Agilent Inert Plus GC/MS System with Hydrogen Carrier Gas

Sommaire

  1. Overview pp. 1-3
    Introduction to hydrogen carrier gas transition and system benefits
  2. Quality pp. 4-6
    Spectral fidelity and library match results with hydrogen
  3. Quality pp. 7-8
    Source maintenance reduction and sensitivity performance data
  4. Spec pp. 9-17
    Application results for hydrocarbons, PAHs, and pesticides
  5. Spec p. 18
    VOC analysis performance with hydrogen carrier
  6. Ordering pp. 19-20
    Upgrade kit configurations and complete source options
  7. p. 21
    Summary of hydrogen carrier gas benefits
  8. p. 20 table, 10 rows
  9. p. 7 table, 21 rows
  10. p. 9 table, 7 rows
  11. p. 19 table, 17 rows
  12. p. 4 table, 2 rows
  13. p. 11 table, 22 rows
  14. p. 12 table, 40 rows
  15. p. 18 table, 4 rows

Table 9. Agilent HydroInert triple quadrupole GC/MS upgrade kit for the Agilent 7000C/D/E triple quadrupole mass spectrometers, with part numbers and item numbers indicated in Figure 15.

Item Number (Figure 15)Part NamePart Number
NAHydroInert GC/TQ upgrade5505-0084
17Insert*G7078-20901
12Repeller*G7078-20902
3Extractor source body*G7078-20903
9**Entrance lens –extended*, **G7006-60926
10Ion focus lens*G7078-20905
4Extractor lens–9 mm*G7078-20909
16H2 EI heater/sensor assembly*G7078-67920
NAWire, extractor lensG7000-60827
11Repeller insulator (x2)G1099-20133

* HydroInert-specific parts. ** The extended entrance lens for the HydroInert MS/MS system is longer than the entrance lens of a single quadrupole MS system.

Table 2. Comparison of LMS for a mixture of SVOCs, pesticides, and fragrance compounds using an extractor source with He carrier gas, an extractor source with H2 carrier gas, and an Agilent HydroInert source with H2 carrier gas.

AnalyteExtractor He LMSExtractor H2 LMSHydroInert H2 LMS
Aniline98.692.397.7
1,4-dichlorobenzene-d496.794.797.5
Nitrobenzene95.780.994.3
Benzoic acid93.387.797.2
Naphthalene-d896.993.997.6
Acenaphthene-d1098.397.397.8
2,4-dinitrophenol95.690.694.8
4-nitrophenol89.583.194.8
4,6-dinitro-2-methylphenol94.393.695.7
Pentachlorophenol90.785.489.1
4-aminobiphenyl96.995.897.9
Phenanthrene-d1097.793.397.2
Musk ambrette86.686.389.5
Fenitrothion95.688.397.0
Musk ketone95.889.898.2
Benzidine92.670.197.5
3,3'-dichlorobenzidine97.389.895.3
Chrysene-d1296.084.893.5
Benzo[b]fluoranthene97.870.198.4
Benzo[k]fluoranthene97.896.898.4
Perylene-d1294.479.493.1

Table 3. Tested compound classes, including example compounds, with performance abilities for hydrogen carrier gas with a typical Inert Plus source and the Agilent HydroInert source.

Compound ClassExample Compound(s)H2 Carrier Gas + Normal GC/MS SourceH2 Carrier Gas + HydroInert Source
Nitro CompoundsNitrobenzene, fenitrothion, ethalfluralinUnacceptableDifferentiating
Heavily Chlorinated CompoundsDDT, Endrin, heptachlor, BHC compounds, pentachlorophenolUnacceptableDifferentiating
PAHsBenzo(b)fluoranthene, benzo[g,h,i]perylene, fluorantheneNeutralNeutral
Alkanes >C24Tetracosane (C24), octatriacontane (C38)NeutralNeutral
PesticidesDeltamethrin, fipronil, permethrin, captanUnacceptableDifferentiating
Fragrance/Flavor CompoundsMusk ketone, musk ambrette, linaloolUnacceptableDifferentiating
VOCs1,4-dioxane, trichloromethane, bromodichloromethaneNeutralDifferentiating

– "Unacceptable" was declared when mass spectral infidelity or compound misidentification was observed. – "Neutral" was determined when results were similar across the normal GC/MS system and HydroInert source in the presence of hydrogen carrier gas. – "Differentiating" was stated when the HydroInert source retained mass spectral fidelity and compounds were identified correctly with a helium library as reference.

Table 7. List of parts in a fully assembled Agilent HydroInert source for an Agilent Inert Plus GC/MS system.

