Induztr
Résumé technique Visionneuse de documents Modèles associés

Fermentors Bioreactors Brochure BioFlo 320 plus Every Move Seamlessly

Technical Reference Summary

Covered Models & Document Scope

Sistema de biorreactor universal de banco BioFlo® 320 plus; vasos de biorreactor de vidrio de 1 L, 3 L, 5 L, 10 L (volúmenes totales 2.5 L, 5.0 L, 7.5 L, 14.0 L); intercambiables con biorreactores de un solo uso BioBLU®; rango de volumen de trabajo total 420 mL – 40 L en una única plataforma de control; aplicaciones de cultivo celular y microbianas; variantes de software DASware® control y DASware control plus.

Technical Data

Parameter Value Condition
Working volume 0.6 – 1.9 L p.6 ↗ glass bioreactor 1 L
Working volume 1.3 – 3.8 L p.6 ↗ glass bioreactor 3 L
Working volume 1.9 – 5.6 L p.6 ↗ glass bioreactor 5 L
Working volume 3.5 – 10.5 L p.6 ↗ glass bioreactor 10 L
Total volume 2.5 L p.6 ↗ glass bioreactor 1 L
Total volume 5.0 L p.6 ↗ glass bioreactor 3 L
Total volume 7.5 L p.6 ↗ glass bioreactor 5 L
Total volume 14.0 L p.6 ↗ glass bioreactor 10 L
Working volume range (platform) 420 mL – 40 L p.3 ↗ single control platform
Electrical supply 100 – 120/208 – 240 VAC p.10 ↗ 50/60 Hz, IEC with regional plug types
Water supply pressure 10 psig p.10 ↗ stainless steel quick-connect
Water supply pressure 0.69 barg p.10 ↗ stainless steel quick-connect
pH control range 2 – 12 p.10 ↗ analog or digital Mettler Toledo ISM
Optical pH control range 6 – 8 p.10 ↗ digital (Presens)
DO control range 0 – 200 % p.10 ↗ analog or digital Mettler Toledo ISM
Optical DO control range 0 – 200 % p.10 ↗ digital Mettler Toledo ISM
Redox control range (-)2000 mV – (+)2000 mV mV p.10 ↗ analog or digital Mettler Toledo ISM
Pump speed 0.1 – 25 rpm p.10 ↗ Pumps 1, 2 & 3, Watson-Marlow 114DVP
Pump speed 0.1 – 150 rpm p.10 ↗ Pump 4, Watson-Marlow 314DV
Pump speed 0.1 – 125 rpm p.10 ↗ Pump 4 dual pump head, Watson-Marlow 314X
Pump speed 0.1 – 220 rpm p.10 ↗ External pumps 1 & 2, Watson-Marlow 120U/DV
Number of parallel units up to eight systems p.3 ↗ multi-unit control via Ethernet
Mass flow controllers up to 5 p.3 ↗ configurable TMFC for overlay and sparge, field upgradable
Integrated pumps up to six p.3 ↗ variable speed

Operating Limits

  • Electrical supply: 100 – 120/208 – 240 VAC, 50/60 Hz p.10 ↗
  • Water supply pressure: 10 psig (0.69 barg) p.10 ↗
  • pH control range: 2 – 12 p.10 ↗
  • DO control range: 0 – 200 % p.10 ↗
  • Redox control range: (-)2000 mV – (+)2000 mV p.10 ↗
  • Max number of controlled systems: eight p.3 ↗