CS 75 by Van der Heijden — Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...15 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperature stability at -10°C0,5 K
Interface digitalRS232
Safety classificationI / NFL
Cooling power at ambient temperature 20°C
at 15°C9 kW
at 10°C7,5 kW
at 0°C5,5 kW
at -10°C3 kW
at -20°C2,4 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Refrigerant quantity4,5 kg
CO2 equivalent6,29 t
Circulation pump:
max. delivery70 l/min
max. delivery pressure3,7 bar
Pump connectionG3/4 male
min. filling capacity160 l
max. filling capacity175 l
Overall dimensions WxDxH **800x850x1665 mm
Net weight265 kg
sound pressure level +/- 4 dB(A)66 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)14 A
Fuse (3 phase)3x20A
Degree of ProtectionIP52
min. ambient temperature5 °C
max. ambient temperature32 °C
from Serial-No.:1.0/24

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply with a 3x20 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device rating before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

The unit is delivered for three-phase connection below 63 A with a cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the bill of materials for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances must be respected simultaneously, and some combinations such as -5% voltage with +2% frequency are not allowed.

Prevents selecting a site supply or generator that would violate the permissible operating envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The chiller must be operated within an ambient temperature range of 5 °C to 32 °C.

Defines the allowable plant-room conditions and rules out cold or hot locations outside this range.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and may drop if the ambient temperature rises.

Prevents oversizing assumptions based on nominal capacity when the actual plant room is warmer.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Selection

The unit uses refrigerant R-449A with a CO2 equivalent of 6.29 t, which matters for refrigerant reporting and safety classification.

Needed for refrigerant inventory, leak reporting and safety classification of the installation.

Source
“R-449A (A1, H280)”
Installation

The chiller has IP52 protection, so it is protected against dust and dripping water but not against water jets or immersion.

Constrains cleaning methods and placement in wet or washdown environments.

Source
“Degree of Protection IP52”
Selection

The tank filling capacity is between 160 l and 175 l, which sets the required heat-transfer fluid volume.

Used to size the fluid charge and expansion volume for the system.

Source
“min. filling capacity 160 l”
Selection

The recommended thermofluid is a 50:50 water-monoethylene glycol mixture.

Ensures compatible fluid selection for the operating temperature range down to -20 °C.

Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Installation

The pump connection is G3/4 male, so mating hoses or adapters must match this thread.

Determines the connection hardware needed for the hydraulic circuit.

Source
“Pump connection G3/4 male”
Operation

Flow rate can be adjusted via the manual bypass valve on the back of the chiller, allowing the pump delivery to be matched to the application.

Important for commissioning and for avoiding excessive flow or pressure in the external circuit.

Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Conformité

The equipment is classified as EN55011 Class A Group 1, which has implications for electromagnetic compatibility in residential areas.

Class A equipment may not be suitable for residential environments without additional EMC measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Space requirements around the unit must be respected according to the operating conditions.

Affects layout planning and air-cooled condenser clearance for reliable operation.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”