Induztr
Résumé technique Visionneuse de documents Modèles associés

Per- and Polyfluoroalkyl Substances Analysis in Soil

Sommaire

  1. Overview p. 2
    Background on PFAS compounds and analytical approach
  2. Spec p. 3
    LC-MS/MS instrument parameters and settings
  3. Procedure p. 4
    Calibration standards and quantitation methodology
  4. p. 5
    Analysis results and interpretation
  5. Quality p. 6
    Blank sample quality control results
  6. Procedure p. 7
    Spike recovery results for sandy loam matrix
  7. Procedure p. 8
    PFAS analysis in reed sedge peat and topsoil
  8. p. 9
    Summary of findings and method performance
  9. p. 10
    Supplementary data tables
  10. p. 11
    Cited literature and sources
  11. p. 3 table, 7 rows
  12. p. 8 table, 40 rows
  13. p. 10 table, 40 rows
  14. p. 11 table, 16 rows

Table 1. PFAS-suitable consumables and supplies.

Agilent Consumables and SuppliesPart Number
Carbon S for PFAS cartridge, 250 mg, 6 mL5610-2247
Polypropylene autosampler screw top vials, 2 mL, and caps5191-8151 and 5191-8121
Centrifuge tubes and caps, 15 mL5610-2039
InfinityLab PFC delay column, 4.6 × 30 mm5062-8100
ZORBAX RRHD Eclipse Plus C18 column, 2.1 × 100 mm, 1.8 µm959758-902
Vac Elut SPS 24 manifold with collection rack for 10 × 75 mm test tubes12234003
Collection rack and funnel set for 12 or 15 mL conical tubes, for Vac Elut SPS 24 manifold12234027

Table 4. Average concentrations of PFAS measured in soil samples.

AcronymReed Sedge PeatTopsoil
10:2 FTCA<MRL<MRL
10:2 FTSA<MRL<MRL
10:2 FTUCA<MRL<MRL
11Cl-PF3OUdS<MRL<MRL
3:3 FTCA<MRL<MRL
4:2 FTSA<MRL<MRL
4-PFecHS<MRL<MRL
5:3 FTCA<MRL<MRL
6:2 FTCA<MRL<MRL
6:2 FTSA<MRL<MRL
6:2 FTUCA<MRL<MRL
6:2/8:2 diPAP<MRL<MRL
6:6 PFPi<MRL<MRL
6:8 PFPi<MRL<MRL
7:3 FTCA<MRL<MRL
8:2 diPAP<MRL<MRL
8:2 FTCA<MRL<MRL
8:2 FTSA<MRL<MRL
8:2 FTUCA<MRL<MRL
8:8 PFPi<MRL<MRL
9Cl-PF3ONS<MRL<MRL
diSAmPAP<MRL<MRL
DONA<MRL<MRL
EtFOSE<MRL<MRL
FBSA<MRL<MRL
FDSA<MRL<MRL
FHxSA<MRL<MRL
HFPO-DA<MRL<MRL
MeFOSE<MRL<MRL
N-EtFOSA<MRL<MRL
N-EtFOSAA<MRL<MRL
NFDHA<MRL<MRL
N-MeFOSA<MRL<MRL
N-MeFOSAA<MRL<MRL
PFBA4.5 ng/g<MRL
PFBS<MRL<MRL
PFDA<MRL<MRL
PFDoDA<MRL<MRL
PFDoS<MRL<MRL
PFDS<MRL<MRL

Таблица напечатана в PDF разбитой на три последовательные части с повторяющимся заголовком Acronym / Reed Sedge Peat / Topsoil; здесь объединена в одну.

Table A1. Target spiking solution and target concentration in matrix.

