KISS 220B Immersion Circulator Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and main features of KISS 220B
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 36 rows

KISS 220B Technical data according to DIN 12876

ParameterValue
Operating temperature range25...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
InterfaceAlarm message
optic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
max. delivery14 l/min
max. delivery pressure0,25 bar
max. delivery (suction)10,5 l/min
max. delivery pressure (suction)0,17 bar
Pump connenction (optional)M16x1 male
Bath volume17 l
min. filling capacity10 l
Height of bath opening205 mm
Width bath opening WxD/ bath depth290x329/ 150 mm
Overall dimensions WxDxH **350x555x375 mm
Net weight14 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:401723

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

14 facts · every claim linked to a page
Selection

The usable temperature range depends on the cooling method chosen, so the required range must be matched to water cooling or a refrigerator.

Determines whether an external water supply or an active refrigerator is needed for the target process temperature.

Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation

The unit accepts 200-240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A, so the electrical supply and protection must be sized accordingly.

Needed to plan the power feed, breaker and wiring before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A”
Installation

Branch circuit protection must be between 10 A and 16 A, which sets the fuse/breaker rating for the installation.

Defines the acceptable protective device rating to avoid nuisance trips or inadequate protection.

Source
“min. Fuse 10A max. Fuse 16A”
Installation

The device requires external branch circuit protection per UL 489, so a listed breaker must be provided in the installation.

Affects panel design and compliance of the electrical installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are limited to ±10% and must not run in opposite directions, so supply quality must be verified.

Prevents selecting a site with supply deviations that would violate the operating envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Operation

The pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic specified.

Critical for media compatibility when choosing the bath fluid and pump wetted parts.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glycerine or ethylene glycol rather than used pure.

Guides heat transfer fluid selection to avoid freezing and damage at low temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Selection

The bath holds 17 l with a minimum filling of 10 l, so the working volume and fluid quantity must be planned within these limits.

Determines fluid volume, expansion space and suitability for the sample load.

Source
“Bath volume 17 l min. filling capacity 10 l”
Selection

The bath opening is 290x329 mm with 150 mm depth and 205 mm height, which constrains the size of immersed vessels and accessories.

Needed to check that sample containers and racks physically fit.

Source
“Height of bath opening 205 mm Width bath opening WxD/ bath depth 290x329/ 150 mm”
Installation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so the location must be dry and within this temperature band.

Affects site selection and enclosure/room conditions for safe operation.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

The pressure pump delivers up to 14 l/min at 0.25 bar and the suction side up to 10.5 l/min at 0.17 bar, which bounds external circuit capability.

Used to verify the pump can serve the external loop resistance and flow demand.

Source
“max. delivery 14 l/min max. delivery pressure 0,25 bar max. delivery (suction) 10,5 l/min max. delivery pressure (suction) 0,17 bar”
Safety

The unit is safety class III/FL with adjustable overtemperature protection per DIN 12876, which is relevant for flammable-liquid operation.

Determines suitability for flammable bath fluids and the required safety concept.

Source
“Safety classification III / FL”
Operation

An autostart function restarts the unit after a power failure, which must be considered for unattended operation.

Affects risk assessment of automatic restart after mains interruption.

Source
“Autostart function for power failure”
Selection

An optional Pt100 sensor connection can display (not control) process temperature and is only available factory fitted.

Must be ordered with the unit and cannot be used for closed-loop control of process temperature.

Source
“Option: Pt100 sensor connection #10688 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge).”