Sensitive Determination of Polycyclic Aromatic Hydrocarbons in Tap Water
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que describe un método SPE en línea UHPLC para la determinación de 16 PAH en agua potable en un sistema Agilent 1200 Infinity Series Online SPE (Agilent 1260 Infinity LC con automuestreador); no se cubren variantes de modelo/tamaño de producto ni códigos de pedido.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Sample volume required | less than 2 mL p.2 ↗ | |
| Overall run time (enrichment + analytical run) | 30 minutes p.2 ↗ | |
| Number of PAHs monitored | 16 p.2 ↗ | |
| Autosampler sample capacity | 2 × 18, 6-mL vials p.2 ↗ | Agilent 1260 Infinity autosampler |
| Fluorescence detector acquisition | Multisignal acquisition, set at λex= 260 nm p.3 ↗ | Agilent 1290 Infinity fluorescence detector |
| Retention time RSD | 0.011–0.090 % p.4 ↗ | n = 7, 16 PAHs |
| Area RSD | 0.3–6.0 % p.4 ↗ | n = 7, 16 PAHs |
| Linearity R² | 0.99834–0.99996 p.5 ↗ | 16 PAHs |
| Limit of detection (LOD, S/N = 3) | 0.0003–0.0233 µg/L p.5 ↗ | 16 PAHs |
| Limit of quantification (LOQ, S/N = 10) | 0.0009–0.0775 µg/L p.5 ↗ | 16 PAHs |
| LOD range | 0.2–23 ng/L p.5 ↗ |
Operating Limits
- Minimum sample volume sufficient for analysis: less than 2 mL p.2 ↗
- Overall run time: 30 minutes p.2 ↗