Introduces the CC-405w circulating bath and its main features
Specpp. 1-2
Lists performance, temperature, and physical specifications per DIN 12876
p. 2table, 6 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Overall dimensions WxDxH **
370x460x679 mm
Power supply requirement
208V 2~ 60Hz
Degree of Protection
IP20
max. ambient temperature
40 °C
min. ambient temperature
5 °C
from Serial-No.:
1.1/25
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
-40...200 °C
Temperature stability at -10°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at 100°C
0,7 kW
Cooling power at 20°C
0,7 kW
Cooling power at 0°C
0,7 kW
Cooling power at -20°C
0,45 kW
Cooling power at -30°C
0,18 kW
Cooling power at -40°C
0,03 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)
Global Warming Potential (GWP)
0
Pressure pump
yes
max. delivery
25 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
18,5 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
46 l/h
min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar
Bath volume
5 l
min. filling capacity
4 l
Width bath opening WxD / bath depth
120 x 110 / 150 mm
Height of bath opening
480 mm
Engineering Facts
13 facts · every claim linked to a page
Selection
The unit uses flammable natural refrigerant R-1270 (A3, H220), so the installation location must be assessed for A3 refrigerant safety requirements.
A3 flammability classification affects room ventilation, ignition-source control and local code compliance at the selection and installation stage.
Source
“Refrigerant (ASHRAE, GHS) R-1270 (A3, H220)”
Installation
The cooling water circuit requires a differential pressure of at least 3 bar and must not exceed 6 bar, so the site water supply must be verified before installation.
Insufficient or excessive cooling water pressure will prevent rated cooling performance or damage the water circuit.
Source
“min. cooling water differential pressure 3 bar”
Installation
Cooling water intake must stay between 5 °C and 20 °C, and pipes must be insulated if water temperature can drop below the condensation point.
Out-of-range cooling water temperature invalidates the published cooling capacity and can cause condensation inside the unit.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Published cooling power is only valid at 20 °C room temperature, 15 °C cooling water inlet and 3 bar differential pressure, so actual capacity will differ at other site conditions.
The engineer must derate the cooling capacity when site conditions deviate from these reference values.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 3 bar differential pressure between cooling water inlet and -”
Installation
The unit requires a 208 V two-phase 60 Hz supply, so the electrical infrastructure must be matched to this non-standard voltage.
A 208 V two-phase supply is not universally available and determines the electrical connection design.
Source
“Power supply requirement 208V 2~ 60Hz”
Conformité
Voltage and frequency tolerances are coupled: ±5 % voltage is only allowed with ±2 % frequency in the same direction, and opposite-sign deviations are not permitted.
Supply quality must be checked against this combined tolerance rule to avoid invalidating operation or warranty.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit is IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C.
IP20 offers no water protection and the ambient limits define the permissible installation environment.
Source
“max. ambient temperature 40 °C”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Water chemistry must be compatible with these materials to avoid corrosion and fouling of the water circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
Three-phase devices below 63 A are delivered with cable but without plug, so the customer must provide the plug and connection.
This determines what electrical termination hardware must be procured for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The pump connections are M16x1 male and the maximum permissible kinematic viscosity is 50 mm²/s, which constrains the heat transfer fluid choice.
Fluid viscosity above this limit will overload the pump and reduce circulation performance.
Source
“max. permissible kin. viscosity 50 mm²/s”
Operation
The bath holds 5 l with a minimum filling capacity of 4 l, so the working fluid volume must be planned within this range.
Underfilling risks pump cavitation and poor temperature control; the bath volume sets the fluid inventory.
Source
“min. filling capacity 4 l”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
The overtemperature cut-out must be set correctly for the application to protect the bath and process.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Conformité
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which has implications for the installation environment.
Class A equipment is intended for industrial environments and may require EMC measures in residential or sensitive areas.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”