Release note purpose and introduction to the software update
Overviewpp. 3-4
New features and changes in this software release
Specp. 5
Supported LC modules and firmware compatibility requirements
Procedurep. 6
Installation instructions and prerequisites for the software
Overviewpp. 8-9
Detailed change history for versions 2.0.14 and 1.1.0
Specpp. 10-12
Required InfinityLab Assist Control Software per LC module
Warningp. 13
Important notes and cautions for users
p. 14
Agilent contact and support information
p. 5table, 2 rows
p. 11table, 11 rows
p. 12table, 17 rows
p. 13table, 4 rows
Supported Operating Systems
Specification
Description Details
Operating system name, version
Windows 10 or 11, Enterprise of Professional, 64-bit
Web browser
Chromium-based browser (Chrome, Edge, Firefox, etc.) with a version higher than 132
Safari-based browser with a version higher than 17.6
Agilent LC – Pumps
Product Number
Description
min. required version
G5654A
Agilent 1260 Infinity II Bio-Inert Pump
1.0
G7104A
Agilent 1290 Infinity II Flexible Pump
1.0
G7104C
Agilent 1260 Infinity II Flexible Pump
1.0
G7110B
Agilent 1260 Infinity II Isocratic Pump
1.0
G7111A
Agilent 1260 Infinity II Quaternary Pump VL
1.0
G7111B
Agilent 1260 Infinity II Quaternary Pump
1.0
G7112B
Agilent 1260 Infinity II Binary Pump
1.0
G7120A
Agilent 1290 Infinity II High-Speed Pump
1.0
G7131A
Agilent 1290 Infinity II Bio Flexible Pump
1.0
G7131C
Agilent 1260 Infinity II Bio Flexible Pump
1.0
G7132A
Agilent 1290 Infinity II Bio High-Speed Pump
1.0
Agilent LC – Detectors
Product Number
Description
min. required version
G7114A
Agilent 1260 Infinity II Variable Wavelength Detector
1.0
G7114B
Agilent 1290 Infinity II Variable Wavelength Detector
1.0
G7115A
Agilent 1260 Infinity II Diode Array Detector WR
1.0
G7117A
Agilent 1290 Infinity II Diode Array Detector FS
1.0
G7117B
Agilent 1290 Infinity II Diode Array Detector
1.0
G7117C
Agilent 1260 Infinity II Diode Array Detector HS
1.0
G7121A
Agilent 1260 Infinity II Fluorescence Detector
1.0
G7121B
Agilent 1260 Infinity II Fluorescence Detector Spectra
1.0
G7162A
Agilent 1260 Infinity II Refractive Index Detector
1.0
G7162B
Agilent 1290 Infinity II Refractive Index Detector
1.0
G7165A
Agilent 1260 Infinity II Multiple Wavelength Detector
1.0
G6160B
Agilent InfinityLab Pro iQ Mass Detector
2.0
G6170A
Agilent InfinityLab Pro iQ Plus Mass Detector
2.0
G7123B
Agilent 1290 Infinity III Fluorescence Detector
2.2 (New)
G6495D
6495D Triple Quadrupole LC/MS
2.2 (New)
G6475A
6475 Triple Quadrupole LC/MS
2.2 (New)
G6465B
Ultivo Triple Quadrupole LC/MS
2.2 (New)
Agilent LC – Clusters
Product
Description
Description min. required version
Agilent Pump Valve Cluster Driver
A combination of one of the following pumps with up to two G1170A and valves 5067-4147 or 5067-4159 Supported pumps: G7111x, G7112x, G5654x, G7104x, G7120x, G7131x, G7132x
1.0
Agilent Valve-Thermostat Cluster Driver
A combination of G7116B, G1170A for combined valve control plus a combination of G7116B and G7130A for combined temperature control. Supports up to 32 columns.
2.2 (New)
Agilent 2D-LC System Driver
Combination of one 2D-capable pump G7120A with up to four G1170A valve drives into a 2D-LC System Driver.
