Induztr
Résumé technique Visionneuse de documents Modèles associés

CC-110A Immersion Circulator — Technical Data Sheet

Sommaire

  1. Overview p. 1
    Product identification and general description of the CC-110A unit
  2. Spec p. 2
    Technical data per DIN 12876 including temperature and performance ratings
  3. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
Order-No.CC-110A
Operating temperature range25...100 °C
with water cooling20...100 °C
with refrigerator15...100 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pump max. delivery27 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connenction (optional)M16x1 male
Bath volume7,5 l
min. filling capacity5,2 l
Width bath opening WxD/ bath depth130x310/ 150 mm
Overall dimensions WxDxH **147x507x330 mm
Net weight6 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method, so the choice of water cooling or refrigerator must be matched to the required process temperature.

Determines whether the bath can reach the required low temperatures without an external chiller or refrigerator.

Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Installation

The unit requires a 200-240 V single- or two-phase supply with a minimum 10 A fuse and maximum 16 A fuse, so the branch circuit must be sized accordingly.

Electrical supply and protection must be planned before installation to avoid nuisance tripping or undersized wiring.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

External branch circuit protection according to UL 489 is required, so a suitable UL-listed breaker must be provided in the installation.

Affects the electrical panel design and compliance of the installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are limited to ±10% and must not run in opposite directions, so supply quality must be verified before commissioning.

Prevents operation outside the permissible supply envelope that could damage the unit or affect performance.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

The device is rated IP20, so it must be installed in a dry indoor location protected from water ingress.

Determines the permissible environment and enclosure requirements at the installation site.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must remain between 5 °C and 40 °C for safe and specified operation.

Room conditions outside this range can degrade performance or trip protective functions.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation

The bath has a 7.5 l volume with a minimum filling capacity of 5.2 l, so the heat transfer fluid charge must be at least 5.2 l.

Underfilling can cause pump cavitation and poor temperature control; overfilling may overflow when hot.

Source
“Bath volume 7,5 l min. filling capacity 5,2 l”
Operation

The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar, which defines the external circuit flow and head available.

Needed to check whether the external loop resistance and required flow can be met.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar”
Selection

The bath is made of polycarbonate and is limited to a maximum of 100 °C, so it is not suitable for higher-temperature applications.

Material compatibility and temperature limit must be checked against the process requirements.

Source
“transparent baths made of polycarbonate with temperature range up to max. 100°C”
Operation

Polyglycol must not be used as heat transfer fluid when the Huber pump adapter is fitted, so fluid selection must account for the pump connection.

Using the wrong fluid with the adapter can damage the pump or compromise safety.

Source
“When using Huber pump adapter: Polyglycol is not permissible to be used as a heat transfer fluid”
Conformité

The unit is classified for electromagnetic disturbance as Class A, Group 1 to EN55011, so it is intended for industrial environments and may cause radio interference in residential areas.

Affects EMC planning and whether the unit can be installed in a residential or light-industrial location.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide a suitable plug and connection.

Affects the electrical installation scope and ordering of mating connectors.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Selection

The pressure equipment category is Art. 4.3 PED, which determines the applicable pressure-equipment compliance route.

Relevant for compliance documentation and acceptance of the unit in the plant.

Source
“Pressure equipment category Art. 4.3 PED”
Maintenance

The unit has adjustable overtemperature protection according to DIN 12876, which must be set and checked to protect the bath and process.

The overtemperature cut-out is a key safety function that must be configured during commissioning and verified during maintenance.

Source
“With adjustable overtemperature protection according to DIN 12876.”