Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model MKF CO₂ 720

Technical Reference Summary

Covered Models & Document Scope

Cámara climática dinámica de CO₂ modelo MKF 720 con control de humedad. Dos variantes de alimentación: MKF720CO2-400V, Art.-No. 9020-0523 (400 V 3~/N/PE 50 Hz) y MKF720CO2-400V, Art.-No. 9020-0525 (480 V 3~/N/PE 60 Hz). Ambas en la versión de modelo Standard.

Technical Data

Parameter Value Condition
Temperature range -48...180 °C p.3 ↗ MKF720CO2-400V, 50 Hz
Temperature range -48...180 °C p.3 ↗ MKF720CO2-400V, 60 Hz
Temperature uniformity dependent on set 0.1...1.0 ± K p.3 ↗ MKF720CO2-400V, 50 Hz
Temperature fluctuation dependent on set 0.1...0.2 ± K p.3 ↗ MKF720CO2-400V, 50 Hz
Cooling down time from 180°C to -40°C 60 min p.3 ↗ MKF720CO2-400V, 50 Hz
Temperature range with humidity 10...95 °C p.3 ↗ MKF720CO2-400V, 50 Hz
Humidity range 10...98 % RH p.3 ↗ MKF720CO2-400V, 50 Hz
Max. heat compensation at 25°C and 90% RH 1,100 W p.3 ↗ MKF720CO2-400V, 50 Hz
Rated voltage 400 V p.3 ↗ MKF720CO2-400V, 50 Hz
Rated voltage 480 V p.3 ↗ MKF720CO2-400V, 60 Hz
Power frequency 50 Hz p.3 ↗ MKF720CO2-400V, 50 Hz
Power frequency 60 Hz p.3 ↗ MKF720CO2-400V, 60 Hz
Nominal power 11 kW p.3 ↗ MKF720CO2-400V, 50 Hz
Unit fuse 25,0 A p.3 ↗ MKF720CO2-400V, 50 Hz
Phase (nominal voltage) 3~/N/PE p.3 ↗ MKF720CO2-400V, 50 Hz
Interior volume 734 L p.3 ↗ MKF720CO2-400V, 50 Hz
Net weight of the unit (empty) 518 kg p.3 ↗ MKF720CO2-400V, 50 Hz
Load per rack 40 kg p.3 ↗ MKF720CO2-400V, 50 Hz
Permitted load 160 kg p.3 ↗ MKF720CO2-400V, 50 Hz
Width net 1,580 mm p.4 ↗ MKF720CO2-400V, 50 Hz
Height net 2,005 mm p.4 ↗ MKF720CO2-400V, 50 Hz
Depth net 1,140 mm p.4 ↗ MKF720CO2-400V, 50 Hz
Sound-pressure level 60 dB(A) p.4 ↗ MKF720CO2-400V, 50 Hz
Energy consumption at 25°C and 60% RH 3,400 Wh/h p.4 ↗ MKF720CO2-400V, 50 Hz
Number of shelves (std./max.) 1/11 p.4 ↗ MKF720CO2-400V, 50 Hz

Operating Limits

  • Temperature range: -48...180 °C p.3 ↗
  • Temperature range with humidity: 10...95 °C p.3 ↗
  • Humidity range: 10...98 % RH p.3 ↗
  • Permitted load: 160 kg p.3 ↗
  • Load per rack: 40 kg p.3 ↗
  • Wall clearance back: 300 mm p.4 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ MKF Series: dynamic climate chamber with humidity control
4-6 p.1 ↗ 720 Model size (720)
7-9 p.1 ↗ CO2 Refrigerant: natural CO₂ (R744)
10-14 p.2 ↗ 400V Rated voltage variant: 400 V (50 Hz) or 480 V (60 Hz) depending on article number

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
MKF720CO2-400V p.2 ↗ 9020-0523 400 V 3~/N/PE 50 Hz, Standard Mains variant 50 Hz
MKF720CO2-400V p.2 ↗ 9020-0525 480 V 3~/N/PE 60 Hz, Standard Mains variant 60 Hz

Modèles liés