Target CompoundCASRet Time (min)Target Quant Ion MRM TransitionIsotope AnalogIsotope MRM Transition
PFBA375-22-42.43213 & 16913C4-PFBA217 & 172
PFMPA377-73-13.57229 & 8513C5-PFPeA268 & 223
3:3 FTCA356-02-53.88241 & 17713C5-PFPeA268 & 223
PFPeA2706-90-33.89263 & 21913C5-PFPeA268 & 223
PFBS375-73-53.97299 & 8013C3-PFBS302 & 80
PFMBA863090-89-54.01279 & 8513C5-PFPeA268 & 223
PFEESA113507-82-74.12315 & 13513C3-PFBS302 & 80
NFDHA151772-58-64.25295 & 8513C5-PFHxA318 & 273
4:2 FTSA757124-72-44.29327 & 30713C2-4:2 FTSA329 & 309
PFHxA307-24-44.33313 & 26913C5-PFHxA318 & 273
PFPeS2706-91-44.39349 & 8013C3-PFHxS402 & 80
HFPO-DA13252-13-64.48285 & 16913C3-HFPO-DA287 & 169
FBSA30334-69-14.58298 & 7813C3-PFHxS402 & 80
PFHpA375-85-94.90363 & 31913C4-PFHpA367 & 322
PFHxS355-46-44.95399 & 8013C3-PFHxS402 & 80
DONA919005-14-44.98377 & 25113C4-PFHpA367 & 322
5:3 FTCA914637-49-35.04341 & 23713C5-PFHxA318 & 273
6:2 FTUCA70887-88-65.06357 & 29313C2-6:2 FTUCA359 & 294
6:2 FTCA53826-12-35.11377 & 29313C2-6:2 FTCA379 & 294
4-PFecHS646-83-35.52461 & 38113C8-PFOS507 & 80
6:2 FTSA27619-97-25.57427 & 40713C2-6:2 FTSA429 & 409
PFOA335-67-15.60413 & 36913C8-PFOA421 & 376
PFHpS375-92-85.63449 & 8013C8-PFOS507 & 80
FHxSA41997-13-16.10398 & 7813C8-PFOS507 & 80
PFNA375-95-16.15463 & 41913C9-PFNA472 & 427
PFOS1763-23-16.17499 & 8013C8-PFOS507 & 80
8:2 FTUCA70887-84-26.36457 & 39313C2-8:2 FTUCA459 & 394
7:3 FTCA812-70-46.37441 & 33713C5-PFHxA318 & 273
8:2 FTCA27854-31-56.40477 & 39313C2-8:2 FTCA479 & 394
9Cl-PF3ONS756426-58-16.55531 & 35113C8-PFOS507 & 80
8:2 FTSA39108-34-47.02527 & 50713C2-8:2 FTSA529 & 509
PFDA335-76-27.07513 & 46913C6-PFDA519 & 474
PFNS68259-12-17.09549 & 8013C8-PFOS507 & 80
N-MeFOSAA2355-31-97.73570 & 419d3-N-MeFOSAA573 & 419
PFDS335-77-38.10599 & 8013C8-PFOS507 & 80
PFUnDA2058-94-88.11563 & 51913C7-PFUnDA570 & 525
N-EtFOSAA2991-50-68.13584 & 419d5-N-EtFOSAA589 & 419
PFOSA754-91-68.32498 & 7813C8-PFOSA506 & 78
10:2 FTUCA70887-94-48.36557 & 49313C2-10:2 FTUCA559 & 494
10:2 FTCA53826-13-48.40577 & 49313C2-10:2 FTCA579 & 494

