Data Sheet Model FD 720
Technical Reference Summary
Covered Models & Document Scope
Cámara de secado y calentamiento modelo FD 720 con convección forzada, serie Avantgarde.Line; versión estándar FD720-400V, Art.-No. 9010-0311, 400 V 3~/N/PE 50/60 Hz, enchufe CEE 16/6 H de 5 polos; el documento también enumera accesorios y servicios.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Article number | 9010-0311 p.2 ↗ | FD720-400V, standard |
| Designation | FD720-400V p.2 ↗ | standard |
| Temperature range | +12 °C above ambient temperature to 300 °C p.2 ↗ | FD720-400V |
| Temperature uniformity at 150°C | 2.5 ± K p.2 ↗ | FD720-400V |
| Temperature fluctuation at 150°C | 0.6 ± K p.2 ↗ | FD720-400V |
| Heating up time to 150°C | 25 min p.2 ↗ | FD720-400V |
| Recovery time after door was opened for 30 s at 150°C | 6 min p.2 ↗ | FD720-400V |
| Air change rate at 100°C | 5 x/h p.2 ↗ | FD720-400V |
| Rated voltage | 400 V p.2 ↗ | FD720-400V |
| Power frequency | 50/60 Hz p.2 ↗ | FD720-400V |
| Nominal power | 4,5 kW p.2 ↗ | FD720-400V |
| Unit fuse | 16,0 A p.2 ↗ | FD720-400V |
| Phase (nominal voltage) | 3~/N/PE p.2 ↗ | FD720-400V |
| Interior volume | 741 L p.2 ↗ | FD720-400V |
| Net weight of the unit (empty) | 166 kg p.2 ↗ | FD720-400V |
| Load per rack | 45 kg p.2 ↗ | FD720-400V |
| Permitted load | 315 kg p.2 ↗ | FD720-400V |
| Wall clearance back | 160 mm p.2 ↗ | FD720-400V |
| Wall clearance sidewise | 100 mm p.2 ↗ | FD720-400V |
| Width net | 1,165 mm p.2 ↗ | FD720-400V, housing dimensions not incl. fittings and connections |
| Height net | 1,590 mm p.2 ↗ | FD720-400V, housing dimensions not incl. fittings and connections |
| Depth net | 870 mm p.2 ↗ | FD720-400V, housing dimensions not incl. fittings and connections |
Operating Limits
- Temperature range: +12 °C above ambient temperature to 300 °C p.2 ↗
- Permitted load: 315 kg p.2 ↗
- Load per rack: 45 kg p.2 ↗
- Wall clearance back: 160 mm p.2 ↗
- Wall clearance sidewise: 100 mm p.2 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| FD720-400V p.1 ↗ | 9010-0311 | Standard, 400 V 3~/N/PE 50/60 Hz, 16,0 A, CEE 16/6 H plug 5-pin | Main unit |