Item Number (Figure 15)Part NamePart Number
1SetscrewsG3870-20446
2ScrewsG3870-20021
3Extractor source body*G7078-20903
4Extractor lens–9 mm*G7078-20909
5Extractor lens insulatorG3870-20445
6FilamentG7005-60061
7Spring washer Flat washer3050-1301 3050-0982
8Lens insulatorG3870-20530
9Entrance lens*G7078-20904
10Ion focus lens*G7078-20905
11Repeller insulatorG1099-20133
12Repeller*G7078-20902
13Flat washer3050-0891
14Belleville spring washer3050-1301
15Repeller nut0535-0071
16H2 EI heater/sensor assembly*G7078-67910
17Insert*G7078-20901

* HydroInert-specific parts.

Table 1. Comparison of extractor and Agilent HydroInert sources when using H2 carrier gas for identification of DDT and DDT breakdown products.

Source with H2 Carrier GasRetention Time (min)Identified CompoundLibrary Match Score
Extractor13.04DDD79
HydroInert13.04DDT85

Table 5. Response Factors (RFs) for select compounds from EPA method 8270E (Table 4 in the EPA method)4, GC/MS single quadrupole analysis with He carrier gas, GC/MS single quadrupole analysis with Agilent HydroInert source and H2 carrier gas, and GC/MS/MS tandem quadrupole analysis with HydroInert source and H2 carrier gas.

CompoundResponse Factor (RF) from EPA Method 8270E4RF He GC/MSRF H2 HydroInert GC/MSRF H2 HydroInert GC/MS/MS
Acenaphthene0.91.31.10.2
Acenaphthylene0.91.91.40.1
Acetophenone0.011.20.41.0
Anthracene0.71.11.00.9
Benzo(a)anthracene0.81.41.51.0
Benzo(a)pyrene0.71.20.90.9
Benzo(b)fluoranthene0.71.41.21.2
Benzo(g,h,i)perylene0.51.11.01.3
Benzo(k)fluoranthene0.71.21.21.3
Bis(2-chloroethoxy)methane0.30.40.30.7
Bis(2-chloroethyl)ether0.70.80.60.5
Bis-(2-ethylhexyl)phthalate0.010.80.50.2
4-Bromophenyl-phenyl ether0.10.30.20.2
Butyl benzyl phthalate0.010.60.30.1
4-Chloroaniline0.010.40.40.6
4-Chloro-3-methylphenol0.20.30.20.4
2-Chloronaphthalene0.82.41.00.8
2-Chlorophenol0.80.80.70.5
4-Chlorophenyl-phenyl ether0.40.70.50.3
Chrysene0.71.21.10.4
Dibenz(a,h)anthracene0.41.11.00.2
Dibenzofuran0.81.71.51.4

Response Factor (RF) from EPA Method 8270E4

CompoundResponse Factor (RF) from EPA Method 8270E4RF He GC/MSRF H2 HydroInert GC/MSRF H2 HydroInert GC/MS/MS
Di-n-butyl phthalate0.011.30.80.8
3,3'-Dichlorobenzidine0.010.50.40.1
2,4-Dichlorophenol0.20.30.20.4
Diethyl phthalate0.011.41.00.6
Dimethyl phthalate0.011.41.00.8
2,4-Dimethylphenol0.20.30.30.4
4,6-Dinitro-2-methylphenol0.010.20.10.01
2,4-Dinitrophenol0.010.20.10.02
2,4-Dinitrotoluene0.20.40.20.03
2,6-Dinitrotoluene0.20.30.20.03
Di-n-octyl phthalate0.011.30.80.4
Fluoranthene0.61.21.20.4
Fluorene0.91.31.21.4
Hexachlorobenzene0.10.30.30.4
Hexachlorobutadiene0.010.20.20.4
Hexachlorocyclopentadiene0.050.30.10.03
Hexachloroethane0.30.20.10.14
Indeno(1,2,3-cd)pyrene0.51.21.21.0
Isophorone0.40.60.40.3
2-Methylnaphthalene0.40.70.71.7
2-Methylphenol0.70.70.60.5
4-Methylphenol0.61.00.30.7
Naphthalene0.71.110.9
2-Nitroaniline0.010.40.230.05
3-Nitroaniline0.010.30.180.03
4-Nitroaniline0.010.30.210.13
Nitrobenzene0.20.30.20.3
2-Nitrophenol0.10.20.110.1
4-Nitrophenol0.010.20.140.05
N-Nitroso-di-n-propylamine0.50.40.40.03
N-Nitrosodiphenylamine0.012.050.92.3
2,2'-Oxybis-(1-chloropropane)0.010.450.540.03
Pentachlorophenol0.050.180.10.1
Phenanthrene0.71.21.11.1
Phenol0.80.90.70.3
Pyrene0.61.31.20.4
1,2,4,5-Tetrachlorobenzene0.010.380.310.23
2,3,4,6-Tetrachlorophenol0.010.360.170.07
2,4,5-Trichlorophenol0.20.30.30.2
2,4,6-Trichlorophenol0.20.30.20.2

Table 6. NIST helium library scores for select low-level (µg/L) trihalomethanes analyzed with H2 carrier gas and the Agilent HydroInert source.

