Induztr
Résumé technique Visionneuse de documents Modèles associés

Magnis and MagnisDx NGS Prep System User Guide

Sommaire

  1. Overview pp. 1-8
    Table of contents, legal notices, symbols, and product description
  2. Spec pp. 9-16
    Instrument specifications, provided materials, cleaning supplies, environmental requirements
  3. Overview pp. 17-21
    Instrument components, back panel, status indicator lights
  4. Procedure pp. 22-33
    Power on, user account management, system settings configuration
  5. Procedure pp. 34-48
    Running protocols, diagnostic tests, UV decontamination, auto-teaching, updates
  6. Procedure pp. 49-57
    Preventative maintenance, cleaning components, UV tube replacement, disposal
  7. Overview pp. 58-90
    Screens for login, settings, protocols, diagnostics, run data
  8. Procedure pp. 91-94
    Troubleshooting suggestions for labware and library issues
  9. pp. 95-96
    Document revision history and part number
  10. p. 9 table, 29 rows
  11. p. 25 table, 1 rows
  12. p. 95 table, 2 rows
  13. p. 11 table, 3 rows
  14. p. 21 table, 4 rows
  15. p. 24 table, 1 rows
  16. p. 26 table, 11 rows
  17. p. 59 table, 8 rows
  18. p. 63 table, 7 rows
  19. p. 67 table, 5 rows

Table 1 Technical specifications of the instrument

System ComponentSpecification
Thermal cycler module
Minimum temperature of thermal block4°C (39.2°F)
Maximum temperature of thermal block99°C (210.2°F)
Heater/Shaker/Magnet module
Heater
Temperature maximum75°C (167°F)
Shaker
Maximum speed (rpm)1800 rpm ±5%
Chiller module
Temperature range4–12°C (39.2–53.6°F)
Power inlet
AC voltage100–240 VAC ± 10%
AC frequency50/60 Hz
Maximum power1000 W
Input/output (I/O) connectors
USB 2.0 port
Maximum voltage rating5 VDC
LAN port (for Cat 5 cable‡)
Maximum voltage rating3.3 VDC
System
ISM ClassificationISM Group 1 Class A According to CISPR 11
Acoustic sound pressure 70 dBA
Dimensions, instrument door closed (Height × Depth × Width)71 cm × 72 cm × 62 cm (28 inches × 28 inches × 24 inches)
Dimensions, instrument door open (Height × Depth × Width)107 cm × 72 cm × 62 cm (42 inches × 28 inches × 24 inches)
Weight95 kg (209 lbs)
Environmental conditions*
TemperatureOperating: 15°C to 25°C (53.6°F to 77°F) Transport and storage: –40°C to 70°C (–40°F to 158°F)
HumidityOperating: 30% to 70%, non-condensing
Altitude2000 m (6562 feet)

* The listed conditions apply to instrument operation. Conditions required for assay performance may differ. ‡The maximum length of the LAN cable used for EMC testing is 1.5 meters.

Table 5 Preparatory activities (continued)

StepDescription

На странице показана только шапка таблицы (Step, Description); строки данных на данной странице отсутствуют.

Revision Log

RevisionChange
E.01• Additional details added to the Warning statement on moving the instrument with a forklift or lift table. • Added 10% allowance to AC voltage specification.
F.00• New troubleshooting guidance on library collection error. • Updated firmware screen images for version 1.5.1. • Symbol usage updated per EN ISO 15223-1:2021/A1.

Table 3 Recommended supplies for cleaning the system instrument

DescriptionPurposeVendor
Diluted bleach (10%) wipesSurface-cleaning of the instrument deckHype-Wipe Bleach Towelettes, or equivalent
Alcohol (70%) wipesSurface-cleaning of the instrument exterior and instrument deckVWR Pre-Moistened Clean Wipes, or equivalent
Dry, lint-free, scratch-free, laboratory wipesSurface-cleaning of the barcode scanner windowKimwipes, or equivalent

Table 4 Colors and descriptions of the LED indicator lights in the plate filling area

LED ColorInstrument StatusDescription
WhiteReadyThe lights are illuminated white when the system is idle but the door is open, when the system is performing auto-teaching or teach-point verification, and when a user is setting up a protocol run.
BlueReadyThe lights are illuminated blue anytime the system is idle and the door is closed, including at completion of a protocol run. The lights are also blue when the system is performing diagnostic tests.
GreenRunningThe lights are illuminated green when the system is running a protocol.
RedErrorThe lights are illuminated red when the system has encountered an error. Check the touchscreen for an error message containing further details.

WARNING: During UV decontamination, the indicator lights are off and the instrument deck is illuminated by the UV light. Do not look directly at the UV light.

Table 5 Preparatory activities

StepDescription
Hardware initializationDuring hardware initialization, the system returns all motorized parts (i.e., the gantry, HSM module, and thermal cycler) back to their home positions. Additionally, the system checks for the presence of any tips on the micropipettor and, if necessary, disposes of those tips in the tip waste bin.

Table 6

ActionPermitted by Standard users?Permitted by Advanced users?
Set up and run protocolsYesYes
View data from their own protocol runsYesYes
Update the number of PCR cycles during run setupNoYes
View data from protocol runs of other usersNoYes
Edit user accounts of other usersNoYes
Edit chiller temperature settingNoYes
Install firmware updatesNoYes
Install protocol updatesNoYes
Delete diagnostic reportsNoYes
Change default protocol versionNoYes
Bypass reagent expiration errorNoYes*

* Only permissible on the G9710A Magnis NGS Prep System.

