Agilent Protein Broad Range P240 Kit
Technical Reference Summary
Covered Models & Document Scope
Kit Agilent Protein Broad Range P240 (n.º de pieza: 5191-6640) para su uso en el sistema Agilent ProteoAnalyzer; especificaciones analíticas y físicas solo para el modo de marcador LM y para los modos de marcador LM y UM.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Sizing range | 10 to 240 kDa p.3 ↗ | LM only |
| Typical sizing accuracy (% sizing error) | < 15 % p.3 ↗ | LM only, BSA, CAII (using reduced conditions) |
| Typical resolution | < 10 % molecular weight resolution p.3 ↗ | between 15 to 150 kDa (based on ladder) |
| Sizing precision | < 8 % CV p.3 ↗ | LM only, BSA, CAII, GREMLIN-1, and NIST mAb (using reduced conditions) |
| Quantitative range | 2 to 2,000 ng/µL p.3 ↗ | BSA in PBS |
| Sensitivity (signal/noise > 3) | 1 ng/µL p.3 ↗ | BSA, CAII in PBS |
| Quantification reproducibility | < 15 % CV p.3 ↗ | 20 – 2,000 ng/µL BSA |
| Total run time | 30 minutes p.3 ↗ | |
| Samples per run | 11 samples + ladder in well 12 p.3 ↗ | |
| Sample volume required | 1 µL p.3 ↗ | |
| Kit stability | Minimum 4 months p.3 ↗ | |
| Human liver crude cell lysate concentration | 2,000 ng/µL p.3 ↗ |
Operating Limits
- Sizing range (LM only): 10 to 240 kDa p.3 ↗
- Sizing range (LM and UM): 10 to 200 kDa p.3 ↗
- Quantitative range: 2 ng/µL to 2,000 ng/µL for BSA in PBS p.3 ↗
- Sensitivity (signal/noise > 3): 1 ng/µL for BSA, CAII in PBS p.3 ↗
- Kit stability: Minimum 4 months p.3 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| Agilent Protein Broad Range P240 kit p.1 ↗ | 5191-6640 | Kit for Agilent ProteoAnalyzer system |