KISS 225B Immersion Circulator

Sommaire

  1. Overview p. 1
    Product introduction and main features of KISS 225B
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 36 rows

KISS 225B Technical data according to DIN 12876

ParameterValue
Operating temperature range25...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
InterfaceAlarm message
optic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
max. delivery14 l/min
max. delivery pressure0,25 bar
max. delivery (suction)10,5 l/min
max. delivery pressure (suction)0,17 bar
Pump connenction (optional)M16x1 male
Bath volume23,5 l
min. filling capacity15 l
Height of bath opening255 mm
Width bath opening WxD/ bath depth290x329/ 200 mm
Overall dimensions WxDxH **350x555x425 mm
Net weight16 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.404517

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

14 facts · every claim linked to a page
Selection

The usable temperature range depends on the cooling method: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.

Determines whether the bath can meet the process temperature window and what cooling utility or chiller must be specified.

Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation

The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A, so branch protection must be sized between 10 A and 16 A.

Needed to confirm site power availability and to select the correct external fuse or circuit breaker.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

An external branch circuit protection device according to UL 489 is mandatory.

Affects electrical panel design and compliance for the installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±10% only if they do not combine in the opposite direction, so a -10% voltage with +3% frequency is not acceptable.

Prevents specifying the unit on a supply where combined voltage and frequency deviations exceed the allowed envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at order.

Ensures the correct plug is ordered and avoids rework at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic must be ordered.

Prevents premature pump failure when selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than using pure water.

Guides fluid selection to avoid freezing damage when operating near or below ambient.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Selection

The bath holds 23.5 l total but requires a minimum filling of 15 l, which sets the usable working volume.

Determines whether the bath volume is sufficient for the application and how much fluid to prepare.

Source
“Bath volume 23,5 l min. filling capacity 15 l”
Performance

The pressure pump delivers up to 14 l/min at 0.25 bar, while the suction side delivers up to 10.5 l/min at 0.17 bar.

Used to check whether the pump can supply an external circuit or connected equipment.

Source
“max. delivery 14 l/min max. delivery pressure 0,25 bar max. delivery (suction) 10,5 l/min max. delivery pressure (suction) 0,17 bar”
Safety

The unit is safety class III / FL with adjustable overtemperature protection according to DIN 12876.

Confirms the safety category required for flammable or sensitive applications and for compliance review.

Source
“Safety classification III / FL”
Installation

The enclosure is IP20, so it must be installed in a dry location protected from water ingress.

Restricts installation location and prevents specifying the unit in wet or washdown areas.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 °C and 40 °C for correct operation.

Defines the allowable room conditions at the installation site.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The bath bridge has a hole for a cooling probe such as an immersion cooler TC45–TC100E, and a holder for these coolers is available as an accessory.

Allows planning for additional cooling without a built-in refrigerator.

Source
“Bath bridge with hole for cooling probe (e.g. for immersion cooler TC45-TC100E).”
Operation

The autostart function restarts the unit automatically after a power failure.

Important for unattended processes where an automatic restart is desired or must be disabled.

Source
“Autostart function for power failure”