Quantitation of Per- and Polyfluoroalkyl Substances (PFAS) in Chicken Eggs for Human Consumption
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre la cuantificación de PFAS en huevos de gallina utilizando cartuchos de extracción en fase sólida Agilent Bond Elut Carbon S y un sistema LC/MS de triple cuadrupolo Agilent 6475; incluye limpieza de muestra QuEChERS y método LC/MS/MS.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Injection volume | 5 µL p.4 ↗ | Agilent 6475 LC/MS |
| Ionization polarity | Negative p.5 ↗ | |
| Sheath gas temperature | 375 °C p.5 ↗ | |
| Sheath gas flow | 11 L/min p.5 ↗ | |
| Nebulizer gas pressure | 25 psi p.5 ↗ | |
| PFOS maximum level in eggs | 1.0 µg/kg p.2 ↗ | EU regulation 2023/915 |
| PFOA maximum level in eggs | 0.30 µg/kg p.2 ↗ | EU regulation 2023/915 |
| PFNA maximum level in eggs | 0.70 µg/kg p.2 ↗ | EU regulation 2023/915 |
| PFHxS maximum level in eggs | 0.30 µg/kg p.2 ↗ | EU regulation 2023/915 |
| Sum of four PFAS maximum level in eggs | 1.7 µg/kg p.2 ↗ | EU regulation 2023/915 |
| Limit of quantification for PFOS in eggs | 0.30 µg/kg p.2 ↗ | EU Commission Recommendation 2022/1431 |
| Limit of quantification for PFOA in eggs | 0.30 µg/kg p.2 ↗ | EU Commission Recommendation 2022/1431 |
| Limit of quantification for PFNA in eggs | 0.30 µg/kg p.2 ↗ | EU Commission Recommendation 2022/1431 |
| Limit of quantification for PFHxS in eggs | 0.30 µg/kg p.2 ↗ | EU Commission Recommendation 2022/1431 |
| PFPeA calibration range | 0.15 to 2.0 µg/kg p.4 ↗ | internal standard 13C5-PFPeA |
| PFBS calibration range | 0.13 to 1.8 µg/kg p.4 ↗ | internal standard 13C3-PFBS |
| PFHxA calibration range | 0.15 to 2.0 µg/kg p.4 ↗ | internal standard 13C5-PFHxA |
| HFPO-DA calibration range | 0.3 to 4.0 µg/kg p.4 ↗ | internal standard 13C5-PFHxA |
| PFOA calibration range | 0.15 to 2.0 µg/kg p.4 ↗ | internal standard 13C8-PFOA |
| PFOS calibration range | 0.14 to 1.92 µg/kg p.4 ↗ | internal standard 13C8-PFOS |
| PFNA calibration range | 0.15 to 2.0 µg/kg p.4 ↗ | internal standard 13C9-PFNA |
| PFOSA calibration range | 0.3 to 4.0 µg/kg p.4 ↗ | internal standard 13C2-PFDoDA |
| PFHpS recovery | 95.5 % p.6 ↗ | RSD 10.3% |
| PFOSA recovery | 109.0 % p.6 ↗ | RSD 17.5% |
| PFDS recovery | 107.4 % p.6 ↗ | RSD 9.0% |
Operating Limits
- Maximum level of PFOS in eggs: 1.0 µg/kg p.2 ↗
- Maximum level of PFOA in eggs: 0.30 µg/kg p.2 ↗
- Maximum level of PFNA in eggs: 0.70 µg/kg p.2 ↗
- Maximum level of PFHxS in eggs: 0.30 µg/kg p.2 ↗
- Maximum level of sum of four PFAS in eggs: 1.7 µg/kg p.2 ↗
- Limit of quantification for PFOS, PFOA, PFNA, PFHxS in eggs: 0.30 µg/kg p.2 ↗