Polyvinylidene Fluoride Analysis on Agilent PLgel 10 µm MIXED-B and GPC/SEC
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación 5990-8328EN: análisis por GPC/SEC de fluoruro de polivinilideno (PVDF) en dimetilsulfóxido a 95 °C utilizando un Agilent PL-GPC 220 con dos columnas Agilent PLgel 10 µm MIXED-B, 300 × 7.5 mm (n.º de ref. PL1110-6100); verificado para el Agilent 1260 Infinity GPC/SEC System.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Column | PLgel 10 µm MIXED-B p.2 ↗ | 2 columns |
| Column dimensions | 300 × 7.5 mm p.2 ↗ | 2 columns |
| Column part number | PL1110-6100 p.2 ↗ | PLgel 10 µm MIXED-B, 300 × 7.5 mm |
| Eluent | Dimethyl sulfoxide p.2 ↗ | |
| Flow rate | 1.0 mL/min p.2 ↗ | |
| Injection volume | 100 µL p.2 ↗ | |
| Analysis temperature | 95 °C p.1 ↗ | column/dissolution temperature |
| Sample concentration | 3 mg/mL p.1 ↗ | nominal |
| Sample dissolution time | 4 h p.1 ↗ | at 95 °C |
| Autosampler hot zone temperature | 95 °C p.1 ↗ | PL-GPC 220 |
| Autosampler warm zone temperature | 40 °C p.1 ↗ | PL-GPC 220 |
| Calibration standards | narrow polydispersity polymethylmethacrylate (PMMA) p.2 ↗ |
Operating Limits
- Analysis temperature: 95 °C p.1 ↗
- Sample dissolution temperature: 95 °C p.1 ↗
- Sample dissolution time: 4 h p.1 ↗
- Autosampler hot zone temperature: 95 °C p.1 ↗
- Autosampler warm zone temperature: 40 °C p.1 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| PLgel 10 µm MIXED-B, 300 × 7.5 mm p.2 ↗ | PL1110-6100 | Column used in the application (2 columns required) |