Agilent NovoCyte Opteon Spectral Flow Cytometer Specifications
Technical Reference Summary
Covered Models & Document Scope
Citómetro de flujo espectral Agilent NovoCyte Opteon; configuraciones con cuatro láseres (349 nm 20 mW, 405 nm 130 mW, 488 nm 100 mW, 561 nm 100 mW, 637 nm 120 mW) y 8, 11, 14, 18 o 19 detectores; configuraciones personalizadas disponibles bajo pedido
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Laser type | Solid state laser with onboard thermal-electric cooling p.1 ↗ | |
| Laser beam configuration | Spacially separated; fixed laser delay and no need for adjustment p.1 ↗ | |
| Photodetector | Avalanche photodiode (APD) p.1 ↗ | |
| Optical alignment | Fixed. No operator alignment required. p.1 ↗ | |
| Forward scatter detection laser | 488 nm p.1 ↗ | FSC |
| Side scatter detection laser | BSSC 488 nm; VSSC 405 nm nm p.1 ↗ | SSC |
| FSC sensitivity | 0.4 µm p.1 ↗ | |
| VSSC sensitivity | 0.08 µm p.1 ↗ | |
| BSSC sensitivity | 0.2 µm p.1 ↗ | |
| FITC fluorescence threshold sensitivity | ≤ 50 MESF p.1 ↗ | B543 channel |
| PE fluorescence threshold sensitivity | ≤ 10 MESF p.1 ↗ | Y582 channel |
| APC fluorescence threshold sensitivity | ≤ 10 MESF p.1 ↗ | R661 channel |
| Pacific Blue fluorescence threshold sensitivity | ≤ 30 MESF p.1 ↗ | V455 channel |
| Fluorescence resolution | < 3 % CV p.1 ↗ | CEN |
| Flow cell dimensions | 170 x 490 µm p.2 ↗ | rectangular quartz flow cell |
| Sample acquisition rate | 100,000 events/second p.2 ↗ | |
| Volumetric absolute count precision | ≤ 5 % p.2 ↗ | |
| Sample flow rate | 5 to 120 µL/min p.2 ↗ | 1 µL increments |
| Sample acquisition volume | 1 µL to 5 mL p.2 ↗ | |
| Sheath fluid container capacity | 6 L p.2 ↗ | |
| Waste fluid container capacity | 6 L p.2 ↗ | |
| Cleaning fluid container capacity | 500 mL p.2 ↗ | |
| Decontamination fluid container capacity | 500 mL p.2 ↗ | automatic liquid volume sensing |
| Number of detectors | 8, 11, 14, 18, 19 p.3 ↗ | four-laser configuration |
Operating Limits
- Sample flow rate range: 5 to 120 µL/min p.2 ↗
- Sample acquisition volume range: 1 µL to 5 mL p.2 ↗
- Volumetric absolute count precision: ≤ 5% p.2 ↗