Kit for Prep SFC, 228-75753-42), f (Fraction Collector Modification Kit for Analytical SFC, 228-79729-41), A (Analytical Fraction MS, 228-68817-41), L (LC/SFC Switching, 228-62767-41), D (Direct/Loop Injection Switching, 228-75296-42 for FCV-32AH / -43 for FCV-0206).

Section-by-Section Technical Breakdown

pp. 1, 13
Piping Kit Selection and System Assignment

The manual assigns each piping kit symbol to a specific Nexera UC system configuration. The I.D. 0.1 mm kits are additional kits that replace the standard I.D. 0.3 mm piping to reduce system dead volume. The symbol in the table corresponds to the symbol in the piping diagram where each kit is used.

  • B is the common basic piping kit for every system.
  • M is required when using MS; O is for the Online SFE-SFC system; E is for SFE.
  • C and P are for 6- or 12-column switching and are ordered by valve type (FCV-34AH vs FCV-0607).
  • m, c and p are I.D. 0.1 mm additional kits that replace standard I.D. 0.3 mm piping to reduce dead volume.
  • F and f modify the FRC-40 fraction collector for Prep SFC and Analytical SFC respectively.
  • A is used for MS-triggered or ELSD-triggered fractionation in the Analytical Fraction system.
  • L and D kits have their own attached instruction manuals.
pp. 2, 14
Piping Kit, Basic (228-61492-41) Components

Lists the 17 line items of the Basic piping kit with part numbers, quantities and installation locations. The kit includes SUS and ETFE tubing, ferrules, male nuts, couplings, labels, wiring bands and a SIL-40 loop-injection port pipe.

  • B1 SUS Tubing (1.0x100) 228-61552-41 connects line filter to ghost trap.
  • B2 SUS Tubing (1.0x2000) 228-61552-42 connects ghost trap to CO2 pump.
  • B3 SUS Sleeve Tubing (0.3x1200) 228-53184-41, qty 2, for modifier pump–mixer and CO2 pump–mixer.
  • B4 SUS Sleeve Tubing (0.3x700) 228-53184-93 for T-joint–column; also mixer–high pressure valve on Chiral Screening System.
  • B5 SUS Tubing (0.3x175) 228-57292-45 for BPR (1)–coupling.
  • B6 ETFE Tubing (1.6x1.0x2000) 228-18495-03 for drain tubing (Chiral Screening System only).
  • B7 Ferrule 1.6F 316L 228-16000-10 qty 13; B8 Male Nut 1.6MN 228-16001 qty 11; B9 Male Nut 1.6MN,W6 228-16001-03 qty 2; B10 Male Nut 1.6MN PEEK 228-35403 qty 2; B11 Coupling 1.6C-0.4 316L 228-16004-13 qty 2.
  • B12 Label, Right Panel 228-61147 qty 11; B13 Wiring Band IT18FL-B 072-60355-07 qty 70; B14 Label, Identification Tag 228-61513 qty 1.
  • B15 SUS Sleeve Tubing (0.3x300) 228-53184-92 for Cell Out–BPR (1).
  • B16 Injection port pipe, I.D. Φ0.17 228-64106-49 for SIL-40 Loop Injection.
  • B17 Male Nut 1.6MN-L 228-16001-01 for BPR (1) IN.
pp. 2, 14
Piping Kit, MS (228-61493-41) Components

Lists the 8 line items of the MS piping kit. It provides the makeup-pump-to-T-joint line, the MS connection tube, a check valve replacing the makeup pump line filter, and the T-joint and fittings.

  • M1 SUS Sleeve Tubing (0.1x200) 228-53184-45 qty 2 for column–T-joint and T-joint–BPR (1).
  • M2 SUS Sleeve Tubing (0.3x1200) 228-53184-41 for makeup pump–T-joint.
  • M3 PEEK TUBE 5999 (0.13x400) 228-32999-01 or 016-42903-01 qty 1 for MS connection.
  • M4 Check Valve 228-53334-96 replaces the line filter of the makeup pump.
  • M5 T-joint 0.4 228-16008-11 qty 1.
  • M6 Ferrule 1.6F 316L 228-16000-10 qty 6; M7 Male Nut 1.6MN 228-16001 qty 5; M8 Male Nut 1.6MN-L 228-16001-01 qty 1.
pp. 2, 14
Piping Kit, Online (228-61494-41) Components

Lists the 11 line items of the Online SFE-SFC piping kit. It includes the SIL HPV–T-joint and T-joint–BPR (2) lines, a T-joint, fittings, ETFE drain tubing and PEEK male nuts.

