Trace Analysis of Ammonia in Ethylene by Gas Chromatography
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que describe una configuración GC/NCD para amoníaco en trazas en etileno: GC Agilent 7890B con NCD Agilent 8255, válvula de muestra de gas de 6 puertos con bucle de muestra de 1 mL, columna megabore p/n CP8590 (Select Low Ammonia), superficies de acero inoxidable desactivadas UltiMetal Plus, OpenLAB CDS.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Valve/valve box temperature | 125 °C p.2 ↗ | |
| Column oven temperature program | 100 (hold 0.25 minutes), ramp 25 °C/min to 150 (hold 2.75 minutes) °C p.2 ↗ | |
| NCD base temperature | 200 °C p.2 ↗ | |
| NCD burner temperature | 900 °C p.2 ↗ | |
| Oxidizer flow during analysis | 30 mL/min p.2 ↗ | maximum flow, during matrix hydrocarbon elution |
| Oxidizer flow at ammonia elution | 4 mL/min p.2 ↗ | just before ammonia eluted to detector |
| Sample loop volume | 1 mL p.2 ↗ | gas sample valve |
| Valve switch-off time after injection | 0.25 minutes p.2 ↗ | |
| Ammonia run area | 264.08 p.3 ↗ | run 6 |
| Ammonia standard deviation | 7.20 p.3 ↗ | run 6 |
| Ammonia relative standard deviation (RSD) | 2.73 % p.3 ↗ |
Operating Limits
- Detection capability: low ppbv levels p.2 ↗
- Oxidizer flow maximum: 30 mL/min p.2 ↗
- Oxidizer flow minimum at ammonia elution: 4 mL/min p.2 ↗