Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model BD 23

Technical Reference Summary

Covered Models & Document Scope

Incubateur standard BINDER BD 23 (série BD, convection naturelle), volume intérieur 20 L. Deux variantes : BD023-230V (Art.-No. 9010-0187, 230 V 1~ 50/60 Hz, fiche CEE 7/7) et BD023UL-120V (Art.-No. 9010-0189, 120 V 1~ 60 Hz, fiche NEMA 5-15).

Technical Data

Parameter Value Condition
Temperature range +5 °C above ambient temperature to 100 °C °C p.2 ↗ BD023-230V
Temperature uniformity at 37°C 0.5 ± K p.2 ↗ BD023-230V
Temperature fluctuation at 37°C 0.2 ± K p.2 ↗ BD023-230V
Heating up time to 37°C 59 min p.2 ↗ BD023-230V
Recovery time after door was opened for 30 s at 37°C 20 min p.2 ↗ BD023-230V
Rated voltage 230 V p.2 ↗ BD023-230V
Power frequency 50/60 Hz p.2 ↗ BD023-230V
Nominal power 0,2 kW p.2 ↗ BD023-230V
Unit fuse 12,5 A p.2 ↗ BD023-230V
Phase (nominal voltage) 1~ p.2 ↗ BD023-230V
Interior volume 20 L p.2 ↗ BD023-230V
Net weight of the unit (empty) 27 kg p.2 ↗ BD023-230V
Load per rack 12 kg p.2 ↗ BD023-230V
Permitted load 25 kg p.2 ↗ BD023-230V
Wall clearance back 100 mm p.2 ↗ BD023-230V
Wall clearance sidewise 50 mm p.2 ↗ BD023-230V
Width net 435 mm p.2 ↗ BD023-230V
Height net 495 mm p.2 ↗ BD023-230V
Depth net 520 mm p.2 ↗ BD023-230V
Interior width 222 mm p.2 ↗ BD023-230V
Interior height 330 mm p.2 ↗ BD023-230V
Interior depth 277 mm p.3 ↗ BD023-230V
Unit doors 1 p.3 ↗ BD023-230V
Energy consumption at 37°C 11 Wh/h p.3 ↗ BD023-230V
Number of shelves (std./max.) 2/4 p.3 ↗ BD023-230V

Operating Limits

  • Temperature range (min): +5 °C above ambient temperature p.2 ↗
  • Temperature range (max): 100 °C p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗
  • Wall clearance sidewise: 50 mm p.2 ↗
  • Permitted load: 25 kg p.2 ↗
  • Load per rack: 12 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ BD Product series: Standard-Incubator with natural convection
3-4 p.1 ↗ 23 Model size (BD 23, 20 L interior volume)
suffix p.1 ↗ UL UL variant for 120 V / 60 Hz mains (NEMA 5-15 plug); absent = standard 230 V / 50/60 Hz (CEE 7/7 plug)
voltage field p.1 ↗ 230V Rated voltage 230 V, 1~, 50/60 Hz
voltage field p.1 ↗ 120V Rated voltage 120 V, 1~, 60 Hz

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
BD023-230V p.1 ↗ 9010-0187 230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Standard model version
BD023UL-120V p.1 ↗ 9010-0189 120 V 1~ 60 Hz, NEMA 5-15 plug, Standard UL model version