Induztr
Résumé technique Visionneuse de documents Modèles associés

Analytical-Scale LC Purification with High Load and Low Dispersion

Technical Reference Summary

Covered Models & Document Scope

Resumen técnico de un flujo de trabajo de purificación a escala analítica que utiliza un Agilent 1260 Infinity II Analytical LC ampliado con el 1260 Infinity II Analytical-Scale Fraction Collector (G1364F); compara las columnas ZORBAX Eclipse Plus C18, InfinityLab Poroshell 120 y Pursuit XRs. No es una ficha técnica de producto con códigos de pedido.

Technical Data

Parameter Value Condition
Maximum flow rate for fraction collection 10 mL/min p.2 ↗ 1260 Infinity II Analytical-Scale Fraction Collector (G1364F)
Injection volume for quantitation 0.5 µL p.3 ↗
Injection volume for purification 75 to 100 µL p.3 ↗
Needle wash 10 s p.3 ↗ acetonitrile/water 1:1 (v:v)
Total run time 13 minutes p.3 ↗
Peak width (10%) — ZORBAX Eclipse Plus C18 0.085 p.3 ↗ ZORBAX Eclipse Plus C18
Resolution (EP) — ZORBAX Eclipse Plus C18 13.1 p.3 ↗ ZORBAX Eclipse Plus C18
Selectivity — ZORBAX Eclipse Plus C18 1.71 p.3 ↗ ZORBAX Eclipse Plus C18
Peak width (10%) — InfinityLab Poroshell 120 0.065 p.3 ↗ InfinityLab Poroshell 120
Resolution (EP) — InfinityLab Poroshell 120 16.3 p.3 ↗ InfinityLab Poroshell 120
Selectivity — InfinityLab Poroshell 120 1.73 p.3 ↗ InfinityLab Poroshell 120
Peak width (10%) — Pursuit XRs 0.085 p.3 ↗ Pursuit XRs
Resolution (EP) — Pursuit XRs 13.8 p.3 ↗ Pursuit XRs
Selectivity — Pursuit XRs 1.60 p.3 ↗ Pursuit XRs
Maximum injection volume without significant coelution — Pursuit XRs 100 µL p.5 ↗ Pursuit XRs, 4 mg target compound
Maximum injection volume without significant coelution — Pursuit XRs (repeated loading) 80 µL p.7 ↗ Pursuit XRs, high-concentration sample
Purity of collected fractions >99 % p.6 ↗ all three columns tested
Recovery 96 to 97 % p.6 ↗ average, all three columns tested
Recovery — Pursuit XRs (repeated loading) 97 % p.7 ↗ average
Maximum purified target compound more than 50 mg p.7 ↗ Pursuit XRs, four runs
Sample concentration — HBA 8 mg/mL p.2 ↗ acetonitrile/water
Sample concentration — target compound 40 mg/mL p.2 ↗ acetonitrile/water

Operating Limits

  • Maximum flow rate for fraction collection: 10 mL/min p.2 ↗
  • Maximum injection volume without significant coelution (Pursuit XRs): 100 µL p.5 ↗
  • Maximum injection volume without significant coelution (Pursuit XRs, high-concentration sample): 80 µL p.7 ↗