Table 7 Descriptions of the top and bottom banner elements

ElementDescription
Top banner
Instrument statusDisplays the status of the instrument (Ready, Running, or Error). See “Instrument Status Indicator Lights” on page 21 for more information on the possible instrument statuses.
Bottom banner
Home iconProvides access to the Home screen. Press to exit the current screen and navigate to the Home screen.
User nameDisplays the user name of the currently logged-in user. Press to access a button for logging out.
Door statusIndicates the position and lock status of the instrument door (Door Opened, Door Closed, or Door Unlocked).
Error iconIndicates the occurrence of a system error or failed diagnostic test. Press to view information on the error(s).
Time and dateDisplays the time and date according to the system settings. Press to open the Date & Time Settings screen.

Table 8 Settings screen buttons

ButtonDescription
User ManagementOpens the “User Management screen” for viewing user accounts and accessing tools for adding and editing user accounts.
System SettingsOpens the “System Settings screen” for accessing tools to configure system-wide settings.
ExportOpens the “Export Files screen” for exporting post-run data files and log files.
Protocol UpdateOpens the “Protocol Update screen” for installing new protocol files on the system.
Auto TeachingOpens the “Auto Teach screen” for running auto-teaching.
Self DiagnosticOpens the “Instrument Diagnostic screen” for running diagnostic tests and viewing reports for diagnostic tests and instrument health checks.
Hardware Usage TrackingOpens the “Hardware Usage Tracking screen” for viewing a usage counter for the UV-tube.

Table 9 System Settings screen buttons

ButtonDescription
Instrument SettingsOpens the “Instrument Settings screen” for viewing and/or setting the instrument name, serial number, firmware software version, and device version.
Date & Time SettingsOpens the “Date & Time Settings screen” for setting the date and time of the system.
Chiller SettingOpens the “Chiller Setting screen” for setting the temperature of the chiller module.
Other SettingsOpens the “Other Settings screen” which has a setting to verify teach points during the IHC.
Firmware UpdateOpens the “Firmware Update screen” for installing a new version of the firmware software from a connected USB drive.

Engineering Facts

14 facts · every claim linked to a page
Installation

The instrument must be installed by an Agilent engineer or authorized service provider, not by the end user.

Installation labor and scheduling must be planned with Agilent before delivery.

Source
“The system must be installed by an Agilent engineer or Agilent authorized service provider.”
Installation

Moving the instrument requires a forklift or lift table rated for at least 100 kg plus two people for final positioning.

Site access and rigging equipment must be arranged for the 95 kg unit.

Source
“To move the instrument, use an automated forklift or lift table that can lift at least 100 kg (220 lbs).”
Installation

Clearance of 10 cm on each side and 18 cm at the back must be maintained for ventilation.

Bench layout must reserve these clearances to avoid overheating.

Source
“Maintain 10 cm (4 inches) of cleared space on each side of the instrument and 18 cm (7 inches) of cleared space at the back of the instrument.”
Installation

The instrument needs a grounded, dedicated or independent AC circuit and must not share a circuit with high-current devices.

Circuit planning must account for the 1000 W load and avoid nuisance trips.

Source
“If possible, connect the instrument to an independent or dedicated AC circuit.”
Installation

Only the Agilent-supplied regional power cord may be used; substitutes are prohibited.

The correct regional cord must be ordered and kept with the unit.

Source
“Do not substitute with a power cord from another source.”
Operation

The instrument must be placed in a post-PCR laboratory environment.

Room assignment must follow the pre-/post-PCR workflow zoning.

Source
“Place the instrument in a post-PCR laboratory environment.”
Safety

Electrostatic discharges above 8000 V can disrupt the USB ports, so anti-static precautions are required in high-static areas.

ESD controls (wrist strap, flooring) must be planned for the installation site.

Source
“Electrostatic discharges greater than 8000 volts may interfere with the normal operation of the USB ports on the instrument.”
Operation

USB ports must not be used while a protocol is running and cannot accept encrypted drives.

File transfer procedures must be scheduled outside of runs.

Source
“Do not use USB ports while a Magnis protocol is running.”
Operation

Prepared libraries are held at 12 °C for up to 72 hours and must be collected within that window.

Workflow scheduling must guarantee sample retrieval within 72 hours.

Source
“which is held at 12°C for up to 72 hours. Collect the samples within that 72-hour period.”
Operation

The chiller stays at its set temperature for 2 hours after library-ready notification before shutting off automatically.

Operators must retrieve libraries within the 2-hour chiller countdown.

Source
“The chiller remains at its set temperature (12°C, by default) for 2 hours before automatically shutting off.”
Maintenance

No calibration is required, but annual preventive maintenance by Agilent is mandatory.

Service contracts and annual downtime must be budgeted.

Source
“The Magnis/MagnisDx NGS Prep System does not require calibration, however, it does require preventive maintenance annually.”
Maintenance

The barcode scanner window should be cleaned only with a soft cloth and water or mild detergent, never touched otherwise.

Cleaning SOPs must protect the scanner window from damage.

Source
“Do not touch the window with anything other than the soft cloth or wipe that you are using for cleaning.”
Performance

Operating humidity outside 30–70% non-condensing can degrade instrument performance.

HVAC requirements must be verified for the lab room.

Source
“Operating the system in humidity levels outside of this range can impact performance.”
Performance

Vibration-generating or vibration-sensitive equipment nearby can impact performance.

Bench placement must avoid adjacent centrifuges and similar equipment.

Source
“Proximity to lab equipment that generates vibrations can impact performance.”