Retention Time (minutes)NIST LMS
Trichloromethane2.10793
Bromodichloromethane2.79286
Dibromochloromethane3.52688
Tribromomethane4.26780

Engineering Facts

14 facts · every claim linked to a page
Installation

Stainless steel tubing is strongly recommended for hydrogen service; if only copper is available, fresh copper lines must be installed.

Material compatibility and cleanliness of the gas path directly affect hydrogen carrier gas reliability and contamination risk.

Source
“It is highly recommended that stainless steel tubing be used with H2. If only copper tubing is available, install fresh copper lines.”
Installation

Gas filters are mandatory, especially with hydrogen generators, using the specified Agilent trap and carrier gas purifier part numbers.

Filtration protects the column and MS source from moisture, oxygen, and hydrocarbons that degrade performance with H2.

Source
“Always use gas filters, especially with H2 generators.”
Operation

Hydrogen's lower viscosity reduces inlet pressure, so column dimensions should be changed or carrier flow increased, with 0.5–1.0 mL/min optimal and 2.0 mL/min maximum for turbo pumps.

Flow and pressure must be re-optimized when switching from helium to avoid poor chromatography or pump overload.

Source
“Optimal range: 0.5 to 1.0 mL/min”
Operation

For GC/MS/MS with hydrogen, quench gas must be disabled by removing helium plumbing and setting quench gas to 0.00 mL/min in the method.

Leaving helium quench gas connected with hydrogen carrier will cause method errors and incorrect MS/MS operation.

Source
“Please remove the helium plumbing from the back of the electronics pressure control module (EPC) and set the quench gas to 0.00 mL/min in any GC/MS/MS method.”
Installation

HydroInert source parts must never be mixed with extractor source parts, especially when using hydrogen carrier gas.

Part intermixing can cause hydrogenation/dechlorination and invalidate mass spectral fidelity.

Source
“Do not mix the HydroInert source parts with the extractor source parts; this can cause unexpected and unwanted results, especially when using hydrogen carrier gas.”
Selection

The HydroInert source is not compatible with older MS models such as the 5975 Series, nor with stainless steel or inert sources.

This determines whether an existing instrument can be upgraded or a new system is required.

Source
“The HydroInert source is not compatible with older mass spectrometer models, such as the Agilent 5975 Series GC/MSD, nor with stainless steel or inert sources.”
Selection

Diffusion pump single quadrupole instruments are not supported; a turbo pump is required for the HydroInert source.

Pump type is a hard selection criterion before purchasing or installing the HydroInert source.

Source
“Diffusion pump single quadrupole instruments are not supported. A turbo pump is required.”
Installation

Before installing the HydroInert source, verify hardware support and ensure required components (transfer line tip seal, source wires, green and brown wires) are present on older systems.

Missing electrical isolation or wiring components will prevent proper source installation and operation.

Source
“Before installing the HydroInert ion source, verify that the instrument hardware supports use of this source.”
Maintenance

With hydrogen and the HydroInert source, 5,200 matrix injections were completed before source failure versus 365 for helium, and the extraction lens and repeller are recommended for replacement when dirty.

This sets realistic maintenance intervals and spare-parts planning for high-throughput hydrogen operation.

Source
“For the HydroInert source, 5,200 matrix injections were completed before the source failed acceptance criteria.”
Performance

The HydroInert source retains mass spectral fidelity so existing helium-based libraries such as NIST can still be used with hydrogen carrier gas.

Library compatibility avoids rebuilding compound libraries and simplifies method transfer.

Source
“HydroInert source with H2 carrier gas retains mass spectral fidelity and can allow users to continue to use existing He-based mass spectral libraries.”
Performance

For GC/MS/MS with hydrogen, four OFN IDL datasets averaged 2.9 fg, all passing the 4.0 fg maximum allowable IDL criterion.

Sensitivity performance must be verified against method IDL requirements before adopting hydrogen.

Source
“The four IDL values were 2.1, 4.0, 2.0, and 3.5 fg for an average of 2.9 fg IDL, all of which pass the criteria.”
Operation

Not every compound, matrix, or analysis will improve with hydrogen; a limited-capacity test is advised before transferring multiple instruments.

Pilot testing prevents costly site-wide conversion failures for incompatible applications.

Source
“Not every compound, matrix, nor analysis will have the same or improved results, and should be tested in a limited capacity before transferring multiple instruments to hydrogen.”
Selection

For VOCs, a 6 mm HydroInert extraction lens is recommended, though a 9 mm HydroInert lens may also be an option.

Lens choice affects sensitivity and spectral fidelity for the target compound class.

Source
“For VOCs, a 6 mm HydroInert extraction lens is recommended, but a 9 mm HydroInert extraction lens may also be an option.”
Installation

Triple quadrupole and single quadrupole HydroInert parts must be kept separate because entrance lenses differ in length and must not be swapped.

Swapping lenses between instrument types causes installation errors and degraded performance.

Source
“the triple quadrupole parts should be kept separate from the single quadrupole parts as the entrance lenses are different lengths and should not be swapped.”