  • O1 SUS Sleeve Tubing (0.3x300) 228-53184-92 for SIL HPV–T-joint.
  • O2 SUS Sleeve Tubing (0.3x700) 228-53184-93 for T-joint–BPR (2).
  • O3 T-joint 0.4 228-16008-11 qty 1.
  • O4 Ferrule 1.6F 316L 228-16000-10 qty 5; O5 Male Nut 1.6MN 228-16001 qty 3; O6 Male Nut 1.6MN-L 228-16001-01 qty 1; O7 Male Nut 1.6MN,W6 228-16001-03 qty 1.
  • O8 SUS Tubing ID0.3xL175 228-57292-45 for BPR (2)–coupling.
  • O9 Coupling 1.6C-0.4 316L 228-16004-13 qty 1.
  • O10 ETFE Tubing OD1.6xID1.0xL2000 228-18495-03 qty 1.
  • O11 Male Nut 1.6MN PEEK 228-35403 qty 2.
pp. 3, 15
Piping Kit, SFE (228-61496-41) Components

Lists the 6 line items of the SFE piping kit. It provides mixer–SFE and SFE–SIL tubing, fittings, and a cushion plus M4x8 bind screw for installing the mixer in the SFE pretreatment system.

  • E1 SUS Sleeve Tubing (0.3x1200) 228-53184-41 qty 2 for mixer–SFE and SFE–SIL.
  • E2 Ferrule 1.6F 316L 228-16000-10 qty 4; E3 Male Nut 1.6MN 228-16001 qty 2; E4 Male Nut 1.6MN,W6 228-16001-03 qty 3.
  • E5 Cushion qty 1 is inserted between the mixer mounting plate and the side of the pump as a protective material.
  • E6 Bind Screw M4x8 qty 1 fastens one point on the side of the pump for mixer installation.
pp. 3, 15
Piping Kit, Column Switching 6 (228-61105-43, -45) Components

Lists the components for 6- or 12-column switching. The -43 variant is for FCV-34AH and the -45 variant is for FCV-0607; quantities differ between variants. The kit includes the MSS2 CD ASSY, manifold, column clamps, color identification tags and plugs.

  • MSS2 CD ASSY 228-64206-41 qty 1 is used to install Method Scouting System.
  • C1 SUS Sleeve Tubing (0.3x300) 228-53184-92 qty 12 for 6 positions FCV–column (6 pc) and column–manifold (6 pc).
  • C2 Ferrule 1.6F 316L 228-16000-10 qty 18 (-43) or 24 (-45); C3 Male Nut 1.6MN 228-16001 qty 18; C6 Male Nut 1.6MN,W6 228-16001-03 qty 6 only for -45.
  • C8 Manifold ASSY,6IN-1OUT 228-54450-91 qty 1.
  • C9 Color Identification Tag Column ASSY 228-54728-92 qty 1; C10 Color Identification Tag MP ASSY 228-54728-93 qty 1.
  • C11 Column Clamp ASSY B5 228-15617-91 qty 4 for -43; Column Clamp 228-72586 qty 4 for -45.
  • C16 Plug 228-16006-10 qty 6.
pp. 4, 15
Piping Kit, Column Switching +6 (228-61105-44, -46) Components

Lists the components for 12-column switching. The -44 variant is for FCV-34AH and the -46 variant is for FCV-0607; quantities differ. The kit adds a second manifold, a T-joint, additional tubing and color tags.

