Induztr
Résumé technique Visionneuse de documents Modèles associés

Hydrocarbon Class Analysis of Conventional and Synthetic AviationTurbine Fuels by ASTM D8396 Flow-modulated GCxGC-FID

Technical Reference Summary

Covered Models & Document Scope

Nota de aplicación sobre el análisis GCxGC-FID de combustibles de turbina de aviación convencionales y sintéticos según ASTM D8396 utilizando el Agilent Reverse Flow Modulator (RFM) y el Agilent 8890 GC; se comparan dos variantes del método: gas portador hidrógeno y gas portador helio.

Technical Data

Parameter Value Condition
Sampler Agilent 7693A Automated Liquid Sampler p.3 ↗ Hydrogen carrier
Sampler Agilent 7693A Automated Liquid Sampler p.3 ↗ Helium carrier
Inlet Split/Splitless p.3 ↗ Hydrogen carrier
Inlet Split/Splitless p.3 ↗ Helium carrier
Column 1 Agilent DB-17, 20 m × 0.18 mm, 0.18 µm p.3 ↗ Hydrogen carrier
Column 1 Agilent DB-17, 20 m × 0.18 mm, 0.18 µm p.3 ↗ Helium carrier
Detector FID p.3 ↗ Hydrogen carrier
Detector FID p.3 ↗ Helium carrier
Carrier gas Hydrogen p.3 ↗ Hydrogen carrier
Runtime 63.33 min p.3 ↗ Hydrogen carrier
Runtime 63.33 min p.3 ↗ Helium carrier
Column 1 flow 0.3 mL/min p.3 ↗ Hydrogen carrier
Column 1 flow 0.3 mL/min p.3 ↗ Helium carrier
Column 2 flow 15 mL/min p.3 ↗ Hydrogen carrier
Restrictor column flow 0.36 mL/min p.3 ↗ Hydrogen carrier
Restrictor column flow 0.36 mL/min p.3 ↗ Helium carrier
Septum purge 6 mL/min p.3 ↗ Hydrogen carrier
Septum purge 6 mL/min p.3 ↗ Helium carrier
Injection volume 0.1 µL p.3 ↗ Hydrogen carrier
Injection volume 0.1 µL p.3 ↗ Helium carrier
Inlet temperature 300 °C p.3 ↗ Hydrogen carrier
Inlet temperature 300 °C p.3 ↗ Helium carrier
Oven initial temperature 40 °C p.3 ↗ Hydrogen carrier

Operating Limits

  • Maximum programmable temperature: 300 °C p.8 ↗