Turbo-V 551 Rack Controller Instruction Manual

Sommaire

  1. Overview pp. 16-18
    General safety and product overview information
  2. Procedure pp. 19-21
    Installation instructions and procedures
  3. Procedure pp. 22-24
    Operating instructions for the controller
  4. Warning pp. 25-27
    Error message descriptions and troubleshooting
  5. Spec pp. 208-213
    Technical specifications and controller details
  6. Procedure pp. 214-216
    Fuse holder assembly and replacement procedure
  7. Spec pp. 223-225
    PURGE-VENT connection pinout and specifications
  8. Spec pp. 226-228
    Electrical input and output specifications
  9. Procedure pp. 229-231
    Controlled venting operation instructions
  10. Spec pp. 232-234
    Serial communication protocol description
  11. Overview pp. 238-240
    Display window meanings and indicators
  12. Procedure pp. 244-246
    Soft start function description and operation
  13. Procedure pp. 250-252
    How to use the front panel controls
  14. Spec pp. 253-255
    Measurement data and parameters
  15. Procedure pp. 259-261
    PUMP SETTING menu configuration
  16. Procedure pp. 262-264
    INPUT/OUTPUT menu configuration
  17. pp. 268-270
    Configuration and operation examples
  18. pp. 283-286
    Vacuum Products Division contact details
  19. p. 255 table, 3 rows
  20. p. 256 table, 2 rows
  21. p. 212 table, 19 rows
  22. p. 242 table, 9 rows
  23. p. 240 table, 7 rows
  24. p. 252 table, 7 rows
  25. p. 281 table, 6 rows
  26. p. 285 table, 5 rows
  27. p. 205 table, 4 rows
  28. p. 206 table, 3 rows

Programming display messages

STATUSXXX HzHIGH SPEED ADJXXXX TO YYYY HzMessage
STATUSXXX HzHIGH SPEED ADJXXXX TO YYYY HzCHANGE OK
STATUSXXX HzHIGH SPEED ADJXXXX TO YYYY HzOUT OF LIMITS
STATUSXXX HzHIGH SPEED ADJXXXX TO YYYY HzCHANGE ABORTED

First level of the configuration menu

STATUSHIGH SPEED ADJ
XXX HzXXXX TO YYYY Hz
XXXX Hz

Значения приведены как напечатано на схеме меню; пустые ячейки — пустые строки

Tab. 2

ParameterSpecification
Input: Voltage100 ÷ 240 Vac (±10 %), 1-phase
Frequency50 to 60 Hz
Power800 VA maximum
Output for pump: Power during pump ramp-up430 W maximum
Power normal (pump water cooled)350 W maximum
Power normal (pump air cooled)220 W maximum
Operating temperature5 °C to +45 °C
Storage temperature-20 °C to +70 °C
Fuse: Mains2 x T10 AH (slow blow) 250 V 5x20 fuses
Serial communication (T-Plus kit)RS232 cable with a 9-pin D type male connector and a 9-pin D type female connector, and T-Plus software (optional)
Power cableWith European or NEMA plug 3 meters long (optional)
Pump cable5 meters long fixed cable
In compliance with normsEN 61010-1 (2001) EN 61326-1 (2006) Emission level : Class B for Residential Application Immunity level : for Industrial Application
Max altitude2000 m
Protection categoryIP 20
Internal use only
Weight (both models)6.3 kg (13.9 lbs)
Installation categoryII
Pollution degree2

Window Meanings

N.Read/WriteData TypeDescriptionAdmitted Values
310 to 399Reserved to Agilent service
400RACRC EPROM (QE)QE8XXXX (where “XXXX” are variable)
402RACRC Param. (PA)PA8XXXX (where “XXXX” are variables)
404RACRC parameter structureXXXX
406RAProgram Listing code & revisionXXXX
407RAParameter Listing code & revisionXXXX
500 to 502Reserved to Agilent service
503R/WNRS 485 address0 to 31 (default = 0)
504R/WLSerial type selection0 = RS 232 1 = RS 485 (default = 0)

Window Meanings

N.Read/WriteData TypeDescriptionAdmitted Values
126R/WNVent valve opening delay expressed in 0.2 sec0 to 65535 Corresponding to 0 to 13107 sec (Default = 15)
143R/WNExternal Fan Configuration0 = Always ON 1 = Automatic 2 = Serial (default = 2)
144R/WLExternal Fan activation0 = OFF 1 = ON (default = 0)
145R/WLPurge Valve open/close (N.C.)0 = close 1 = open (Default =1)
147R/WNVent Valve opening (Time expressed in 0.2 s)0 to 65535 bit 0 = infinite min 0.2 s, Max 13107 s (default = 0)
155RNPower Limit Applied expressed in W
157R/WLGas Load Type0 = Ar 1 = N2 (default = 0)

Counters / Controller Data

DisplayDescription
CYCLE XXXXNumber of start/stop cycles
TIME XXXXXmTime of last cycle (minutes)
P.LIFE XXXXXXtotal operating time (hours) of the pump
MNModel Number of the controller
SNSerial Number
FWFirmware version
PAParameter Listing version

Pressing once the COUNTERS pushbutton displays CYCLE, TIME, P.LIFE; pressing twice displays MN, SN; pressing a third time displays FW, PA.