  • P1 SUS Sleeve Tubing (0.3x300) 228-53184-92 qty 15 for 2 positions FCV–6 positions FCV (1), 6 positions FCV (2)–column (6 pc), column–manifold (6 pc), manifold (1)–T-joint and manifold (2)–T-joint.
  • P3 T-joint 0.4 228-16008-11 qty 1.
  • P4 Ferrule 1.6F 316L 228-16000-10 qty 26 (-44) or 34 (-46); P5 Male Nut 1.6MN 228-16001 qty 23 (-44) or 24 (-46); P6 Male Nut 1.6MN,W6 228-16001-03 qty 2 (-44) or 10 (-46).
  • P7 Coupling 1.6C-0.4 316L 228-16004-13 qty 1; P8 Manifold ASSY,6IN-1OUT 228-54450-91 qty 1.
  • P9 Color Identification Tag Column ASSY 228-54728-92 qty 1; P10 Color Identification Tag Column ASSY 228-54728-94 qty 1; P11 Color Identification Tag MP ASSY 228-54728-93 qty 1.
  • P12 Column Clamp ASSY B5 228-15617-91 qty 4 for -44; Column Clamp 228-72586 qty 4 for -46.
  • P13 Male Nut 1.6MN-L 228-16001-01 qty 1; P14 SUS Sleeve Tubing (0.1x200) 228-53184-45 qty 1 for SIL–2 positions FCV; P15 SUS Sleeve Tubing (0.3x700) 228-53184-93 qty 1 for 2 positions FCV–6 positions FCV (2); P16 Plug 228-16006-10 qty 6.
pp. 4, 16
Piping Kit, for High pressure gas (228-61497-41) Components

Lists the 6 line items of the high-pressure gas piping kit. It includes a cylinder joint, PTFE gaskets, a line filter, a ghost trap for trapping foreign objects from cylinders, a check valve and an MR mixer.

  • H1 Cylinder Joint 228-61086 qty 1 for connection to cylinder.
  • H2 Gasket (PTFE) 228-61087-10 qty 3.
  • H3 Line Filter 228-58173-42 qty 1.
  • H4 Ghost trap 228-61088-41 qty 1 used to trap foreign objects from cylinders.
  • H5 Check valve 228-53334-96 qty 1; replace it with a line filter.
  • H6 MR mixer, 180UL SFC 228-61089-41 qty 1.
pp. 4, 16
Piping Kit, MS for I.D.0.1 mm (228-63782-41) Components

Lists the 6 line items of the I.D. 0.1 mm additional MS piping kit. It provides BPR IN and BPR OUT tubing plus fittings for the reduced-dead-volume MS flow path.

  • m1 SUS Tubing, BPR IN 228-57271-41 qty 1 for BPR (1).
  • m2 SUS Tubing, BPR OUT 228-57292-41 qty 1 for BPR (1)–coupling.
  • m3 Ferrule 1.6F 316L 228-16000-10 qty 3; m4 Male Nut 1.6MN 228-16001 qty 2; m5 Male Nut 1.6MN-L 228-16001-01 qty 1; m6 Male Nut 1.6MN PEEK 228-35403 qty 1.
pp. 5, 16
Piping Kit, Column Switching 6 for I.D.0.1 mm (228-63783-41, -43) Components

Lists the components of the I.D. 0.1 mm additional kit for 6- or 12-column switching. The -41 variant is for FCV-0607 and the -43 variant is for FCV-34AH; the -43 variant adds ferrules and bushes for FCV-34AH connections.

  • c1 SUS Sleeve Tubing (0.1x200) 228-53137-45 qty 12 for 6 positions FCV–column (6 pc) and column–manifold (6 pc).
  • c3 Ferrule 1.6F 316L 228-16000-10 qty 24; c4 Male Nut 1.6MN 228-16001 qty 18; c5 Male Nut 1.6MN,W6 228-16001-03 qty 6.
  • c6 FERRULE, 7010-010 670-11412-04 qty 7 only for -43, used for FCV-34AH piping connection.
  • c7 BUSH, 7010-011 670-11412-02 qty 6 only for -43; c8 BUSH, 7010-062 670-11412-03 qty 1 only for -43.
pp. 5, 16
Piping Kit, Column Switching +6 for I.D.0.1 mm (228-63783-42, -44) Components

Lists the components of the I.D. 0.1 mm additional kit for 12-column switching. The -42 variant is for FCV-0607 and the -44 variant is for FCV-34AH; the -44 variant adds ferrules and bushes for FCV-34AH connections.

