Introduces the CC-415wl immersion circulator and its main features
Specp. 2
Technical data per DIN 12876 including temperature and performance ratings
p. 2table, 15 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
max. cooling water pressure
6 bar
Bath volume
5 l
min. filling capacity
4 l
Width bath opening WxD / bath depth
120 x 110 / 150 mm
Height of bath opening
565 mm
Overall dimensions WxDxH **
410x480x764 mm
Net weight
61 kg
Power supply requirement
220-240V 1~/2~ 50/60Hz
max. current
9 A
min. Fuse
10A
max. Fuse
16A
Degree of Protection
IP20
max. ambient temperature
40 °C
min. ambient temperature
5 °C
from Serial-No.:
S433624
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-40...200 °C
Temperature stability at -10°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
1,6 kW
Heating power at 230V
1,5 kW
Heating power at 220V
1,3 kW
Cooling power at 100°C
1,2 kW
Cooling power at 20°C
1,2 kW
Cooling power at 0°C
1 kW
Cooling power at -20°C
0,6 kW
Cooling power at -30°C
0,2 kW
Cooling power at -40°C
0,05 kW
Refrigeration machine
air- and water-cooled, CFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
0,38 kg
CO2 equivalent
0,8 t
max. delivery
25 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
18,5 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
72 l/h
Consumption at water 15°C, flow 0°C
66 l/h
Consumption at water 15°C, flow -40°C
42 l/h
min. cooling water differential pressure
3 bar
Order-No.: CC-415wl; 2018.0002.01
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires cooling water with a minimum differential pressure of 3 bar between inlet and outlet, so the facility water supply must be verified before installation.
Insufficient differential pressure will prevent proper cooling and may cause the unit to trip or underperform.
Source
“min. cooling water differential pressure”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the cooling water must be compatible with these materials.
Aggressive or incompatible water will corrode the circuit and void reliable operation.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
If the cooling water temperature is below the dew point, the cooling water pipes inside the unit must be insulated to prevent condensation.
Condensation can cause electrical faults and water damage; insulation is a mandatory installation measure.
Source
“If the cooling water temperature is below the dew point, it is essential to insulate the cooling water pipes in the unit.”
Operation
The unit is designed for cooling water temperatures between 5 °C and 20 °C; higher water temperature reduces cooling power and may require increased flow.
Operating outside this range will degrade performance and may exceed the unit's cooling capacity.
Source
“The temperature control unit is designed for cooling water temperatures between 5 °C and 20 °C.”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
This asymmetric tolerance rule must be respected when assessing site power quality to avoid damage or malfunction.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Missing plug specification at order time will delay commissioning or require field wiring.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety
The refrigerant R-452A has a GWP of 2141 and a charge of 0.38 kg, corresponding to 0.8 t CO2 equivalent, which affects leak-check and F-gas obligations.
Operators must know the CO2 equivalent to comply with refrigerant leak-check and reporting regulations.
Source
“Global Warming Potential (GWP)”
Safety
The refrigerant is classified A1 with H280 hazard statement, indicating a non-flammable but pressurized gas hazard.
Personnel must be aware of the pressurized-gas hazard during handling and maintenance of the refrigeration circuit.
Source
“R-452A (A1, H280)”
Installation
The unit has IP20 degree of protection, so it must be installed in a dry indoor location protected from water ingress.
IP20 offers no water protection; installing in a wet or washdown area risks electrical failure.
Source
“Degree of Protection”
Installation
The ambient temperature must be between 5 °C and 40 °C; higher ambient temperature may reduce cooling capacity.
Site ambient conditions directly affect achievable cooling power and must be considered in selection.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation
The maximum cooling water pressure is 6 bar, so the supply must be regulated or limited to avoid overpressure damage.
Exceeding 6 bar can damage the cooling water circuit and internal components.
Source
“max. cooling water pressure”
Conformité
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which has implications for residential or sensitive environments.
Class A equipment may cause interference in residential areas and may require additional EMC measures.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation
The pump connection is M16x1 male and the cooling water connection is G1/2 male, so mating fittings must be procured accordingly.
Knowing the thread types prevents delays and ensures correct hose/pipe adaptation during installation.
Source
“Pump connection”
Maintenance
The maximum permissible kinematic viscosity of the pumped medium is 50 mm²/s, which limits the fluids that can be circulated.
Using a fluid with higher viscosity will overload the pump and reduce circulation performance.