c101-e181
Technical Reference Summary
Covered Models & Document Scope
Espectrofotómetro UV-Vis UV-1900i Plus; el documento abarca el modelo UV-1900i Plus y sus accesorios (portacubetas rectangular de paso largo, sippers 160L/160T/160V, portacubetas termostatizado, memoria ampliada P/N 207-23119-41, software LabSolutions UV-Vis).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Photometric accuracy (correlation coefficient) | 0.9998 p.8 ↗ | UV-1900i Plus |
| Wavelength range | 300.0-700.0 nm p.9 ↗ | UV-1900i Plus, spectrum display |
| Absorbance of blank solution (273 nm), n=3 | -0.00002 Abs p.9 ↗ | UV-1900i Plus |
| Absorbance of blank solution (273 nm), n=4 | 0.00002 Abs p.9 ↗ | UV-1900i Plus |
| Absorbance of blank solution (273 nm), n=6 | -0.00003 Abs p.9 ↗ | UV-1900i Plus |
| Absorbance of blank solution (273 nm), n=7 | 0.00001 Abs p.9 ↗ | UV-1900i Plus |
| Standard deviation of blank absorbance | 0.000025 Abs p.9 ↗ | UV-1900i Plus |
| Power consumption reduction in sleep mode | more than 85 % p.4 ↗ | Shutdown/Wakeup function vs normal operation |
| Recycled resin content of exterior panels | 50 % p.2 ↗ | UV-1900i Plus |
| Number of basic measurement modes | 6 p.5 ↗ | UV-1900i Plus |
| Maximum number of wavelengths in photometric mode | up to p.5 ↗ | Photometric mode |
| Long-path rectangular cell holder optical path lengths | 10, 20, 30, 50, 70 mm p.15 ↗ | P/N 204-23118-01 |
| Sample volume, 160L standard sipper | 2.0 mL p.15 ↗ | P/N 206-23790-51 |
| Sample volume, 160T triple-pass sipper | 1.5 mL p.15 ↗ | P/N 206-23790-52 |
| Sample volume, 160V 50 mm sipper | 5.0 mL p.15 ↗ | P/N 206-23790-56 |
Operating Limits
- Power consumption in sleep mode relative to normal operation: more than 85 % reduction p.4 ↗
- Recycled material content of exterior resin panels: 50 % p.2 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| Expanded Memory p.15 ↗ | 207-23119-41 | Accessory | ||
| Long-Path Rectangular Cell Holder p.15 ↗ | 204-23118-01 | Optical path lengths 10, 20, 30, 50, 70 mm | ||
| 160L (Standard Sipper) p.15 ↗ | 206-23790-51 | Sample volume 2.0 mL | ||
| 160T (Triple-Pass Sipper) p.15 ↗ | 206-23790-52 | Sample volume 1.5 mL | ||
| 160V (50 mm Sipper) p.15 ↗ | 206-23790-56 | Sample volume 5.0 mL | ||
| Temperature Controlled Cell Holder p.15 ↗ | 206-29510-42 | Variant -42 | ||
| Temperature Controlled Cell Holder p.15 ↗ | 206-29510-43 | Variant -43 | ||
| Temperature Controlled Cell Holder p.15 ↗ | 206-29510-58 | Variant -58 |