  • p1 SUS Sleeve Tubing (0.1x200) 228-53137-45 qty 15 for 2 positions FCV–6 positions FCV (1), 6 positions FCV (2)–column (6 pc), column–manifold (2) (6 pc), manifold (1)–T-joint and manifold (2)–T-joint.
  • p2 SUS Sleeve Tubing (0.1x600) 228-53137-94 qty 1 for 2 positions FCV–6 positions FCV (2).
  • p3 Ferrule 1.6F 316L 228-16000-10 qty 32; p4 Male Nut 1.6MN 228-16001 qty 22; p5 Male Nut 1.6MN,W6 228-16001-03 qty 10.
  • p6 FERRULE, 7010-010 670-11412-04 qty 7 only for -44, used for FCV-34AH piping connection.
  • p7 BUSH, 7010-011 670-11412-02 qty 6 only for -44; p8 BUSH, 7010-062 670-11412-03 qty 1 only for -44.
pp. 6, 17
Fraction Collector Modification Kit for Prep SFC (228-75753-42) Components

Lists the components that modify the FRC-40 for Prep SFC. It includes a check valve with defined flow orientation, a line filter joint, FCV-to-GLS tubing, stop plugs and fittings. Items X1–X3 are not included in the kit and must be purchased additionally.

  • F2 CV,OUT-30ADSF 228-45705-41: valve joint side is upstream and check valve side is downstream.
  • F3 SUS Joint, for Line Filter, Nexera UC 228-50707-01 qty 1.
  • F4 SUS Sleeve Tubing (0.1x600) 228-53137-96 for T-joint–FCV (L), port 2.
  • F5 PIPE (1.6x1.0x300 ORANGE) 228-68910-61 qty 2 for T-joint–FCV (R), port 1; only 1 is used, keep the other as spare.
  • F6 PEEK TUBE / SUS TUBE 0.5 mm I.D. 228-77523-42 (included in GLS kit) for FCV (R), port 2–GLS; F7 ETFE TUBE 0.5 mm I.D. for FCV (L), port 1–GLS.
  • F8 PLUG 228-16006-10 qty 3: use stop plugs on port 3 of FCV (R) and ports 3 and 6 of FCV (L); if not closed, leakage may occur depending on valve position.
  • F9 MALE NUT 1.6MN 228-16001 qty 6; F10 FERRULE 1.6F 316L 228-16000-10 qty 6.
  • X1 SUS Sleeve Tubing (0.3x300) 228-53137-42 qty 2, X2 SUS Sleeve Tubing (0.3x1200) 228-53184-41 qty 1 and X3 FEMALE,1.6FT316L-0.4 228-16008-11 qty 1 are not included in this kit; use additionally purchased piping.
pp. 6, 17
Fraction Collector Modification Kit for Analytical SFC (228-79729-41) Components

Lists the components that modify the FRC-40 for Analytical SFC. It includes makeup-pump tubing, a check valve with defined flow orientation, a line filter joint, PEEK SIL tubing, ETFE tubing for main and rinse flow, and stop plugs.

  • f1 SUS Tubing (1.6x0.3x2000) 228-50579-91 for FRC makeup pump–T-joint.
  • f4 CV,OUT-30ADSF 228-45705-41: valve joint side is upstream and check valve side is downstream.
  • f5 SUS Joint, for Line Filter, Nexera UC 228-50707-01 qty 1.
  • f6 PEEK SIL TUBE (0.05x100) 225-42910-30 for T-joint–coupling.
  • f7 SUS Sleeve Tubing (0.3x300) 228-53137-42 qty 3 for T-joint–valve joint, check valve–T-joint and T-joint–FCV port 2.
  • f8 ETFE TUBE 0.3 mm I.D. 228-79342-43 (included in GLS kit) for T-joint–solenoid valve (main flow); f9 ETFE TUBE 0.3 mm I.D. with 'Rinse' tag for FCV port 1–GLS (rinse flow); f10 ETFE TUBE 1.0 mm I.D. for solenoid valve–drain.
  • f11 PLUG 228-16006-10 qty 1: use stop plug on port 3 of FCV; if the valve is not closed, leakage may occur.
  • f12 MALE NUT 1.6MN 228-16001 qty 10; f13 FERRULE 1.6F 316L 228-16000-10 qty 10.
pp. 7, 18
Piping Kit, Analytical Fraction MS (228-68817-41) Components

Lists the components for MS-triggered or ELSD-triggered fractionation in the Analytical Fraction system. It includes delay tubing for capacity adjustment, PEEK SIL tubing cut to length for MS connection, a backflow-prevention check valve and drain tubing.

