Agilent 9500 ICP-QQQ Dual-Cell System (DCS) and Advanced Helium Mode (AHM)

Sommaire

  1. Overview p. 2
    Introduces DCS and AHM capabilities for ICP-QQQ
  2. Overview p. 3
    Explains helium kinetic energy discrimination enhancement
  3. Overview p. 4
    Overview and advantages of Advanced Helium Mode
  4. p. 5
    Positions DCS and AHM as new standard
  5. p. 6
    Website link for more information

Engineering Facts

12 facts · every claim linked to a page
Selection

AHM consolidates no gas, He, and HEHe modes into a single mode, so a single method can cover most routine applications without multi-tune switching.

Reduces the number of required tune setups and method complexity when selecting the instrument configuration.

Source
“AHM not only replaces traditional He KED but also consolidates multiple tuning modes including no gas, He, and HEHe into a single, streamlined mode.”
Performance

AHM delivers roughly 20-fold sensitivity gain for low-mass elements and about 2-fold for mid-to-high mass elements versus conventional He mode.

Quantifies the sensitivity benefit that justifies choosing DCS/AHM over traditional He-KED.

Source
“Around a 20-fold sensitivity improvement for low-mass elements, such as beryllium (Be) and boron (B)”
Performance

AHM cuts data acquisition time by more than 33% while maintaining detection limits.

Shorter acquisition directly affects throughput and cost per sample in routine labs.

Source
“A reduction of more than 33% in data acquisition time while maintaining detection limits (DLs), accelerating analytical measurement times”
Installation

The DCS sits between Q1 and Q2 and uses two independently biased ion guides controlled by Front and Rear Cell bias voltages.

Understanding the dual-bias architecture is essential for correct setup and tuning of the cell.

Source
“Each of the two ion guides is independently controlled using bias voltages, referred to as the Front Cell bias voltage and Rear Cell bias voltage.”
Operation

In He KED, raising He flow increases collisions but gains plateau around 6 mL/min because excessive thermalization lowers KED effectiveness.

Guides practical He flow settings and explains why simply increasing flow does not keep improving performance.

Source
“while higher He flow rates boost both sensitivity and BEC, the gains plateau around 6 mL/min.”
Operation

Increasing KED voltage improves BEC but reduces analyte sensitivity, so a trade-off must be tuned per method.

Operators must balance background suppression against signal loss when setting KED voltage.

Source
“increasing the KED voltage enhances BEC but reduces sensitivity.”
Operation

To keep KED effective at higher He flows, ion cell entry energy must be raised by lowering the ion guide bias voltage.

Explains the bias-voltage adjustment needed when running elevated He flows.

Source
“it is necessary to increase the ion’s cell entry energy, which is achieved by lowering the ion guide bias voltage.”
Operation

In AHM, KED is effectively disabled for low-mass elements (m/z < 23) while standard settings are kept for m/z ≥ 23, switched in sync with Q2.

Clarifies the automatic mass-dependent behavior operators should expect during acquisition.

Source
“different voltages are applied to the DCS (effectively disabling KED) for low-mass elements, while keeping standard settings for elements with m/z ≥ 23.”
Operation

Mode switching between no gas, He, and HEHe introduces delays due to cell pressure stabilization requirements.

Explains why consolidating modes into AHM saves time and reduces stabilization waits.

Source
“Each mode switch introduces a delay due to cell pressure stabilization requirements.”
Performance

AHM achieves about one-third the sensitivity of no gas mode for Be, while still far exceeding He mode.

Sets realistic expectations for low-mass sensitivity relative to no gas mode.

Source
“For Be, AHM achieves about one-third the sensitivity of no gas mode.”
Performance

In a 20 ppm Ba matrix, AHM gave twice the sensitivity of He mode with lower BECs for both Eu isotopes.

Demonstrates interference removal performance for oxide interferences in challenging matrices.

Source
“AHM provided twice the sensitivity of He mode and lower BECs for the determination of Eu”
Selection

The DCS dual ion guide design uses a combined KED + CID approach that surpasses traditional HEHe and He KED limitations.

Key differentiator to weigh when comparing against older CRC-based instruments.

Source
“Its dual ion guide design and dual-mechanism approach (KED + CID) surpass the limitations of traditional high-energy HEHe and He KED modes”