Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre el análisis del Método <467> de la USP para disolventes residuales en el sistema Agilent 8890 GC con muestreador de espacio de cabeza 7697A, FIDs duales y columnas J&W DB-Select 624 UI for 467 y J&W HP-INNOWax
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Carrier gas | Helium constant flow mode, 2 mL/min on column p.3 ↗ | |
| Inlet temperature | 140 °C p.3 ↗ | |
| Oven temperature program | 40 °C (hold 5 min) to 180 °C at 18 °C/min °C p.3 ↗ | |
| FID temperature (both channels) | 250 °C p.3 ↗ | |
| Loop equilibration time | 0.05 minutes p.3 ↗ | 7697A headspace sampler |
| Software | Agilent OpenLab CDS – Version 2.2 p.3 ↗ |
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| 10-mL Clear crimp top headspace vials p.2 ↗ | 5190–2285 | Vials | ||
| Advanced Green nonstick inlet septa p.2 ↗ | 5183–4759-100 | Septa | ||
| Inert tee for capillary flow technology p.2 ↗ | G3184-60065 | Splitter | ||
| Short graphite ferrules for 0.1 to 0.32 mm columns, 10/pk p.2 ↗ | 5080-8853 | Ferrules | ||
| UltiMetal Plus flexible metal ferrules for 0.32 mm fused silica tubing, 10/pk p.2 ↗ | G3188-27502 | Ferrules | ||
| 2 mm Splitless straight deactivated inlet liner p.2 ↗ | 5181–8818 | Inlet liner |