  • A1 SUS Sleeve Tubing (0.3x300) 228-53137-42, I.D. 0.3 mm × 300 mm.
  • A2 PEEK SIL TUBE 018-30510-51, I.D. 0.025 mm × 500 mm; 2 tubes included but only 1 is used, keep the other as spare.
  • A3 SUS316LTP (1.6x0.3x2000) 228-50579-91 qty 5, I.D. 0.3 mm × 2000 mm: 1 pc for LC-40D–T-joint and 4 pc for Cell–SFC-30A; use 1 pc for MS makeup pump outlet piping and 4 pc for delay capacity adjustment piping.
  • A4 PEEK TUBE (0.13x3000) 228-33833-91, I.D. 0.13 mm × 3000 mm: cut as needed to reach MS; cut 200 mm for T-joint–LCMS and 164 mm for T-joint–coupling.
  • A5 FEMALE,1.6FT316L-0.4 228-16008-11 qty 3; A6 Coupling 1.6C-0.4 316L 228-16004-13 qty 5.
  • A7 Drain tube 228-25495-93 qty 1; when using ELSD, connect with PEEK tube of 0.25 mm I.D. instead of the drain tube (sold separately, 228-33833-91 PEEK TUBE 0.25×3000).
  • A8 Ferrule 1.6F 316L 228-16000-10 qty 16; A9 Male Nut 1.6MN 228-16001 qty 16; A10 Male Nut 1.6MN PEEK 228-35403 qty 5 for both ends of A4*2 and A4*3 and No. 6 connections.
  • A11 CV,OUT-UHP TP 228-53334-96 qty 1 is a check valve to prevent backflow; A12 Pipe bands qty 3 for bundling delay pipe (A3).
pp. 8-12
Piping Examples by System Configuration

Provides piping diagrams showing where each kit symbol is used in the Online SFE-SFC-MS, Analytical Fraction, Analytical Fraction MS, SFC-MS, SFE pretreatment (FRC-40 SF and FRC-10A), Chiral Screening, Column Switching 6/+6 for I.D.0.1 mm, and SFC-UV/PDA systems. The manual refers to the Nexera UC System Guide section 2.1.3 for analytical flow line tubing examples.

