Agilent 1260 Infinity II High Temperature GPC System User Manual

Sommaire

  1. Overview pp. 7-18
    System description, oven, autosampler, injection sequence, DataStream
  2. Spec pp. 19-26
    Power, bench space, extraction, physical and performance specifications
  3. Procedure pp. 27-34
    Unpacking and installing pump and main module
  4. Procedure pp. 35-56
    GUI, modes, configuration editors, methods, security, monitor
  5. Procedure pp. 57-68
    Solvent introduction, columns, loops, samples, leak handling
  6. Procedure pp. 69-102
    Diagnostics, start-up options, error messages and causes
  7. Procedure pp. 103-126
    Safety, cleaning, overtemperature tests, BPR, DRI, autosampler
  8. Ordering pp. 127-130
    Accessory kits and pump spares
  9. Warning pp. 131-142
    General safety, precautions, WEEE, radio, sound emission
  10. p. 25 table, 15 rows
  11. p. 24 table, 12 rows
  12. p. 12 table, 7 rows
  13. p. 18 table, 5 rows
  14. p. 11 table, 8 rows
  15. p. 66 table, 9 rows
  16. p. 79 table, 7 rows
  17. p. 80 table, 9 rows
  18. p. 81 table, 8 rows
  19. p. 92 table, 3 rows

Table 2 Performance Specifications of the 1260 Infinity II High Temperature GPC System

Unit TypeSpecification
Pump1 Flow Range0.0 – 5.0 mL/min (software controlled)
Flow Rate Reproducibility1≤0.07 % RSD, or ≤0.02 min SD whatever is greater, based on retention time at constant room temperature
Oven Temperature RangeAmbient to 220 ° C
Temperature Stability<0.05 ° C/h
Column Capacity6 x 30 cm columns
Detector Differential Refractometer Deflection Cell Volume8 µL
Noise2≤100 µV
Drift2≤10 mV/h
Wavelength890 nm
Autosampler Carousel Capacity40 x 2 mL Vials
Injection Volumeflushed fixed loop; 5 – 500 µL
Precision1≤1%
Instrument Control Microprocessor
Instrument InterfaceWindows 7 and 10, 64bit
Advanced PC Control

1 Refer to 1260 Infinity II Isocratic Pump User Manual for complete pump specifications 2 The technical specifications have been determined by experimentation and reflect the total system performance. The values quoted have been calculated from analyses performed in THF at 40 °C and in TCB at 160 °C.

Table 1 Physical Specifications 1260 Infinity II High Temperature GPC System

TypeSpecificationComments
Line voltage230 VAC (±10 %)Single phase with earth
Power consumption3300 VA, 2310 W/7900 BTU
Line frequency50 or 60 Hz (±5 %)
Size (width × depth × height)900 mmx 530 mm x 620 mm
Packaged Size (width × depth × height)930 mm x 760 mmx 860 mm
Weight136 kg (300 lbs.)
Ambient operating temperature15 – 30 °C (59 – 86 ° F)At constant temperature
Ambient non-operating temperature-40 – 70 °C (-4 – 158 ° F)
Operating altitudeUp to 2000 m (6562 ft.)
Non-operating altitudeUp to 4600 m (15091ft.)
Humidity<80 % at 25 – 30 °C
Safety standards: IEC, CSA, ULInstallation category II, Pollution degree 2For indoor use only.

Figure 5 Standard Oven Configuration

NumberComponent
1Oven Fan with Heater
2Temperature Sensors
3Differential Refractive Index Detector
4Sample Needle Assembly
5Column Mounts
6Injection Valve
7Vapour Sensor

Agilent 1260 DataStream

No.DescriptionNo.Description
1Channel start button2Status LEDs
3Channel analogue input4CTR connector
6PL-CAN connector (not used)7Aux I/O connector
8Power Switch9Voltage selector
5RS232 connector10Power inlet

