CC-304B Technical Data Sheet

Sommaire

  1. Overview p. 1
    Introduces the CC-304B model and document purpose
  2. Spec p. 2
    Technical data table per DIN 12876 standard
  3. p. 2 table, 3 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S392793

Engineering Facts

13 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method: 28–300 °C without cooling, 20–300 °C with water cooling, and -20–300 °C with a refrigerator.

Determines whether the bath can reach sub-ambient or near-ambient setpoints and which cooling utility must be provided.

Source
“Operating temperature range 28...300 °C with water cooling 20...300 °C with refrigerator -20...300 °C”
Performance

Temperature stability at 70 °C is 0.02 K, which is the achievable control uniformity for sensitive processes.

Needed to judge suitability for applications requiring tight temperature uniformity.

Source
“Temperature stability at 70°C 0,02 K”
Installation

The unit requires a 200–240 V single- or two-phase 50/60 Hz supply with a maximum current of 14 A and a 16 A fuse.

Defines the electrical branch circuit, breaker sizing, and whether the site supply is compatible.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 14 A Fuse 16 A”
Installation

Heating power varies with supply voltage from 3 kW at 230–240 V down to 2.2 kW at 200 V, so available voltage affects heat-up performance.

Low-voltage sites will see reduced heating capacity and longer heat-up times.

Source
“Heating power at 240V 3 kW Heating power at 230V 3 kW Heating power at 220V 2,8 kW Heating power at 208V 2,5 kW Heating power at 200V 2,2 kW”
Installation

The pump connections are M16x1 male, so hoses and adapters must match this thread.

Prevents ordering incompatible hoses or fittings during installation.

Source
“Pump connection M16x1 male”
Operation

The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s, limiting the choice of thermofluid at operating temperature.

Ensures the pump can circulate the fluid and protects against overload or poor flow.

Source
“max. permissible kin. viscosity 50 mm²/s”
Selection

The minimum filling capacity is 3.2 l and the bath volume is 5 l, setting the fluid charge and expansion allowance.

Determines the required fluid quantity and whether the bath volume suits the application.

Source
“min. filling capacity 3,2 l Bath volume 5 l”
Installation

The unit is rated IP20, so it must be installed in a dry location protected from water ingress.

Affects placement and enclosure requirements at the installation site.

Source
“Degree of Protection IP20”
Operation

The ambient temperature must remain between 5 °C and 40 °C for specified operation.

Room conditions outside this range can degrade performance or trip protections.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Conformité

External branch circuit protection according to UL 489 is required, so a UL-listed breaker must be provided in the supply.

Mandatory for compliant electrical installation in applicable markets.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±10% only if they do not run in opposite directions, e.g. -10% voltage with +3% frequency is not allowed.

Prevents installation on supplies whose combined voltage/frequency deviation exceeds the allowed envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Ensures the correct plug is supplied for the destination country and avoids rework.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the fluid and application.

Critical safety function that limits damage if control fails; must be configured during commissioning.

Source
“With adjustable overtemperature protection according to DIN 12876.”