Using a Deans Switch to Divert Derivatizing Agents from the EI Source for Improved Robustness

Technical Reference Summary

Covered Models & Document Scope

Nota de aplicación sobre el uso de un conmutador Deans para desviar agentes de derivatización (MSTFA) de la fuente de EI en flujos de trabajo de GC/MS/MS para antidopaje. Abarca dos configuraciones de instrumentos: Agilent 7000 Series GC/TQ con gas portador hidrógeno y Agilent 7010 Series GC/TQ con gas portador helio; dos tipos de entrada (entrada split/splitless SSL y entrada de vaporizador de temperatura programable).

Technical Data

Parameter Value Condition
Purge flow to split vent 50 mL/min p.6 ↗ SSL inlet, at 0.75 min
Purge flow to split vent 50 mL/min p.6 ↗ PTV inlet, at 0.75 min
Injection volume 1.5 µL p.6 ↗ SSL inlet
Switching time 4.9 min p.6 ↗ SSL inlet, from vent to MS
Column temperature program 2.5 °C/min to 199 °C (0 min hold) °C p.6 ↗ SSL inlet
Carrier gas and flow rate Hydrogen, 1 mL/min constant flow p.6 ↗ SSL inlet
Transfer line temperature 300 °C p.6 ↗ SSL inlet
Electron energy 70 eV p.6 ↗ SSL inlet
Electron energy 70 eV p.6 ↗ PTV inlet
Quench gas (helium) 2.25 mL/min p.6 ↗ PTV inlet
Collision gas (nitrogen) 1.5 mL/min p.6 ↗ SSL inlet
Collision gas (nitrogen) 1.5 mL/min p.6 ↗ PTV inlet
Ion source temperature 280 °C p.6 ↗ SSL inlet
Ion source temperature 280 °C p.6 ↗ PTV inlet
Tune Etune: atunes.eiex.tune.xml p.6 ↗ SSL inlet
Cleaning interval 350–400 to ≥ 1,500 injections p.1 ↗ Agilent 7000 Series GC/TQ, hydrogen carrier gas

Operating Limits

  • Source cleaning interval improvement (7000 Series GC/TQ, hydrogen): from approximately 350–400 to ≥ 1,500 injections p.1 ↗
  • Source cleaning interval improvement (7010 Series GC/TQ, helium): from approximately every 300 to 350 injections to approximately 900 injections p.1 ↗
  • Routine batch size: 80 samples p.1 ↗
  • Source cleaning frequency with routine batches: shifted from weekly to every 2 to 3 weeks p.1 ↗