Figure 4 Instrument rear view and connections

NumberConnection
1Refractive Index signal output ±1V FSD
2Auxiliary I/O - Pump stop
3Oven Power Cable connector
4Injection synchronisation to Agilent DataStream
5Serial connection to PC (Instrument Control software)
6LAN connection to Agilent pump
7Oven RI link cable connector
8Oven Digital link cable connector

Table 7 Auto-ignition temperatures of common GPC solvents with ELS detection

SolventAuto-ignition Temperature (°C)
DMSO215
Dibutoxymethane220
Tetrahydrofuran321
Dimethylformamide445
Toluene480
Acetone465
1,2,4 Trichlorobenzene (TCB)570
m-Cresol558
Chloronaphthalene558

Table 3 Error overview

TypeGUI Error MessageConditionsResponse Time (Timeout, Retries)LEDBeeperResulting ModeNotes
OvenTEMPERATURE "Oven temperature has become unstable"Temperature setpoint reached, however subsequently increased to > 5 °C above setpoint- (-,-)Flash RedNO-Not STBY
FAN“The fan is no longer operating correctly"Frequency < 10 kHz5 s (5 s)Flash RedYESSTBY
DOOR"Illegal opening of the oven door has occurred"I/O state change- (-,-)Flash OrangeYES-Heaters OFF while door open
V/F"Temperature sensor is not responding correctly"Temp = 0 °C- (-,-)Flash RedYESSTBY
SENSOR"The vapour sensor is not responding correctly"Open/Short circuit5 s (5 s)Flash RedYESSTBY
OVEN VAPOUR"Spillage may have occurred within the oven"Voltage < set threshold- (-, -)Flash OrangeYES-
2 min (-, -)Flash RedYESSTBY

Table 3 Error overview

TypeGUI Error MessageConditionsResponse Time (Timeout, Retries)LEDBeeperResulting ModeNotes
RIAZFAIL"Attempts to autozero the RI have failed"Timeout or False Zero20 min (10 min, 2)Flash OrangeNO-Not while initializing. Injection sequence is aborted.
PURGE"The purge valve is not functioning correctly"Timeout5 s (5 s)Flash OrangeNO-Not while initializing. Injection sequence is aborted.
AutosamplerINITIALIZATION"Failed to initialize autosampler"Command failed (response from autosampler)- (-, 0)Flash RedNO-
MOVE"Failed to move carousel correctly"- (-, 3)Flash RedNO-
VIALLOWER"Failed to lower vial from the oven"- (-, 0)Flash RedNO-
VIALLIFT"Failed to lift vial from the carousel"- (-, 0)Flash RedNO-
SYRINGE"Failed to move syringe"- (-, 0)Flash RedNO-
LOAD"Failed to switch valve to load position"- (-, 3)Flash RedNO-
INJ"Failed to switch valve to inject position"- (-, 3)Flash RedNO-

Table 3 Error overview

TypeGUI Error MessageConditionsResponse Time (Timeout, Retries)LEDBeeperResulting ModeNotes
CAROUSEL LOCK"Failed to lock carousel after being closed"Command failed during autosampler run- (-, 0)Flash RedNO-
AUTOCOMMS"The autosampler is no longer responding"Packet errors/ no ACK response from autosampler1.5 s (500 ms, 3)Flash RedNOno command success response
variable,(variable, 0)Flash RedNO
Pump
PUMPCOMMSERR“The pump is no longer responding.”No response, or invalid response from pump.~30s (~10s, 3)Flash RedYESSTANDBY
PUMPBADTYPEID"Pump has wrong type id"Pump is not of the type G7110B.- (-, -)Flash RedYESSTANDBY
PUMPERRCODEAn error code, along with an appropriate message. The message will vary based on the error code.Pump is in an error state.- (-, -)Flash RedYESSTANDBYSee Pump General Error Messages and Pump Module Error Messages
SEQABORTPUMPNRDY“Pump became not ready. See error log.”Pump became not-ready while the autosampler was running an injection sequence.- (-, -)Flash OrangeYESInjection sequence is aborted.

MCGV Fuse

Probable causeSuggested actions
1 Defective gradient valve.Restart the quaternary pump. If the error message appears again, exchange the gradient valve.
2 Defective connection cable (front panel to main board).Please contact your Agilent service representative.
3 Defective main board.Please contact your Agilent service representative.

