Refinery Gas Analysis Using the Agilent 990 Micro GC

Technical Reference Summary

Covered Models & Document Scope

Nota de aplicación que describe dos configuraciones de RGA (análisis de gases de refinería) en el Agilent 990 Micro GC: Configuración RGA 1 (cuatro canales: 10 m CP-Molesieve 5 Å backflush (RTS), 10 m CP-PoraPLOT U backflush, 10 m CP-AL2O3/KCL backflush, 8 m CP-Sil 5CB straight) y Configuración RGA 2 (tres canales: 10 m CP-Molesieve 5 Å backflush (RTS), 10 m CP-PoraPLOT U backflush, 10 m CP-AL2O3/KCL backflush-to-detector).

Technical Data

Parameter Value Condition
Channel 1 column 10 m, CP-Molesieve 5 Å, backflush (RTS) p.2 ↗ RGA Configuration 1 and 2
Channel 1 compounds analyzed Permanent gas except CO2 p.2 ↗ RGA Configuration 1 and 2
Channel 2 column 10 m CP-PoraPLOT U, backflush p.2 ↗ RGA Configuration 1 and 2
Channel 2 compounds analyzed CO2, C2H4, C2H6, C2H2, H2S p.2 ↗ RGA Configuration 1 and 2
Channel 3 column 10 m CP-AL2O3/KCL, backflush p.2 ↗ RGA Configuration 1
Channel 3 compounds analyzed C3–C5 paraffins and C3–C5 olefins p.2 ↗ RGA Configuration 1
Channel 4 column 8 m, CP-Sil 5CB, straight p.2 ↗ RGA Configuration 1
Channel 4 compounds analyzed Detailed C6 and C6+ hydrocarbons p.2 ↗ RGA Configuration 1
Carrier gas Argon p.2 ↗ Channel 1, RGA Configuration 1
Injector temperature 110 °C p.2 ↗ all channels
Injection time 40 ms p.2 ↗ all channels
Column head pressure 200 kPa p.2 ↗ Channel 1
Column temperature 80 °C p.2 ↗ Channel 1

Operating Limits

  • Backflush time, Channel 1: 7 seconds p.2 ↗
  • Backflush time, Channel 2: 7.5 seconds p.2 ↗
  • Backflush time, Channel 3 (RGA Configuration 1): 25 seconds p.2 ↗
  • Backflush time, Channel 4 (RGA Configuration 1): NA p.2 ↗
  • Backflush time, Channel 3 (RGA Configuration 2): 4.5 seconds p.2 ↗
  • Invert signal window, Channel 3 (RGA Configuration 2): From 5 to 12 seconds p.2 ↗