Induztr
Résumé technique Visionneuse de documents Modèles associés

CC-130A Visco 3 Immersion Circulator Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and general description of the CC-130A Visco 3
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 4 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:394687

Technical data according to DIN 12876

CharacteristicCC-130A Visco 3
Operating temperature range28...100 °C
with water cooling15...100 °C
with refrigerator15...100 °C
Temperature stability at 70°C0,01 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pump max. delivery27 l/min
max. delivery pressure0,7 bar
Pump connectionM16x1 male
Bath volume30 l
min. filling capacity25,5 l
Width bath opening (mm)90 mm
Length bath opening (mm)90 mm
Number of bath openings3
Bath depth310 mm
Height of bath opening323 mm
Overall dimensions WxDxH **500x240x490 mm
Net weight14 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method: 28–100 °C without cooling, 15–100 °C with water cooling or refrigerator.

Determines whether the bath can reach sub-ambient or near-ambient setpoints required by the viscometer method.

Source
“Operating temperature range 28...100 °C with water cooling 15...100 °C with refrigerator 15...100 °C”
Installation

The unit requires a 200–240 V single- or two-phase supply with a minimum 10 A fuse and draws up to 10 A.

Electrical supply and branch protection must be sized accordingly before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A”
Installation

External branch circuit protection according to UL 489 is required.

Dictates the type of upstream breaker that must be provided in the installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±10% only if they do not run in opposite directions.

Site power quality must be checked; combined opposite deviations are not permitted.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

The correct plug type must be stated on the order to avoid rework at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

Ambient temperature must stay between 5 °C and 40 °C and the enclosure is IP20.

Defines the permissible room conditions and limits protection against water and dust ingress.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation

The bath holds 30 l with a minimum filling of 25,5 l and has three bath openings of 90 x 90 mm, 310 mm deep.

Determines the required bath fluid volume and the space available for inserting viscometers.

Source
“Bath volume 30 l min. filling capacity 25,5 l”
Performance

The pressure pump delivers up to 27 l/min at a maximum pressure of 0,7 bar through an M16x1 male connection.

Needed to match external circuit pressure drop and hose/connector thread.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar Pump connection M16x1 male”
Performance

Temperature stability at 70 °C is 0,01 K, supported by a Pt100 internal sensor and an external Pt100 connection.

Confirms the bath meets the tight stability needed for capillary viscometry and density measurement.

Source
“Temperature stability at 70°C 0,01 K”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

Overtemperature cut-out must be set and verified during commissioning for safe unattended operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Conformité

The unit is classified as pressure equipment under Art. 4.3 PED.

Affects the regulatory classification and documentation required for the installation.

Source
“Pressure equipment category Art. 4.3 PED”
Conformité

The unit is EMC class A, group 1 according to EN55011.

Class A equipment is intended for non-residential environments and may cause interference in domestic settings.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

An integral cooling coil for cooling water is provided for operation below or near ambient temperature.

Requires a cooling water supply to be plumbed in when sub-ambient operation is needed.

Source
“An integral cooling coil for cooling water is provided for temperature applications below or near ambient temperatures.”
Selection

A drain valve and a holder for Ubbelohde viscometers are optional accessories, not included.

These must be ordered separately if draining or viscometer mounting is required.

Source
“Optional accessories: drain valve, holder Ubbelohde viscosimeter #9586”