Error ID: 2043; Valve Fuse 0: Channels A and B; Valve Fuse 1: Channels C and D. The gradient valve in the quaternary pump has drawn excessive current causing the electronic fuse to open.

Engineering Facts

14 facts · every claim linked to a page
Installation

The instrument requires a dedicated 230 VAC single-phase earthed supply with an RCCB, so site power must be planned before delivery.

Electrical infrastructure and breaker selection must be confirmed during site preparation.

Source
“The mains supply should be fitted with an RCCB (Residual Current operated Circuit-Breaker).”
Installation

The power plug must remain reachable at all times so the unit can be disconnected in an emergency.

Bench layout must leave free access behind the instrument for safe disconnection.

Source
“Make sure the power connector of the instrument can be easily reached and unplugged.”
Installation

Only the Agilent power cord matched to the region may be used, and the outlet must be grounded.

Prevents electric shock, short circuit and EMC non-compliance at commissioning.

Source
“Never use a power cord other than the Agilent Technologies power cord designed for your region.”
Installation

The bench must accommodate the footprint and the upward-opening oven, which needs 97 cm clear height.

Overhead obstructions above the bench must be checked before installation.

Source
“The oven opens upwards to a full height of 97 cm (38 in).”
Selection

The system draws 3300 VA / 2310 W, which drives the electrical supply and heat-load sizing.

Lab power capacity and HVAC load must be verified against this consumption.

Source
“3300 VA, 2310 W/7900 BTU”
Selection

The 136 kg weight and 900 x 530 x 620 mm size dictate floor loading and bench strength requirements.

Bench or floor must be rated for the mass and dimensions of the main module.

Source
“136 kg (300 lbs.)”
Operation

The oven reaches 220 °C with better than 0.05 °C/h stability, enabling high-temperature GPC at 160 °C.

Confirms the thermal envelope needed for TCB-based polyolefin analysis.

Source
“Ambient to 220 ° C Temperature Stability <0.05 ° C/h”
Operation

The autosampler carousel holds 40 x 2 mL vials, with position zero reserved as the index vial.

Sample throughput planning must account for only 39 usable injection positions.

Source
“The carousel will accommodate 40 x 2 mL sample vials, 39 of these (positions 1 to 39) are used for sample injection.”
Operation

An empty, clean, capped vial must always sit in position zero or the autosampler cannot initialize and waste disposal fails.

Missing index vial blocks initialization and risks polymer precipitation in the loop.

Source
“It is therefore important to keep an empty vial fitted with a clean, shiny cap in the zero position of the carousel at all times.”
Operation

Solvents must be degassed and filtered to 0.45 µm and be compatible with the column before use.

Improper solvent preparation causes baseline noise, blockages and column damage.

Source
“should be thoroughly degassed, filtered (0.45 µm) and fully compatible with the column(s).”
Safety

Solvents with auto-ignition temperatures below 190 °C must not be used in the system.

Solvent selection at method development must respect this fire-safety limit.

Source
“Do not use solvents with auto-ignition temperatures <190°C with the HT GPC System”
Operation

Ambient lab temperature must exceed the solvent melting point, e.g. above 19 °C for TCB, before pumping.

Cold labs can freeze TCB and block the pump; HVAC setpoint must be verified.

Source
“the ambient temperature must exceed 19°C for 1,2,4-trichlorobenzene before the solvent can be pumped”
Maintenance

Column changes require cooling to 40 °C, setting idle/low flow to 0.2 mL/min, then STANDBY before opening the oven.

Defines the safe, leak-checked procedure for column exchange at high temperature.

Source
“Change the Idle and Low flow rates to 0.2 mL/min and the Idle temperature to 40 °C.”
Maintenance

After column change the DRI reference cell must be purged for at least 20 minutes before use.

Insufficient purging leaves air bubbles and unstable RI baseline.

Source
“Purge the DRI reference cell for at least 20 min using the Control software.”