Evaluating the Performance of the LotusStream Gas-Liquid Separator for Preparative Supercritical Fluid Chromatography
Technical Reference Summary
Covered Models & Document Scope
Informe técnico C190-E250 que evalúa el separador gas-líquido LotusStream en un sistema de cromatografía de fluidos supercríticos preparativa Nexera UC Prep; abarca pruebas de tasa de recuperación, arrastre y contaminación. No se indican variantes de modelo/tamaño del producto ni códigos de pedido.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| System | Nexera UC Prep Stacked fraction system p.2 ↗ | Recovery rate test |
| Flow rate | 125 mL/min p.2 ↗ | Recovery rate test |
| Detection wavelength | 272 nm p.2 ↗ | Caffeine, recovery rate test |
| BPR pressure | 10 MPa p.2 ↗ | Recovery rate test |
| BPR temperature | 50 °C p.2 ↗ | Recovery rate test |
| Sample concentration | 10 mg/mL p.2 ↗ | Caffeine, recovery rate test |
| Injection volume | 1 mL p.2 ↗ | Recovery rate test |
| Modifier | Methanol p.1 ↗ | LotusStream separator test |
| Modifier concentration | 20 % p.1 ↗ | LotusStream separator test |
| Modifier flow rate | 100 mL/min p.1 ↗ | LotusStream separator test |
| System | Nexera UC Prep Multi-fraction system p.3 ↗ | Contamination test |
| Makeup | Methanol p.3 ↗ | Contamination test |
| Makeup flow rate | 15 mL/min p.3 ↗ | Contamination test |
| BPR pressure | 10 MPa p.3 ↗ | Contamination test |
| Sample | Linalool 1 % p.3 ↗ | Contamination test |
| Fraction collector | FRC-40 SF p.3 ↗ | Contamination test |
| Column | Silica column (250 mm L. × 20 mm I.D.) p.4 ↗ | Carryover test |
| Makeup flow rate | 10 mL/min p.4 ↗ | Carryover test |
| Sample concentration | 30 mg/mL p.4 ↗ | Hydrocortisone, carryover test |
| Modifier | Methanol p.4 ↗ | Carryover test |
| Flow rate | 3 mL/min p.4 ↗ | Carryover test |
Operating Limits
- Contamination (Linalool): 0.006 % max. p.3 ↗
- Carryover (Hydrocortisone): 0.024 % p.4 ↗