Data Sheet Model FD 115
Technical Reference Summary
Covered Models & Document Scope
Chambre de séchage et de chauffage modèle FD 115 à convection forcée (Avantgarde.Line). Deux variantes standard : FD115-230V (Art.-No. 9010-0305, 230 V 1~ 50/60 Hz, fiche CEE 7/7) et FD115UL-120V (Art.-No. 9010-0306, 120 V 1~ 60 Hz, fiche NEMA 5-15).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | +10 °C above ambient temperature to 300 °C p.2 ↗ | FD115-230V |
| Temperature uniformity at 150°C | 1.7 ± K p.2 ↗ | FD115-230V |
| Temperature fluctuation at 150°C | 0.3 ± K p.2 ↗ | FD115-230V |
| Heating up time to 150°C | 18 min p.2 ↗ | FD115-230V |
| Recovery time after door was opened for 30 s at 150°C | 4 min p.2 ↗ | FD115-230V |
| Air change rate at 100°C | 32 x/h p.2 ↗ | FD115-230V |
| Rated voltage | 230 V p.2 ↗ | FD115-230V |
| Power frequency | 50/60 Hz p.2 ↗ | FD115-230V |
| Nominal power | 1,3 kW p.2 ↗ | FD115-230V |
| Unit fuse | 6,3 A p.2 ↗ | FD115-230V |
| Phase (nominal voltage) | 1~ p.2 ↗ | FD115-230V |
| Interior volume | 116 L p.2 ↗ | FD115-230V |
| Net weight of the unit (empty) | 54 kg p.2 ↗ | FD115-230V |
| Load per rack | 30 kg p.2 ↗ | FD115-230V |
| Permitted load | 150 kg p.2 ↗ | FD115-230V |
| Wall clearance back | 160 mm p.2 ↗ | FD115-230V |
| Wall clearance sidewise | 100 mm p.2 ↗ | FD115-230V |
| Width net | 710 mm p.2 ↗ | FD115-230V |
| Height net | 735 mm p.2 ↗ | FD115-230V |
| Depth net | 605 mm p.2 ↗ | FD115-230V |
| Interior width | 550 mm p.3 ↗ | FD115-230V |
| Interior height | 550 mm p.3 ↗ | FD115-230V |
| Interior depth | 385 mm p.3 ↗ | FD115-230V |
| Sound-pressure level | 43 dB(A) p.3 ↗ | FD115-230V |
| Number of shelves (std./max.) | 2/5 p.3 ↗ | FD115-230V |
Operating Limits
- Temperature range: +10 °C above ambient temperature to 300 °C p.2 ↗
- Permitted load: 150 kg p.2 ↗
- Load per rack: 30 kg p.2 ↗
- Wall clearance back: 160 mm p.2 ↗
- Wall clearance sidewise: 100 mm p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-2 p.1 ↗ | FD | Series: drying and heating chamber with forced convection (Avantgarde.Line) |
| 3-5 p.1 ↗ | 115 | Model size / interior volume class (116 L) |
| 6-8 p.1 ↗ | UL | UL variant for 120 V / 60 Hz mains (absent = standard 230 V variant) |
| 9-12 p.1 ↗ | 230V / 120V | Rated voltage variant: 230 V (50/60 Hz) or 120 V (60 Hz) |
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| FD115-230V p.1 ↗ | 9010-0305 | Standard, 230 V 1~ 50/60 Hz, CEE 7/7 plug | ||
| FD115UL-120V p.1 ↗ | 9010-0306 | Standard, 120 V 1~ 60 Hz, NEMA 5-15 plug |