Tab. 8

DescriptionPart number
Turbo-V 551 Rack ControllerX3501-64002
T-Plus communication software and serial cable kit9699883
Turbo-V HT Rack Air Cooling kitX3501-68001
Mains cable NEMA plug 3m long9699958
Mains cable European plug 3m long9699957
Turbo-V ventX3501-68002

RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION

EUROPENORTH AMERICAPACIFIC RIM
Fax00 39 011 9979 3301 781 860 9252
Fax Free00 800 345 345 00
Toll Free00 800 234 234 00800 882 7426, Option 3
vpt-customercare@agilent.comvpl-ra@agilent.comPlease visit our website for individual office information
http://www.agilent.com

Tab. 1

MESSAGEDESCRIPTIONREPAIR ACTION
CHECK CONNECTION TO PUMPWrong connection between the pump and the controller.Check connections between controller and pump. Press the START push-button twice to start the pump.
PUMP WAITING INTERLOCKThe interlock signal of J1 connector is activated by an interruption of the link between pin 3 and 9 or between pin 4 and 15 of P1 connector, or because the external interlock signal is open.Reset the short circuit between pin 3 - 9 and between pin 4 - 15 of J1 connector, or close the external interlock signal.
FAULT: PUMP OVERTEMP.The pump upper bearing temperature exceeded 60 °C.Wait until the temperature decrease below threshold value. Press the START push-button twice to start the pump.
FAULT: CONTROLLER OVERTEMPERATUREThe controller internal temperature exceeded 65 °C.Wait until the temperature decrease below threshold value. Check that the pump cable length has been set correctly. Press the START push-button twice to start the pump.

Error Messages

MESSAGEDESCRIPTIONREPAIR ACTION
FAULT: TOO HIGH LOADIn normal operation, the current drawn by the pump is higher than programmed.Check that the pump rotor is free to rotate. Press the START push-button twice to start the pump.
FAULT: SHORT CIRCUITDuring normal functioning (after the start-up phase), a short-circuit condition has been detected in the power circuit.Check connections and shortages between pump and controller. Press the START push-button twice to start the pump.
OVERVOLTAGEA failure has occurred in the controller power supply section, or the controller has received a disturbance from the mains.Press the START button twice to start the pump again. If the message is redisplayed, contact Agilent for maintenance.

Engineering Facts

14 facts · every claim linked to a page
Installation

The controller must be powered from a properly grounded 3-wire socket to avoid electric shock and meet CE requirements.

Grounding and approved cord/plug are mandatory for safe installation and CE compliance.

Source
“Use this power cord and plug in conjunction with a properly grounded power socket to avoid electrical shock and to satisfy CE requirements.”
Safety

The input power cable must be disconnected before any servicing because high voltage inside can be lethal.

Defines the lockout step required before maintenance work.

Source
“High voltage developed in the controller can cause severe injury or death. Before servicing the unit, disconnect the input power cable.”
Installation

The controller must be placed so that free air flows through its ventilation holes, whether used on a bench or in a rack.

Obstructed airflow causes overheating and overtemperature faults.

Source
“The Turbo-V controller can be used as a bench unit or a rack module, but it must be positioned so that free air can flow through the holes.”
Installation

The controller must not be installed in explosive, dusty, wet, or high-fire-risk environments.

Site selection must exclude hazardous and contaminated atmospheres.

Source
“Do not install or use the controller in an environment exposed to atmospheric agents (rain, snow, ice), dust, aggressive gases, or in explosive environments or those with a high fire risk.”
Operation

Operating ambient must stay between 0 °C and +45 °C with 0–95 % non-condensing humidity.

Exceeding the ambient range risks overtemperature trips and reduced lifetime.

Source
“temperature: from 0 °C to +45 °C;”
Operation

The pump must never be run unless its inlet is connected to the system or blanked off.

Prevents injury and equipment damage during commissioning.

Source
“Never operate the Turbo-V pump if the pump inlet is not connected to the system or blanked off.”
Maintenance

The controller itself requires no maintenance, but any system work must be done by authorized personnel only.

Defines who may service the unit and that no routine controller maintenance is needed.

Source
“The Turbo-V 551 Rack Controller does not require any maintenance. Any work performed on the system must be carried out by authorized personnel.”
Operation

After an automatic fault stop, START must be pressed once to reset and a second time to restart the pump.

Operators must know the two-press reset/restart sequence after faults.

Source
“If the pump has been stopped automatically by a fault, this push-button must be pressed once to reset the controller and a second time to restart the pump.”
Operation

The interlock is satisfied only when pins 3–9 and 4–15 of J1 are shorted or the external interlock is closed.

Wiring of the interlock loop is required for the pump to start.

Source
“Reset the short circuit between pin 3 - 9 and between pin 4 - 15 of J1 connector, or close the external interlock signal.”
Operation

A pump upper-bearing temperature above 60 °C triggers a pump overtemperature fault.

Cooling and thermal design must keep bearing temperature below this threshold.

Source
“The pump upper bearing temperature exceeded 60 °C.”
Operation

A controller internal temperature above 65 °C triggers a controller overtemperature fault, and the pump cable length setting must be verified.

Rack airflow and correct cable-length configuration are needed to avoid this trip.

Source
“The controller internal temperature exceeded 65 °C.”
Installation

Remote I/O and serial cables longer than 3 m must be shielded.

Shielding is required for EMC when routing long signal cables.

Source
“All connecting cables for remote I/O and serial connections, if longer than 3 m, must be of the shielded type.”
Installation

Serial and I/O cables must be shielded, no longer than 30 m, with the shield bonded to the connector case at least on the controller side.

Sets the maximum cable run and shield termination for reliable, EMC-compliant wiring.

Source
“A shielded cable of 30 m maximum length has to be utilized for both serial and I/O port connections.”
Operation

Purge and vent valves are normally closed, so they stay closed on power failure.

Fail-safe valve behavior must be considered in vent/purge system design.

Source
“The two valves are Normally Closed (N.C.) so if a power fail occurs, the valves will remain closed.”