Agilent AA Spectrometer User's Guide

Sommaire

  1. Overview pp. 2-4
    Table of contents and regulatory notices
  2. Overview pp. 6-8
    Introduction to the guide and Agilent AA documentation
  3. Warning pp. 7-30
    Electrical, heat, gas, UV and flashback safety warnings
  4. Overview pp. 31-36
    Front and back views of flame and furnace instruments
  5. Procedure pp. 37-41
    Electrical connections, fuse information and gas supplies
  6. Procedure pp. 42-50
    Connecting, configuring IEEE addresses, autosampler and drivers
  7. Overview pp. 51-56
    Worksheet, reports windows and SpectrAA software usage
  8. Procedure pp. 57-60
    Power and startup sequence for the system
  9. Procedure pp. 61-66
    Schedule, cleaning, gas cylinders and flashback prevention
  10. p. 61 table, 5 rows
  11. p. 13 table, 1 rows

Maintenance and Troubleshooting

РазделСтраница
Schedule62
Cleaning63
Gas cylinders64
Preventing flashbacks65
Spare parts65

Hazard Type Classification Description

Hazard TypeClassificationDescription
Ultraviolet HazardRisk Group 3UV emitted from this product. Avoid eye and skin exposure to unshielded product.

Engineering Facts

14 facts · every claim linked to a page
Safety

Only trained operators should use the instrument, and improper use can create explosion, fire, or other hazards.

Determines personnel qualification requirements before installation and operation.

Source
“Only trained operators should use this instrument.”
Installation

The instrument must be connected to a properly grounded 3-wire outlet rated for the load and using only the Agilent-supplied power cord.

Electrical supply and grounding must be verified during site preparation to avoid shock and fire hazards.

Source
“Always use a 3-wire outlet with ground connection that is adequately rated for the load.”
Safety

Gas cylinders must be stored vertically, secured, ventilated, and kept below 52 °C because pressure relief devices release contents above that temperature.

Cylinder storage location and ventilation must be planned to prevent overheating and uncontrolled gas release.

Source
“Cylinders have pressure relief devices that will release the contents of the cylinder if the temperature exceeds 52 °C (125 °F).”
Safety

The flame door suppresses UV radiation and must always be reinstalled after removal; exposure with the door open must be limited.

Maintenance procedures that remove the flame door require controlled UV exposure and prompt reinstallation.

Source
“The flame door has the effect of suppressing radiation from the flame to the outside of the instrument.”
Operation

Only air, nitrous oxide, and acetylene may be used for flame operation; oxygen or oxygen-enriched air must never be used as oxidant.

Gas supply selection and cylinder procurement must exclude oxygen to prevent explosion.

Source
“This spectrometer is to be used only with air, nitrous oxide, and acetylene for flame operation.”
Operation

Acetylene supply pressure must remain below 105 kPa (15 psig), with the instrument designed for 65–100 kPa (9.5–14.5 psig).

Regulator selection and delivery pressure settings must be verified before flame ignition.

Source
“Use acetylene at pressures lower than 105 kPa (15 psig).”
Operation

Acetylene must be packed in acetone; other packing materials can corrode the gas control module and must not be used.

Cylinder specification must be checked at procurement to avoid corrosion and flashback risk.

Source
“Use only acetylene that is packed in acetone.”
Maintenance

Burner slot width limits are 0.47 mm for nitrous-oxide/acetylene and 0.54 mm for air/acetylene; exceeding them requires burner replacement.

Burner inspection and spare-parts planning must include slot-width measurement against these limits.

Source
“If the width of the burner slot exceeds 0.47 mm (0.0185 in) for a nitrous-oxide/acetylene burner or 0.54 mm (0.021 in) for an air/acetylene burner, the burner cannot be refurbished and must be replaced with a new one.”
Safety

After a flashback, the burner must be considered damaged and must not be reused; the incident must be reported to Agilent service before resuming operation.

Post-flashback procedure requires burner replacement and service notification before restart.

Source
“Burners involved in a flashback event should not be re-used.”
Operation

The graphite tube atomizer must never use pure hydrogen; only a prepackaged 95% argon (or nitrogen) and 5% hydrogen mixture is permitted.

Furnace gas supply selection must exclude pure hydrogen and use only approved premixed gas.

Source
“Never use pure hydrogen with the graphite tube atomizer since this could result in leakage and a potentially explosive accumulation of hydrogen.”
Safety

The Zeeman magnet produces up to 0.8 Tesla peak at mains frequency, so pacemakers and magnetic storage media must be kept at least 300 mm away.

Lab layout and operator access around the Zeeman instrument must respect the 300 mm exclusion distance.

Source
“To avoid interference with heart pacemakers or magnetic storage media, keep them at least 300 mm from the magnet.”
Installation

A USB-GPIB HS or HS+ converter is required to interface the PC with the Agilent AA spectrometer.

PC hardware and driver selection must include the correct GPIB converter before software installation.

Source
“A USB-GPIB HS or HS+ converter must be installed in your computer to interface the computer and the Agilent AA.”
Installation

In a Duo system, the instrument configured as Instrument #1 must retain its default IEEE address (1 for 55B or 9 for 240FS/240Z/280FS/280Z).

IEEE address configuration must be verified during Duo system commissioning to avoid communication failure.

Source
“Duo mode will only work if the instrument selected as ‘Instrument #1’ in the SpectrAA Instrument Configuration dialog box retains its default IEEE address (that is, 1 for 55B or 9 for 240FS, 240Z, 280FS or 280Z.”
Installation

SpectrAA software accepts only COM port numbers 1 to 4 for USB-to-RS-232 adapters.

COM port assignment must be constrained during PC setup to ensure software communication with accessories.

Source
“SpectrAA software only accepts values from 1 to 4.”