  • Online SFE-SFC-MS System diagram.
  • Analytical Fraction System and Analytical Fraction MS System diagrams.
  • SFC-MS System and Piping Kit MS for I.D.0.1 mm diagram.
  • SFE pretreatment system diagrams for FRC-40 SF and FRC-10A.
  • Chiral Screening System diagram.
  • Piping Kit Column Switching 6/+6 for I.D.0.1 mm diagram.
  • SFC-UV/PDA System diagram.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication S'applique à
The I.D. 0.1 mm piping kits (m, c, p) replace the standard I.D. 0.3 mm piping to reduce system dead volume. p.1 ↗ Select an I.D. 0.1 mm kit when reduced dead volume is required; otherwise the standard I.D. 0.3 mm kit applies. m (228-63782-41), c (228-63783-41/-43), p (228-63783-42/-44)
Column Switching 6 kit 228-61105-43 is for FCV-34AH and 228-61105-45 is for FCV-0607. p.1 ↗ Order the kit variant matching the installed FCV valve model. C kit
Column Switching +6 kit 228-61105-44 is for FCV-34AH and 228-61105-46 is for FCV-0607. p.1 ↗ Order the kit variant matching the installed FCV valve model. P kit
I.D. 0.1 mm Column Switching 6 kit 228-63783-41 is for FCV-0607 and 228-63783-43 is for FCV-34AH. p.1 ↗ The -43 variant includes ferrules 670-11412-04 (qty 7) and bushes 670-11412-02 (qty 6) and 670-11412-03 (qty 1) for FCV-34AH connections. c kit
I.D. 0.1 mm Column Switching +6 kit 228-63783-42 is for FCV-0607 and 228-63783-44 is for FCV-34AH. p.1 ↗ The -44 variant includes ferrules 670-11412-04 (qty 7) and bushes 670-11412-02 (qty 6) and 670-11412-03 (qty 1) for FCV-34AH connections. p kit
Basic kit B16 Injection port pipe has I.D. Φ0.17 mm (228-64106-49) and is used for SIL-40 Loop Injection. p.2 ↗ Use this specific injection port pipe when configuring SIL-40 loop injection. B kit
MS kit M4 Check Valve 228-53334-96 replaces the line filter of the makeup pump. p.2 ↗ Remove the makeup pump line filter and install the check valve in its place when using the MS kit. M kit
High pressure gas kit H5 Check valve 228-53334-96 is to be replaced with a line filter. p.4 ↗ In the high-pressure gas kit, the check valve position is replaced by a line filter. H kit
SFE kit E5 Cushion is inserted between the mixer mounting plate and the side of the pump as a protective material; E6 Bind Screw M4x8 fastens one point on the side of the pump. p.3 ↗ Follow this mounting arrangement when installing the mixer in the SFE pretreatment system. E kit
Prep SFC modification kit F2 CV,OUT-30ADSF 228-45705-41 must be oriented with the valve joint side upstream and the check valve side downstream. p.6 ↗ Incorrect orientation will prevent proper check-valve function. F kit (228-75753-42)
Analytical SFC modification kit f4 CV,OUT-30ADSF 228-45705-41 must be oriented with the valve joint side upstream and the check valve side downstream. p.6 ↗ Incorrect orientation will prevent proper check-valve function. f kit (228-79729-41)
Prep SFC kit F8 PLUG 228-16006-10 qty 3: use stop plugs on port 3 of FCV (R) and ports 3 and 6 of FCV (L); if the valve is not closed with a stop plug, leakage may occur depending on the valve position. p.6 ↗ Always plug the specified unused FCV ports to prevent leakage. F kit (228-75753-42)
Analytical SFC kit f11 PLUG 228-16006-10 qty 1: use stop plug on port 3 of FCV; if the valve is not closed, leakage may occur. p.6 ↗ Always plug the specified unused FCV port to prevent leakage. f kit (228-79729-41)
Analytical Fraction MS kit A3 SUS316LTP (1.6x0.3x2000) 228-50579-91 qty 5: use 1 pc for MS makeup pump outlet piping and 4 pc for delay capacity adjustment piping (LC-40D–T-joint 1 pc; Cell–SFC-30A 4 pc). p.7 ↗ Allocate the five tubes exactly as specified to achieve the intended delay volume. A kit (228-68817-41)
Analytical Fraction MS kit A4 PEEK TUBE (0.13x3000) 228-33833-91: cut 200 mm for T-joint–LCMS and 164 mm for T-joint–coupling; cut as much as necessary to reach MS for T-joint–T-joint. p.7 ↗ Use these cut lengths to maintain the intended flow path geometry. A kit (228-68817-41)
When using ELSD with the Analytical Fraction MS kit, connect with a PEEK tube of 0.25 mm I.D. instead of the A7 Drain tube (sold separately, 228-33833-91 PEEK TUBE 0.25×3000). p.7 ↗ Do not use the standard drain tube with ELSD; use the specified PEEK tube. A kit (228-68817-41)
Analytical Fraction MS kit A11 CV,OUT-UHP TP 228-53334-96 is a check valve to prevent backflow. p.7 ↗ Install it to protect the flow path from backflow. A kit (228-68817-41)
Prep SFC kit items X1 (SUS Sleeve Tubing 0.3x300, 228-53137-42, qty 2), X2 (SUS Sleeve Tubing 0.3x1200, 228-53184-41, qty 1) and X3 (FEMALE,1.6FT316L-0.4, 228-16008-11, qty 1) are not included in the kit. p.6 ↗ Order these items additionally when installing the Prep SFC modification kit. F kit (228-75753-42)
Basic kit B3 SUS Sleeve Tubing (0.3x1200) 228-53184-41 qty 2 is used for modifier pump–mixer and CO2 pump–mixer; B4 (0.3x700) 228-53184-93 is used for T-joint–column and, only on the Chiral Screening System, mixer–high pressure valve. p.2 ↗ Route these tubes to the specified connections; B4 has an additional use only on the Chiral Screening System. B kit
Basic kit B6 ETFE Tubing (1.6x1.0x2000) 228-18495-03 is for drain tubing and is used only on the Chiral Screening System. p.2 ↗ Install B6 only when configuring the Chiral Screening System. B kit

Modèles liés