Polyamide Analysis on Agilent PL HFIPgel with Gel Permeation Chromatography
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre el análisis de poliamidas mediante cromatografía de permeación en gel (GPC) utilizando columnas Agilent PL HFIPgel (conjunto de dos columnas) con detección de índice de refracción en HFIP + 0.02 M NaTFAc; calibración con estándares estrechos de PMMA Agilent PM-1.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Solvent | HFIP p.2 ↗ | GPC mobile phase |
| Salt additive concentration | 0.02 M p.2 ↗ | NaTFAc in HFIP |
| Column set | two PL HFIPgel columns p.2 ↗ | two-column set |
| Detector | refractive index p.2 ↗ | |
| Sample concentration | 0.2 % (w/v) p.2 ↗ | polyamide sample |
| Calibration standard | PM-1 PMMA narrow standards p.2 ↗ | |
| Highest molecular weight standard excluded | 1,900,000 g/mol p.2 ↗ | Mp, excluded on PL HFIPgel columns |
| Standard Mp values | 790,000; 281,700; 144,000; 59,800; 28,900; 13,300; 5,720; 1,936; 1,020 g/mol p.2 ↗ | PMMA standards |
| Analysis temperature | 40 °C p.3 ↗ |
Operating Limits
- Highest molecular weight standard usable on PL HFIPgel columns: 1,900,000 g/mol excluded p.2 ↗
- Analysis temperature: 40 °C p.3 ↗