2.1
Agilent 1200 Infinity Series High Dynamic Range DAD Solution
Combines two G7117A/B for high dynamic range DAD applications
2.2 (New)
NOTE Modules which are part of Agilent Pump Valve Cluster and Agilent 2D-LC System are displayed with a cluster icon in status. Task execution for these clusters is not available in this version and will be added in a future release.
Engineering Facts
13 facts · every claim linked to a page
Selection
VTC and HDR cluster configurations require Agilent InfinityLab LC and CE Drivers version 3.11 or later, so driver version must be verified before specifying these clusters.
Prevents selecting cluster hardware that cannot be fully controlled without the correct driver version.
Source
“Full support for Valve-Thermostat Cluster (VTC) and High Dynamic Range (HDR) detector cluster configurations requires Agilent InfinityLab LC and CE Drivers version 3.11 or later.”
Selection
The software requires a 64-bit Windows 10 or 11 Professional/Enterprise PC and a sufficiently recent browser, which must be confirmed before deployment.
Defines the minimum PC platform needed to run the browser-based user interface.
Source
“Windows 10 or 11, Enterprise of Professional, 64-bit”
Selection
The browser used to access the UI must be Chromium-based above version 132 or Safari-based above version 17.6.
An outdated browser will block access to the control software interface.
Source
“Chromium-based browser (Chrome, Edge, Firefox, etc.) with a version higher than 132”
Conformité
The minimum supported Agilent LC and CE Drivers version is 3.7 SR1, so older driver installations are not supported.
Ensures the CDS/driver stack meets the minimum compatibility baseline before installation.
Source
“3.7 SR1. Minimum supported version.”
Selection
OpenLab CDS must be at version 2.6 or higher (including 2.8 FP1, FP2 and Update 9 Client/Server) for browser access support.
Confirms which CDS versions provide the required browser access for the Assist software.
Source
“2.8 FP1, FP2 and Update 9 Client/Server (provides browser access)”
Selection
For Empower integration, the minimum is Empower 3 FR4 with Agilent Driver for Waters Empower 4.1 (Agilent LC Drivers 3.8).
Defines the minimum CDS and driver combination for Empower-based laboratories.
Source
“Min. Empower 3 FR4”
Selection
For Chromeleon integration, the Agilent Driver for Thermo Fisher Chromeleon 3.2 requires Agilent LC Drivers 3.11.
Prevents mismatched driver versions when connecting to a Chromeleon CDS.
Installation and setup must follow the official Agilent InfinityLab LC Series Assist User Manual procedures.
Ensures correct installation, configuration and initial operation per Agilent specifications.
Source
“users are advised to follow the procedures outlined in the official documentation”
Maintenance
The Software Status Bulletin lists known limitations, incompatibilities and fixes/workarounds and should be consulted for this and previous versions.
Known limitations may affect commissioning and require workarounds during setup.
Source
“The Software Status Bulletin lists known limitations, incompatibility and information about available fixes or workarounds for this and previous versions.”
Maintenance
Agilent recommends using the latest driver, firmware and software updates, available from the InfinityLab Assist Updates site.
Keeping firmware and software current avoids known defects and ensures compatibility.
Source
“Agilent continuously improves its drivers, firmware and software and recommends using latest updates.”
Operation
Modules in the Pump Valve Cluster and 2D-LC System are shown with a cluster icon, but task execution for these clusters is not available in this version.
Prevents planning automated task workflows around clusters that cannot yet execute tasks.
Source
“Task execution for these clusters is not available in this version and will be added in a future release.”
Selection
The Valve-Thermostat Cluster supports up to 32 columns, which must be considered when sizing a multi-column switching system.
Defines the maximum column capacity of the VTC configuration for method planning.
Source
“Supports up to 32 columns.”
Selection
The 1290 Infinity III Hybrid Multisampler (G7137B) requires Assist Control Software version 2.0 or higher.
Confirms the minimum software version needed to support this multisampler.