Appendix A

Table

Target CompoundCASRet Time (min)Target Quant Ion MRM TransitionIsotope AnalogIsotope MRM Transition
6:6 PFPi40143-77-98.90701 & 40113C2-PFDoDA615 & 570
PFDoS79780-39-59.09699 & 8013C8-PFOS507 & 80
PFTrDA72629-94-89.12663 & 61913C2-PFDoDA615 & 570
PFTDA376-06-79.49713 & 66913C2-PFTDA715 & 670
N-MeFOSA31506-32-89.50512 & 219d3-N-MeFOSA515 & 169
FDSAN/A9.52598 & 7813C8-PFOSA506 & 78
6:8 PFPi610800-34-59.54801 & 401(13C2)2-6:2 diPAP993 & 97
MeFOSE24448-09-79.54616 & 59d7-MeFOSE623.1 & 59
N-EtFOSA4151-50-29.88526 & 219d5-N-EtFOSA531 & 169
EtFOSE1691-99-29.89630 & 59d9-EtFOSE639.1 & 59
6:2/8:2 diPAP943913-15-310.02889 & 443(13C2)2-6:2 diPAP793 & 97
8:8 PFPi40143-79-110.11901 & 501(13C2)2-6:2 diPAP793 & 445
PFHxDA67905-19-510.18813 & 26913C2-PFHxDA815 & 770
8:2 diPAP678-41-110.55989 & 543(13C2)2-8:2 diPAP993 & 97
PFODA16517-11-610.81913 & 36913C2-PFHxDA815 & 770
diSAmPAP2965-52-811.101,203 & 526(13C2)2-8:2 diPAP993 & 97

Engineering Facts

11 facts · every claim linked to a page
Selection

Choose Carbon S for PFAS cartridges when both low PFAS background and high recovery of GCB-sensitive analytes are required.

It justifies selecting Carbon S over conventional GCB for PFAS soil cleanup.

Source
“Carbon S sorbent provides a better balance between analyte recovery and pigment removal efficiency than traditional GCB sorbent.”
Selection

The cartridge is manufactured specifically for PFAS work so common PFAS contaminants stay below typical reporting limits.

Low-residue consumables are essential to reach ng/g detection limits without false positives.

Source
“Bond Elut Carbon S for PFAS has been manufactured for PFAS applications, ensuring that common PFAS contaminants are below typical reporting limits.”
Installation

Use a PFC delay column in the LC flow path to keep PFAS background under control.

The delay column is a required hardware item for low-level PFAS analysis.

Source
“InfinityLab PFC delay column, 4.6 × 30 mm”
Operation

Condition the Carbon S cartridge with two 5 mL methanol rinses and discard the rinsate before loading sample extract.

Proper conditioning prevents carryover and ensures consistent cleanup performance.

Source
“Rinse the Carbon S cartridge with two 5 mL volumes of methanol and discard.”
Operation

Load the extract in two portions and elute under gravity or slight vacuum to avoid forcing matrix through the sorbent.

Controlled flow protects recovery and pigment removal efficiency.

Source
“Decant the extract into Carbon S cartridges in two portions, elute under gravity or with a slight vacuum.”
Operation

Neutralize the collected eluate with 50 µL acetic acid before combining with internal standard and water.

Neutralization is required before LC/MS/MS injection for consistent analyte response.

Source
“Neutralize with 50 µL of acetic acid.”
Performance

Expect average recovery near 99% with RSD around 13.5% at low spike levels for the 59 PFAS tested.

This sets realistic acceptance criteria for method validation.

Source
“The average recovery for all compounds was 99.3% with an RSD of 13.5%.”
Performance

Carbon S cleanup improves early-eluting PFBA peak shape and retention time consistency in peat extracts.

Matrix cleanup directly affects chromatographic data quality for early eluters.

Source
“the use of Carbon S improved the peak shape integrity and retention consistency of PFBA compared to extracts without the use of Carbon S cleanup.”
Conformité

The extraction procedure follows ASTM D7968-17a except that syringe filtration is replaced by Carbon S passthrough cleanup.

Knowing the deviation from the validated standard is needed for method documentation and compliance.

Source
“The sample preparation closely followed the extraction procedure in ASTM D7968‑17a except for replacing the syringe filtration step with a passthrough cleanup using the Carbon S for PFAS cartridge.”
Maintenance

Run extraction blanks and cartridge rinsate blanks with each extraction set to confirm the system and cartridges are PFAS-free.

Blank monitoring is the primary ongoing contamination control measure.

Source
“Cartridge blanks (rinsate collected from the methanol rinse) and matrix blanks were also analyzed to ensure the system and cartridges were free from PFAS contamination before sample analysis.”
Operation

Use a tube rotator to fully invert sample tubes during the 1-hour extraction for reproducible contact.

Consistent agitation is needed to meet recovery and precision targets.

Source