Analyse des dioxines dans l’alimentation humaine et animale
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación (publicación de Agilent 5991-9399FR) sobre el análisis de dioxinas en alimentos para consumo humano y animal mediante un sistema GC-QQQ Intuvo 9000/7010; abarca dos métodos cromatográficos: un método de 52.5 minutos y un método acelerado de 31.12 minutos.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Injection volume | 1 μL p.2 ↗ | 52.5-minute method |
| Column flow | 1.6 mL/min p.2 ↗ | 52.5-minute method |
| Column flow | 1.6 mL/min p.2 ↗ | 31.12-minute method |
| Guard Chip temperature | 310 °C p.2 ↗ | 52.5-minute method |
| Guard Chip temperature | 310 °C p.2 ↗ | 31.12-minute method |
| Bus temperature | 325 °C p.2 ↗ | 52.5-minute method |
| Bus temperature | 325 °C p.2 ↗ | 31.12-minute method |
| Splitless pulsed pressure | 60 psi p.2 ↗ | 0.6 minute, 50 mL/min |
| Calibration levels | 6 points p.3 ↗ | 52.5-minute method |
| Calibration range (Tetra) | 50 to 80 000 fg/µL p.3 ↗ | L1–L6 |
| Calibration range (Penta) | 100 to 160 000 fg/µL p.3 ↗ | L1–L6 |
| Calibration range (Hexa) | 100 to 160 000 fg/µL p.3 ↗ | L1–L6 |
| Calibration range (Hepta) | 200 to 320 000 fg/µL p.3 ↗ | L1–L6 |
| Calibration range (Octa) | 200 to 320 000 fg/µL p.3 ↗ | L1–L6 |
| Coefficient of determination R² (2378-TCDD) | 0.99997 p.3 ↗ | 52.5-minute method |
| Coefficient of determination R² (OCDD) | 0.99999 p.3 ↗ | 52.5-minute method |
| MDL (2378-TCDF) | 3.8184 fg/µL p.4 ↗ | |
| LOQ (2378-TCDF) | 13.5334 fg/µL p.4 ↗ | |
| LOD (2378-TCDF) | 4.06 fg/µL p.4 ↗ | |
| MDL (OCDD) | 16.7276 fg/µL p.4 ↗ | |
| LOQ (OCDD) | 59.2874 fg/µL p.4 ↗ | |
| LOD (OCDD) | 17.7862 fg/µL p.4 ↗ | |
| Retention time (2378-TCDD) | 27.835 min p.4 ↗ | |
| Retention time (OCDD) | 48.83 min p.4 ↗ |
Operating Limits
- Method duration (long method): 52.5 min p.2 ↗
- Method duration (accelerated method): 31.12 min p.2 ↗
- Maximum injection volume: 1 μL p.2 ↗
- Minimum injection volume: 0.5 μL p.2 ↗