Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model KT 53

Technical Reference Summary

Covered Models & Document Scope

Ficha técnica del modelo KT 53 incubador refrigerado con tecnología Peltier, volumen interior 53 L. Dos variantes: KT053-230V (Art.-No. 9020-0311, 200...240 V 1~ 50/60 Hz, enchufe CEE 7/7) y KT053UL-120V (Art.-No. 9020-0312, 100...120 V 1~ 50/60 Hz, enchufe NEMA 5-20).

Technical Data

Parameter Value Condition
Temperature range 4...100 °C p.2 ↗ KT053-230V / KT053UL-120V
Temperature uniformity at 37°C 0.3 ± K p.2 ↗ KT053-230V / KT053UL-120V
Temperature fluctuation at 37°C 0.1 ± K p.2 ↗ KT053-230V / KT053UL-120V
Max. heat compensation at 40°C 100 W p.2 ↗ KT053-230V / KT053UL-120V
Recovery time after door was opened for 30 s at 37°C 2 min p.2 ↗ KT053-230V / KT053UL-120V
Rated Voltage 200...240 V p.2 ↗ KT053-230V
Rated Voltage 100...120 V p.2 ↗ KT053UL-120V
Power frequency 50/60 Hz p.2 ↗ KT053-230V / KT053UL-120V
Nominal power 0,4 kW p.2 ↗ KT053-230V / KT053UL-120V
Unit fuse 10,0 A p.2 ↗ KT053-230V / KT053UL-120V
Phase (Nominal voltage) 1~ p.2 ↗ KT053-230V / KT053UL-120V
Interior volume 53 L p.2 ↗ KT053-230V / KT053UL-120V
Net weight of the unit (empty) 63 kg p.2 ↗ KT053-230V / KT053UL-120V
Load per rack 15 kg p.2 ↗ KT053-230V / KT053UL-120V
Permitted load 40 kg p.2 ↗ KT053-230V / KT053UL-120V
Wall clearance back 100 mm p.2 ↗ KT053-230V / KT053UL-120V
Wall clearance sidewise 240 mm p.2 ↗ KT053-230V / KT053UL-120V
Width net 660 mm p.2 ↗ KT053-230V / KT053UL-120V
Height net 635 mm p.2 ↗ KT053-230V / KT053UL-120V
Depth net 630 mm p.2 ↗ KT053-230V / KT053UL-120V
Interior width 400 mm p.2 ↗ KT053-230V / KT053UL-120V
Interior height 400 mm p.2 ↗ KT053-230V / KT053UL-120V
Interior depth 334 mm p.3 ↗ KT053-230V / KT053UL-120V
Sound-pressure level 48 dB(A) p.3 ↗ KT053-230V / KT053UL-120V
Energy consumption at 25°C 75 Wh/h p.3 ↗ KT053-230V / KT053UL-120V

Operating Limits

  • Temperature range: 4...100 °C p.2 ↗
  • Max. heat compensation at 40°C: 100 W p.2 ↗
  • Permitted load: 40 kg p.2 ↗
  • Load per rack: 15 kg p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗
  • Wall clearance sidewise: 240 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ KT Product series: cooling incubator with Peltier technology
3-5 p.1 ↗ 053 Model size / interior volume 53 L
suffix p.1 ↗ -230V Mains voltage variant 200...240 V 1~ 50/60 Hz, CEE 7/7 plug
suffix p.1 ↗ UL-120V Mains voltage variant 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
KT053-230V p.1 ↗ 9020-0311 200...240 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Standard model version
KT053UL-120V p.1 ↗ 9020-0312 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug, Standard Standard model version