Data Sheet Model CB S 260
Technical Reference Summary
Covered Models & Document Scope
Incubateurs à CO2 modèle CB-S 260 avec stérilisation à air chaud. Deux variantes : CBS260-230V (Art.-No. 9040-0223, 200...230 V 1~ 50/60 Hz, fiche CEE 7/7) et CBS260UL-120V (Art.-No. 9040-0224, 100...120 V 1~ 50/60 Hz, fiche NEMA 5-20).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Interior volume | 267 L p.1 ↗ | CBS260-230V / CBS260UL-120V |
| Rated voltage | 200...230 V p.2 ↗ | CBS260-230V |
| Rated voltage | 100...120 V p.2 ↗ | CBS260UL-120V |
| Power frequency | 50/60 Hz p.2 ↗ | both variants |
| Phase (nominal voltage) | 1~ p.2 ↗ | both variants |
| Nominal power | 1,5 kW p.2 ↗ | both variants |
| Unit fuse | 10,0 A p.2 ↗ | CBS260-230V |
| Unit fuse | 16,0 A p.2 ↗ | CBS260UL-120V |
| Temperature range | +6 °C above ambient temperature to 50 °C p.2 ↗ | both variants |
| Temperature uniformity at 37°C | 0.4 ± K p.2 ↗ | both variants |
| Temperature fluctuation at 37°C | 0.1 ± K p.2 ↗ | both variants |
| Recovery time after door was opened | 8 min p.2 ↗ | both variants |
| Humidity range | 90 ...95 % RH p.2 ↗ | both variants |
| CO₂ range | 0 ...20 Vol.-% CO₂ p.2 ↗ | both variants |
| CO₂ measuring technology | IR p.2 ↗ | both variants |
| Net weight of the unit (empty) | 116 kg p.2 ↗ | both variants |
| Load per rack | 10 kg p.2 ↗ | both variants |
| Permitted load | 40 kg p.2 ↗ | both variants |
| Wall clearance back | 100 mm p.2 ↗ | both variants |
| Wall clearance sidewise | 50 mm p.2 ↗ | both variants |
| Width net | 740 mm p.2 ↗ | both variants |
| Height net | 1,020 mm p.3 ↗ | both variants |
| Depth net | 785 mm p.3 ↗ | both variants |
| Sound-pressure level | 41 dB(A) p.3 ↗ | both variants |
| Energy consumption at 37°C | 55 Wh/h p.3 ↗ | both variants |
Operating Limits
- Temperature range: +6 °C above ambient temperature to 50 °C p.2 ↗
- Humidity range: 90 ...95 % RH p.2 ↗
- CO₂ range: 0 ...20 Vol.-% CO₂ p.2 ↗
- Permitted load: 40 kg p.2 ↗
- Load per rack: 10 kg p.2 ↗
- Wall clearance back: 100 mm p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-6 p.1 ↗ | CBS260 | Model series CB-S, size 260 |
| suffix p.1 ↗ | UL | UL variant with 100...120 V power supply and NEMA 5-20 plug (absent = standard 200...230 V, CEE 7/7) |
| suffix p.1 ↗ | 230V | Standard mains voltage variant 200...230 V 1~ 50/60 Hz, unit fuse 10,0 A |
| suffix p.1 ↗ | 120V | UL mains voltage variant 100...120 V 1~ 50/60 Hz, unit fuse 16,0 A |
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| CBS260-230V p.1 ↗ | 9040-0223 | 200...230 V 1~ 50/60 Hz, 10,0 A, CEE 7/7 plug, Standard | Standard model version | |
| CBS260UL-120V p.1 ↗ | 9040-0224 | 100...120 V 1~ 50/60 Hz, 16,0 A, NEMA 5-20 plug, Standard | UL model version |