Innovative Freeform Optical Design

Sommaire

  1. Overview pp. 1-2
    Introduction and advantages of freeform optics
  2. Overview p. 3
    Explanation of what freeform optics are
  3. Spec p. 4
    Detector specifications and capabilities
  4. Spec p. 5
    Signal readout performance details
  5. p. 6
    Summary of freeform optics benefits
  6. p. 3 table, 7 rows

Table 1. Typical resolution performance of the Agilent 5800 and 5900 ICP-OES (based on FWHM).

Element and Wavelength (nm)Typical Resolution of 5800 and 5900 (pm)
As 188.980< 6.5
Mo 202.032< 7
Zn 213.857< 7.5
Pb 220.353< 7.5
Cr 267.716< 9.5
Cu 327.395< 13
Ba 614.171< 32

Engineering Facts

12 facts · every claim linked to a page
Performance

The freeform optics deliver full wavelength coverage from 167 to 785 nm with true simultaneous measurement, which matters when selecting an instrument for multi-element throughput.

Defines the spectral range the instrument can cover in a single simultaneous measurement, a key selection criterion.

Source
“full wavelength coverage from 167 to 785 nm”
Performance

Detection limits improve by about 40% on average versus traditional optics, which affects achievable trace-level reporting.

Quantifies expected sensitivity gain relevant to method requirements and LOD targets.

Source
“improvements in detection limits by 40% on average”
Installation

The optics volume is halved, so purge and warm-up times are shorter, reducing wait time before samples can be measured.

Shorter purge/warm-up affects lab workflow and startup time after installation.

Source
“The polychromator optical system has reduced in volume by 50% compared to previous designs”
Installation

The compact footprint saves bench space, which matters for lab layout planning.

Footprint is a direct constraint when planning bench space and installation.

Source
“As one of the smallest ICP-OES available, the 5800 and 5900 ICP-OES save valuable bench space in the laboratory.”
Operation

The optical system has no moving parts and is temperature-controlled, supporting long-term stability during operation.

No moving parts and thermal control reduce drift and maintenance concerns in routine operation.

Source
“The system has no moving parts and is temperature-controlled to provide excellent long-term stability.”
Performance

The focal length is reduced to 253 mm while resolution and sensitivity improve, which is relevant when comparing optical designs.

Focal length and resolution trade-offs are central to evaluating optical performance claims.

Source
“reduce the optical focal length to just 253 mm, while improving resolution and sensitivity across the detector”
Operation

The detector is Peltier cooled to -40 °C to minimize dark current and readout noise, which affects low-light measurement quality.

Detector cooling temperature influences noise floor and thus detection capability.

Source
“The detector is Peltier cooled to –40 °C to minimize dark current and readout noise.”
Performance

The CCD has about 70,000 pixels across 70 DLAs covering 167 to 785 nm, which defines the detector's wavelength coverage architecture.

Detector pixel/DLA layout determines wavelength coverage and simultaneous measurement capability.

Source
“approximately 70,000 light sensitive pixels spread across 70 diagonal linear arrays (DLAs)”
Performance

The full spectrum from 167 to 785 nm can be measured in less than half a second, which sets throughput expectations.

Readout speed directly affects sample throughput and analysis time planning.

Source
“the entire spectrum from 167 to 785 nm can be measured in less than half a second”
Operation

Adaptive Integration Technology sets optimum integration time per wavelength, avoiding sequential grouped measurements and overrange signals.

Explains how the system handles mixed concentration ranges in a single measurement.

Source
“AIT automatically sets the optimum integration time for each wavelength”
Operation

Antiblooming protection on every pixel prevents intense signals from bleeding into neighboring pixels, enabling trace analysis alongside high concentrations.

Critical for samples with high matrix concentrations where trace elements must still be measured.

Source
“Antiblooming protection ensures that trace elements can be accurately measured in the presence of high concentrations of other elements.”
Performance

The freeform mirror corrects spherical aberration, coma, and astigmatism across the wavelength range, improving peak shape and resolution.

Aberration correction underpins the claimed resolution and detection limit improvements.

Source
“The high degree freeform surface corrects spherical aberrations, coma, and astigmatism”